MAHURKAR 8888123435Recalls & reported problems
Mozarc Medical US LLC · Chronic Carbothane Catheter Sport Pack Staggered Tip with Sideholes · Class 2 · Implant
All models of MAHURKAR (81)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10884521012516), received between May 11, 2018 and May 11, 2018 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Occlusion Within Device” (10 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K111372); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jun 8, 2022Class IOpen, Classified
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Z-1320-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Occlusion Within Device10
Patient problems
- No Consequences Or Impact To Patient10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO, AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT ON MONDAY, THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN AN HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7509228 ↗MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO, AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT ON (B)(6), THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN A HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7509173 ↗MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO, AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT , THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN AN HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7509051 ↗MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO, AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT, THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN AN HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7508950 ↗MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO, AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT ON MONDAY, THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN AN HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7508691 ↗MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO, AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT ON MONDAY, THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN A HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7508666 ↗MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT ON MONDAY, THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN A HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7508635 ↗MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO, AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT ON MONDAY, THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN AN HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7508495 ↗MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO, AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT ON MONDAY, THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN AN HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7508462 ↗MalfunctionMay 11, 2018Occlusion Within Device; Occlusion Within Device
ACCORDING TO THE REPORTER, THE CATHETER¿S VENOUS RETURN WAS ZERO, AND THE CENTRAL LINE WAS OFTEN BLOCKED. IT WAS STATED THAT CONSEQUENCES FOR THE USERS ARE LONGER HOSPITALIZATION, TREATMENT LONGER OR POSTPONED (THEREFORE NOT COMPLETED), MORE CATHETER HANDLING, INCREASED RISK OF INFECTION AND OF VENOUS THROMBOSIS. IT WAS REPORTED THAT THE CATHETER WAS PATENT AND WORKING WELL IN THE RADIOLOGY SUITE, HOWEVER WHEN THE PATIENT SHOWED UP FOR TREATMENT ON MONDAY, THE CATHETER WAS NOT PATENT. THE PATENCY WAS CHECKED PRIOR TO HOOKING IN THE MACHINE AND IT WAS NOT WORKING. THE TREATMENT WAS NEVER STARTED. THE PROCEDURE WAS AN APHERESIS AND THEY WERE UNABLE TO RETRIEVE BLOOD. THE CATHETER WAS NOT MANIPULATED AND DRESSED USING ASEPTIC TECHNIQUE. THERE WAS NO REPORTED BLOOD LOSS AND THE PATIENT WAS JUST GIVEN AN HEPARIN 5000 UNITS (ANTICOAGULANT).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7508393 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111372. One submission often covers many models, so check the product and lot details in each record.
- Z-2632-2023 · Aug 22, 2023 · Open, ClassifiedPalindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm
- Z-1316-2022 · Jun 8, 2022 · Open, ClassifiedPalindrome Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page