Magellan 8881892950Recalls & reported problems
Cardinal Health, Inc. · 1/2 mL Insulin Safety Syringe,Permanent Needle · Class 2
All models of Magellan (29)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10884521005945), received between Oct 31, 2018 and Aug 7, 2025 — classified as 6 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Fail-Safe Problem” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fail-Safe Problem8
- Activation Problem1
- Device Contaminated During Manufacture or Shipping1
- Material Separation1
- Nonstandard Device1
Patient problems
- Needle Stick/Puncture9
- No Patient Involvement2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2025Fail-Safe Problem; Fail-Safe Problem
THE CUSTOMER REPORTED THAT THEY HAD AN ACCIDENTAL NEEDLESTICK INCIDENT INVOLVING SYRINGE MODEL 8881892950 DISPENSED BY PHARMACY FOR LONG-ACTING INSULIN ADMINISTRATION. THE CORE SAFETY CONCERN IS THAT THE NEEDLE GUARD ON THIS SYRINGE DOES NOT ENGAGE SMOOTHLY. INITIALLY, IT SLIDES UP AND APPEARS TO STOP, WHICH CAN GIVE THE IMPRESSION THAT THE NEEDLE IS FULLY COVERED. HOWEVER, NURSES NEED TO APPLY ADDITIONAL FORCE BEYOND THIS INITIAL RESISTANCE TO FULLY ACTIVATE THE SAFETY GUARD AND PROPERLY COVER THE NEEDLE. THIS INCOMPLETE CLOSURE CREATES A RISK FOR ACCIDENTAL NEEDLESTICKS. ADDITIONAL INFORMATION RECEIVED ON 23JUL2025 STATED THAT THERE WAS NO MEDICAL TREATMENT NEEDED DUE TO IT BEING A CLEAN NEEDLE STICK.
Patient problems: Needle Stick/Puncture
MAUDE report 22741860 ↗MalfunctionMar 6, 2024Fail-Safe Problem; Fail-Safe Problem
THE CUSTOMER REPORTED THAT TWO ASSOCIATES TESTED A CLEAN UNUSED NEEDLE. THEY ENGAGED THE SAFETY DEVICE AND HEARD THE "CLICK". THEY WENT TO DISPOSE OF THE SYRINGE AND THE SAFETY HAD SLID DOWN ENOUGH THAT THE NEEDLE WAS IN FACT UNCOVERED. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18848988 ↗MalfunctionMar 6, 2024Fail-Safe Problem; Fail-Safe Problem
THE CUSTOMER REPORTED THAT THE ASSOCIATE ENGAGED THE SAFETY DEVICE AND HEARD THE "CLICK". SHE WENT TO DISPOSE OF THE SYRINGE AND THE SAFETY HAD SLID DOWN ENOUGH THAT THE NEEDLE WAS IN FACT UNCOVERED. THIS RESULTED IN A DIRTY NEEDLESTICK ON THE SIDE OF HER FINGER. ADDITIONAL INFORMATION RECEIVED ON 19FEB2024 STATED THAT NO FIRST AID OR PROPHYLACTIC TREATMENT WAS PROVIDED. THE ASSOCIATE DID HAVE BLOOD WORK DONE AND THEY HAVE THE RESULTS BUT THE INITIAL REPORTER DOES NOT HAVE ACCESS TO THAT INFORMATION.
Patient problems: Needle Stick/Puncture
MAUDE report 18848506 ↗MalfunctionAug 18, 2022Nonstandard Device; Nonstandard Device
THE CUSTOMER REPORTED THAT THE DEFECTIVE NEEDLE CAUSED A NURSE TO GET STUCK AS THE NEEDLE WENT THROUGH THE CAP AS WELL AS A SYRINGE WITH TWO NEEDLES.
Patient problems: Needle Stick/Puncture
MAUDE report 15254902 ↗InjuryDec 21, 2020Fail-Safe Problem; Fail-Safe Problem
THE CUSTOMER REPORTED THAT THE SAFETY LOCK DID NOT ENGAGE CORRECTLY AND THE NURSE WAS STUCK BY THE NEEDLE. THE RN WENT TO THE EMERGENCY DEPARTMENT FOR TESTING.
Patient problems: Needle Stick/Puncture
MAUDE report 11051283 ↗InjuryDec 21, 2020Fail-Safe Problem; Fail-Safe Problem
THE CUSTOMER REPORTED THAT THE SAFETY LOCK DID NOT ENGAGE CORRECTLY AND THE NURSE WAS STUCK BY THE NEEDLE. THE RN WENT TO ED FOR TESTING. ADDITIONAL INFORMATION RECEIVED STATED THE NEEDLE STICK OCCURRED WHEN THE NURSE WAS DISPOSING OF THE NEEDLE AFTER GIVING AN INSULIN INJECTION.
Patient problems: Needle Stick/Puncture
MAUDE report 11051266 ↗InjuryAug 31, 2020Fail-Safe Problem; Fail-Safe Problem
THE CUSTOMER REPORTED, AFTER GIVING AN INSULIN INJECTION, THE NURSE PUSHED UP ON THE SAFETY SHIELD TO ACTIVATE IT AND THE NEEDLE BENT AT ABOUT A 90 DEGREE ANGLE AND PUNCTURED THROUGH THE SAFETY CAP STABBING THE NURSE'S INDEX FINGER. A BANDAGE WAS PLACED ON THE FINGER AND NO FURTHER TESTING WAS DONE.
Patient problems: Needle Stick/Puncture
MAUDE report 10474084 ↗MalfunctionAug 11, 2020Material Separation; Material Separation
THE CUSTOMER REPORTED THAT THE INSULIN SYRINGE WAS FOUND BROKEN AT THE LEVEL OF THE NEEDLE AND THE SAFETY SHIELD IN THE PACKAGING. THE ISSUE WAS OBSERVED WITHOUT OPENING THE PACKAGE.
Patient problems: No Patient Involvement
MAUDE report 10393387 ↗InjuryFeb 17, 2020Activation Problem; Activation Problem
THE CUSTOMER REPORTED THAT AFTER ADMINISTERING THE INSULIN, WHEN ACTIVATING THE SAFETY ON INSULIN SYRINGE IT JAMMED AND THE TIP OF THE NEEDLE PRICKED THE USERS PALM.
Patient problems: Needle Stick/Puncture
MAUDE report 9717358 ↗InjuryAug 15, 2019Fail-Safe Problem; Fail-Safe Problem
THE CUSTOMER REPORTED THAT THE MAGELLAN INSULIN NEEDLE FLIPPED OUT OF THE SAFETY CAP CAUSING A NEEDLE STICK.
Patient problems: Needle Stick/Puncture
MAUDE report 8897454 ↗InjuryAug 15, 2019Fail-Safe Problem; Fail-Safe Problem
THE CUSTOMER REPORTED THAT THE MAGELLAN INSULIN NEEDLE FLIPPED OUT OF THE SAFETY CAP CAUSING A NEEDLE STICK.
Patient problems: Needle Stick/Puncture
MAUDE report 8897249 ↗MalfunctionOct 31, 2018Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
THE CUSTOMER REPORTS: WHEN USING (OPENING THE CAP) IT WAS NOTICED, THAT A PIECE OF PLASTIC WAS STICKING AT THE NEEDLE (COLOR SIMILAR TO THE COLOR OF THE CAP).
Patient problems: No Patient Involvement
MAUDE report 8025256 ↗
Other Syringe, piston devices
- BD Luer-Lok 309657Becton, Dickinson And Company · Syringe, piston1,266 reports
- BD Luer-Lok 309653Becton, Dickinson And Company · Syringe, piston948 reports
- BD Luer-Lok 309628Becton, Dickinson And Company · Syringe, piston862 reports
- N/A 302995Becton, Dickinson And Company · Syringe, piston1 recall537 reports
- BD Plastipak Syringe 300865Becton, Dickinson And Company · Syringe, piston515 reports
- N/A 302830Becton, Dickinson And Company · Syringe, piston462 reports
- N/A 302832Becton, Dickinson And Company · Syringe, piston424 reports
- BD Ultra-Fine 328468Becton, Dickinson and Company Inc · Syringe, piston323 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page