Skip to content

ORBIT GALAXY ENPOWER 641CF0721Recalls & reported problems

Medos International Sàrl · ORBIT GALAXY G2 microcoil Stretch Resistant Complex Fill Coil with Tight Distal Loop Technology 7mm x 21cm 190cm Contents: 1 Detachable Coil System Coil Type: Stretch Resistant Complex Fill Coil with Tight Distal Loop Technology · Class 2 · Implant

All models of ORBIT GALAXY ENPOWER (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10878528009617), received between Jun 2, 2017 and Nov 5, 2019 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Stretched” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Stretched2
  • Material Twisted/Bent1
  • Positioning Failure1
  • Positioning Problem1
  • Unraveled Material1

Patient problems

  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2019Stretched; Unraveled Material; Stretched; Unraveled Material

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COIL EMBOLIZATION, AN UNSPECIFIED MICROCATHETER (MC) WAS PLACED IN THE ANEURYSM. WHILE ADVANCING A 7MM X 21CM ORBIT GALAXY G2 COIL (641CF0721, S13606) AS THE SECOND COIL IN THE MICROCATHETER, PROXIMALLY IT DIDN'T SHOW ANY RESISTANCE AFTER THE DOCTOR PULLED THE COIL SHEATH BACKWARDS, HOWEVER, WHEN ADVANCING THE COIL, RESISTANCE OCCURRED AT THE MICROCATHETER HUB. WHEN THE COIL WAS ATTEMPTED TO RESHEATHED, IT BECAME STRETCHED. IT WAS NOT POSITIONED TO BE USED FOR THE COILING PROCEDURE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9280125 ↗
  • MalfunctionDec 14, 2018Stretched; Material Twisted/Bent; Positioning Problem; Stretched; Material Twisted/Bent; Positioning Problem

    THE HEALTHCARE PROFESSIONAL REPORTED THAT THE 7MM X 21CM G2 STRETCH RESISTANT COMPLEX FILL COIL (641CF0721/LOT# NOT AVAILABLE) ENCOUNTERED RESISTANCE FRICTION AND WAS BENT DURING THE PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY OR ADVERSE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8165215 ↗
  • MalfunctionJun 2, 2017Positioning Failure; Positioning Failure

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING COIL EMBOLIZATION OF A BASILAR ARTERY ANEURYSM, A G2TDL COMPLEX FILL COIL (641CF0721/ S12761) FAILED TO DETACH WITH USING CABLES WITH DIFFERENT LOTS AND A DETACHMENT CONTROL BOX (DCB00000500/C20654) . THE COIL WAS REMOVED FROM THE MICROCATHETER WITHOUT DIFFICULTY AND THE PROCEDURE WAS CONTINUED WITHOUT CONSEQUENCE FOR THE PATIENT; HOWEVER, THE PROCEDURE WAS DELAYED AND RISKIER DUE TO THE EVENT AND THE PROCEDURE WAS DELAYED 20 MINUTES. A PRE-DEPLOYMENT ELECTRICAL CHECK HAD BEEN PERFORMED. UPON PRESSING THE POWER BUTTON, ALL LIGHTS ILLUMINATED AND THE SYSTEM READY LIGHT ILLUMINATED. DURING THE DETACHMENT CYCLE, THE DETACHMENT LIGHT ILLUMINATED AND THE AUDIBLE SIGNAL BEEPED. ALL CONNECTIONS FIT PROPERLY WITHOUT NEED FOR EXCESSIVE FORCE. IT WAS REPORTED THAT PRODUCT WOULD BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6607575 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →