Skip to content

MAHURKAR recallZ-1320-2022 · Covidien Llc · initiated Jun 8, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

FDA root cause category: Process control

What the company is doing

Medtronic distributed a recall notification to customers by mail on 06/08/2022. Customers were instructed to immediately quarantine and discontinue use for specific lots of chronic hemodialysis catheters, which are to be returned to Medtronic. If customers have further distributed product, they are asked to forward the information from the recall notification to those recipients. All customers are asked to complete and return the Consignee Product Retrieval Confirmation Form to [email protected]. Additionally, consignees are asked to share the notice within their organization and retain it for their records regarding the retrieval of unused product. If a consignee is aware of any adverse events associated with recalled product, they are to submit a report to FDA via Medwatch. Customers with any questions can contact their Medtronic Representative or call Customer Service at 800-962-9888, option 2. ***Updated 8/16/22*** Medtronic has created a website to help customers identify affected product: https://www.medtronic.com/covidien/en-us/products/c/renal-care-recall-lookup-tool.html#lotLookupTool

Recall details

Recall number
Z-1320-2022
Status
Open, Classified
Initiated by the company
Jun 8, 2022
Posted by the FDA
Jul 7, 2022
Recalling firm
Covidien Llc
Quantity
132,443 units
Product code
MSD
FDA submissions
K111372
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, DC, GU, & PR. The countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Republic of Korea, Kosovo, Kuwait, Lebanon, Luxembourg, Macao, Malaysia, Maldives, Malta, Mayotte, Mexico, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Viet Nam, Virgin Islands, and Zimbabwe.

Product description

Mahurkar Chronic Carbothane (Maxid) Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.

Code information (models, lots, serial numbers)

Show the full code information (20k characters)

Code Information organized as "Model No., Serial No., GTIN": 8888119362, 1928800230, 10884521056879; 8888119362, 1825400094, 20884521056876; 8888119362, 1912600107, 20884521056876; 8888119362, 1917600139, 20884521056876; 8888119362, 1917700080, 20884521056876; 8888119362, 1925200098, 20884521056876; 8888119362, 1927300146, 20884521056876; 8888119362, 1928800230, 20884521056876; 8888119362, 2023300144, 20884521056876; 8888119362, 2028200216, 20884521056876; 8888119362, 2122900239, 20884521056876; 8888119385, 1720700105, 20884521012421; 8888119385, 1724900109, 20884521012421; 8888119385, 2113300429, 20884521012421; 8888123402, 1927500121, 10884521056886; 8888123402, 2028200203, 10884521056886; 8888123402, 2110900059, 10884521056886; 8888123402, 1921700145, 20884521056883; 8888123402, 1924000138, 20884521056883; 8888123402, 1925200091, 20884521056883; 8888123402, 1927500120, 20884521056883; 8888123402, 1927500121, 20884521056883; 8888123402, 2023300143, 20884521056883; 8888123402, 2026000055, 20884521056883; 8888123402, 2028200203, 20884521056883; 8888123402, 2034400155, 20884521056883; 8888123402, 2104100105, 20884521056883; 8888123402, 2110900059, 20884521056883; 8888123402, 2113300432, 20884521056883; 8888123402, 2122900252, 20884521056883; 8888123435, 2034400156, 10884521012516; 8888123435, 1716500170, 20884521012513; 8888123435, 1716700115, 20884521012513; 8888123435, 1825400095, 20884521012513; 8888123435, 1831000101, 20884521012513; 8888123435, 2003700038, 20884521012513; 8888123435, 2034400156, 20884521012513; 8888123435, 2106900231, 20884521012513; 8888123435, 2113300430, 20884521012513; 8888128452, 2023300145, 10884521056916; 8888128452, 1716500171, 20884521056913; 8888128452, 1831000102, 20884521056913; 8888128452, 1904200079, 20884521056913; 8888128452, 2023300145, 20884521056913; 8888128485, 1716500172, 20884521012544; 8888128485, 1718600097, 20884521012544; 8888128485, 1720000119, 20884521012544; 8888133502, 1715100148, 20884521056920; 8888133535, 1924000139, 20884521013947; 8888145001, 1715800112, 10884521014084; 8888145001, 1722300112, 10884521014084; 8888145001, 1722600094, 10884521014084; 8888145001, 1723400093, 10884521014084; 8888145001, 1831000103, 10884521014084; 8888145001, 1904900119, 10884521014084; 8888145001, 1910200086, 10884521014084; 8888145001, 1912600085, 10884521014084; 8888145001, 1919700113, 10884521014084; 8888145001, 1925200099, 10884521014084; 8888145001, 2003700040, 10884521014084; 8888145001, 2008400137, 10884521014084; 8888145001, 2014900153, 10884521014084; 8888145001, 2014900166, 10884521014084; 8888145001, 2017500086, 10884521014084; 8888145001, 2017600091, 10884521014084; 8888145001, 2017600092, 10884521014084; 8888145001, 2017800060, 10884521014084; 8888145001, 1715300102, 20884521014081; 8888145001, 1715800112, 20884521014081; 8888145001, 1717200147, 20884521014081; 8888145001, 1717900143, 20884521014081; 8888145001, 1719300092, 20884521014081; 8888145001, 1720700107, 20884521014081; 8888145001, 1721300152, 20884521014081; 8888145001, 1722100148, 20884521014081; 8888145001, 1722300110, 20884521014081; 8888145001, 1722300111, 20884521014081; 8888145001, 1722300112, 20884521014081; 8888145001, 1722600094, 20884521014081; 8888145001, 1723400093, 20884521014081; 8888145001, 1723400094, 20884521014081; 8888145001, 1729100101, 20884521014081; 8888145001, 1810000171, 20884521014081; 8888145001, 1811500160, 20884521014081; 8888145001, 1827600114, 20884521014081; 8888145001, 1828200109, 20884521014081; 8888145001, 1829500106, 20884521014081; 8888145001, 1831000103, 20884521014081; 8888145001, 1834500194, 20884521014081; 8888145001, 1904200086, 20884521014081; 8888145001, 1904900119, 20884521014081; 8888145001, 1904900120, 20884521014081; 8888145001, 1905600120, 20884521014081; 8888145001, 1905600121, 20884521014081; 8888145001, 1906000118, 20884521014081; 8888145001, 1906000119, 20884521014081; 8888145001, 1910200086, 20884521014081; 8888145001, 1910200087, 20884521014081; 8888145001, 1910600255, 20884521014081; 8888145001, 1910600258, 20884521014081; 8888145001, 1912600085, 20884521014081; 8888145001, 1912600086, 20884521014081; 8888145001, 1919700113, 20884521014081; 8888145001, 1924000150, 20884521014081; 8888145001, 1925200099, 20884521014081; 8888145001, 2003700040, 20884521014081; 8888145001, 2008400137, 20884521014081; 8888145001, 2014900153, 20884521014081; 8888145001, 2014900164, 20884521014081; 8888145001, 2014900166, 20884521014081; 8888145001, 2014900167, 20884521014081; 8888145001, 2014900168, 20884521014081; 8888145001, 2014900169, 20884521014081; 8888145001, 2015300068, 20884521014081; 8888145001, 2017500086, 20884521014081; 8888145001, 2017600091, 20884521014081; 8888145001, 2017600092, 20884521014081; 8888145001, 2017800060, 20884521014081; 8888145001, 2124600201, 20884521014081; 8888145002, 1833100106, 10884521014145; 8888145002, 1914800162, 10884521014145; 8888145002, 1914800164, 10884521014145; 8888145002, 1920700130, 10884521014145; 8888145002, 2009300119, 10884521014145; 8888145002, 2027200195, 10884521014145; 8888145002, 2027200206, 10884521014145; 8888145002, 2034300098, 10884521014145; 8888145002, 1725100109, 20884521014142; 8888145002, 1728400115, 20884521014142; 8888145002, 1731200161, 20884521014142; 8888145002, 1732600137, 20884521014142; 8888145002, 1831600193, 20884521014142; 8888145002, 1833100106, 20884521014142; 8888145002, 1833800130, 20884521014142; 8888145002, 1906700180, 20884521014142; 8888145002, 1914400129, 20884521014142; 8888145002, 1914800162, 20884521014142; 8888145002, 1914800164, 20884521014142; 8888145002, 1917600065, 20884521014142; 8888145002, 1920700130, 20884521014142; 8888145002, 2009300119, 20884521014142; 8888145002, 2027200195, 20884521014142; 8888145002, 2027200206, 20884521014142; 8888145002, 2034300098, 20884521014142; 8888145003, 1727700140, 20884521014180; 8888145003, 1828900129, 20884521014180; 8888145003, 2001400108, 20884521014180; 8888145004, 1715100150, 20884521014203; 8888145004, 1725500097, 20884521014203; 8888145004, 1801000133, 20884521014203; 8888145027, 1720700106, 10884521056930; 8888145027, 1726300133, 10884521056930; 8888145027, 1805200080, 10884521056930; 8888145027, 1814300195, 10884521056930; 8888145027, 1822000073, 10884521056930; 8888145027, 1830300164, 10884521056930; 8888145027, 1926100268, 10884521056930; 8888145027, 1830300164, 20884521056937; 8888145027, 1916500080, 20884521056937; 8888145027, 1926100268, 20884521056937; 8888145027, 2003700043, 20884521056937; 8888145027, 2009300081, 20884521056937; 8888145027, 2026000074, 20884521056937; 8888145027, 2027300039, 20884521056937; 8888145027, 2034400151, 20884521056937; 8888145027, 2102500099, 20884521056937; 8888145027, 2104600166, 20884521056937; 8888145028, 1728400117, 10884521056947; 8888145028, 1805200081, 10884521056947; 8888145028, 1834500198, 10884521056947; 8888145028, 2034400152, 10884521056947; 8888145028, 2102500100, 10884521056947; 8888145028, 1834500198, 20884521056944; 8888145028, 1917600137, 20884521056944; 8888145028, 2002100143, 20884521056944; 8888145028, 2023300108, 20884521056944; 8888145028, 2031800136, 20884521056944; 8888145028, 2034400152, 20884521056944; 8888145028, 2102500100, 20884521056944; 8888145028, 2124600209, 20884521056944; 8888145029, 2031800133, 20884521056951; 8888145029, 2034400153, 20884521056951; 8888145030, 1729700114, 20884521056968; 8888145251, 1715100151, 10884521014220; 8888145251, 1715800113, 10884521014220; 8888145251, 1716500173, 10884521014220; 8888145251, 1717500004, 10884521014220; 8888145251, 1717900144, 10884521014220; 8888145251, 1718600098, 10884521014220; 8888145251, 1719300093, 10884521014220; 8888145251, 1720000120, 10884521014220; 8888145251, 1720700108, 10884521014220; 8888145251, 1721300153, 10884521014220; 8888145251, 1723400095, 10884521014220; 8888145251, 1727100080, 10884521014220; 8888145251, 1727700141, 10884521014220; 8888145251, 1731900229, 10884521014220; 8888145251, 1732600131, 10884521014220; 8888145251, 1733300154, 10884521014220; 8888145251, 1733700133, 10884521014220; 8888145251, 1733700188, 10884521014220; 8888145251, 1734700109, 10884521014220; 8888145251, 1800300132, 10884521014220; 8888145251, 1800300140, 10884521014220; 8888145251, 1805200106, 10884521014220; 8888145251, 1806600084, 10884521014220; 8888145251, 1808600071, 10884521014220; 8888145251, 1812900138, 10884521014220; 8888145251, 1812900139, 10884521014220; 8888145251, 1814600080, 10884521014220; 8888145251, 1827600117, 10884521014220; 8888145251, 1827600118, 10884521014220; 8888145251, 1831600195, 10884521014220; 8888145251, 1831600196, 10884521014220; 8888145251, 1833800140, 10884521014220; 8888145251, 1834500196, 10884521014220; 8888145251, 1834500206, 10884521014220; 8888145251, 1835100102, 10884521014220; 8888145251, 1835300203, 10884521014220; 8888145251, 1913000094, 10884521014220; 8888145251, 1914000293, 10884521014220; 8888145251, 1914000294, 10884521014220; 8888145251, 1914400122, 10884521014220; 8888145251, 1916500098, 10884521014220; 8888145251, 1916500099, 10884521014220; 8888145251, 1916500100, 10884521014220; 8888145251, 1926100264, 10884521014220; 8888145251, 1926100265, 10884521014220; 8888145251, 1926100266, 10884521014220; 8888145251, 1934400191, 10884521014220; 8888145251, 1934400192, 10884521014220; 8888145251, 2002100144, 10884521014220; 8888145251, 2014900172, 10884521014220; 8888145251, 2017500089, 10884521014220; 8888145251, 2017500132, 10884521014220; 8888145251, 2017500137, 10884521014220; 8888145251, 2019500267, 10884521014220; 8888145251, 2021000067, 10884521014220; 8888145251, 2021000068, 10884521014220; 8888145251, 2023300104, 10884521014220; 8888145251, 2034400176, 10884521014220; 8888145251, 2106900163, 10884521014220; 8888145251, 2106900204, 10884521014220; 8888145251, 2113300431, 10884521014220; 8888145251, 1715100151, 20884521014227; 8888145251, 1715800113, 20884521014227; 8888145251, 1716500173, 20884521014227; 8888145251, 1717500004, 20884521014227; 8888145251, 1717900144, 20884521014227; 8888145251, 1718600098, 20884521014227; 8888145251, 1719300093, 20884521014227; 8888145251, 1720000120, 20884521014227; 8888145251, 1720700108, 20884521014227; 8888145251, 1721300153, 20884521014227; 8888145251, 1723400095, 20884521014227; 8888145251, 1727100080, 20884521014227; 8888145251, 1727700141, 20884521014227; 8888145251, 1731900229, 20884521014227; 8888145251, 1732600131, 20884521014227; 8888145251, 1733300154, 20884521014227; 8888145251, 1733700133, 20884521014227; 8888145251, 1733700188, 20884521014227; 8888145251, 1734700109, 20884521014227; 8888145251, 1800300132, 20884521014227; 8888145251, 1800300140, 20884521014227; 8888145251, 1805200106, 20884521014227; 8888145251, 1806600084, 20884521014227; 8888145251, 1808600071, 20884521014227; 8888145251, 1812900138, 20884521014227; 8888145251, 1812900139, 20884521014227; 8888145251, 1814600080, 20884521014227; 8888145251, 1827600117, 20884521014227; 8888145251, 1827600118, 20884521014227; 8888145251, 1831600195, 20884521014227; 8888145251, 1831600196, 20884521014227; 8888145251, 1834500206, 20884521014227; 8888145251, 1835100102, 20884521014227; 8888145251, 1835300203, 20884521014227; 8888145251, 1913000094, 20884521014227; 8888145251, 1914000293, 20884521014227; 8888145251, 1914000294, 20884521014227; 8888145251, 1914400122, 20884521014227; 8888145251, 1914400123, 20884521014227; 8888145251, 1916500098, 20884521014227; 8888145251, 1916500099, 20884521014227; 8888145251, 1916500100, 20884521014227; 8888145251, 1926100264, 20884521014227; 8888145251, 1926100265, 20884521014227; 8888145251, 1926100266, 20884521014227; 8888145251, 1934400191, 20884521014227; 8888145251, 1934400192, 20884521014227; 8888145251, 2002100144, 20884521014227; 8888145251, 2008400141, 20884521014227; 8888145251, 2014900158, 20884521014227; 8888145251, 2014900172, 20884521014227; 8888145251, 2017500089, 20884521014227; 8888145251, 2017500132, 20884521014227; 8888145251, 2017500137, 20884521014227; 8888145251, 2019500267, 20884521014227; 8888145251, 2021000067, 20884521014227; 8888145251, 2021000068, 20884521014227; 8888145251, 2023300104, 20884521014227; 8888145251, 2034400174, 20884521014227; 8888145251, 2034400175, 20884521014227; 8888145251, 2034400176, 20884521014227; 8888145251, 2106900163, 20884521014227; 8888145251, 2106900204, 20884521014227; 8888145251, 2110900089, 20884521014227; 8888145251, 2113300431, 20884521014227; 8888145251, 2124600198, 20884521014227; 8888145251, 2201800126, 20884521014227; 8888145252, 1715100152, 10884521014244; 8888145252, 1717900145, 10884521014244; 8888145252, 1719300094, 10884521014244; 8888145252, 1720000121, 10884521014244; 8888145252, 1720700109, 10884521014244; 8888145252, 1721300168, 10884521014244; 8888145252, 1724300117, 10884521014244; 8888145252, 1727700139, 10884521014244; 8888145252, 1728400118, 10884521014244; 8888145252, 1729700117, 10884521014244; 8888145252, 1732700130, 10884521014244; 8888145252, 1801700129, 10884521014244; 8888145252, 1803100113, 10884521014244; 8888145252, 1803800135, 10884521014244; 8888145252, 1805900114, 10884521014244; 8888145252, 1805900133, 10884521014244; 8888145252, 1806600083, 10884521014244; 8888145252, 1807300175, 10884521014244; 8888145252, 1808600069, 10884521014244; 8888145252, 1811500157, 10884521014244; 8888145252, 1813600094, 10884521014244; 8888145252, 1817100080, 10884521014244; 8888145252, 1826200145, 10884521014244; 8888145252, 1826200167, 10884521014244; 8888145252, 1826800153, 10884521014244; 8888145252, 1828200107, 10884521014244; 8888145252, 1828200108, 10884521014244; 8888145252, 1828900131, 10884521014244; 8888145252, 1828900139, 10884521014244; 8888145252, 1828900140, 10884521014244; 8888145252, 1829500126, 10884521014244; 8888145252, 1830300140, 10884521014244; 8888145252, 1832300104, 10884521014244; 8888145252, 1832300105, 10884521014244; 8888145252, 1833100107, 10884521014244; 8888145252, 1833800142, 10884521014244; 8888145252, 1907800084, 10884521014244; 8888145252, 1907800085, 10884521014244; 8888145252, 1908400301, 10884521014244; 8888145252, 1913000179, 10884521014244; 8888145252, 1913000180, 10884521014244; 8888145252, 1913000181, 10884521014244; 8888145252, 1914000241, 10884521014244; 8888145252, 1914800206, 10884521014244; 8888145252, 1915800101, 10884521014244; 8888145252, 1917200226, 10884521014244; 8888145252, 1935100108, 10884521014244; 8888145252, 1935100109, 10884521014244; 8888145252, 2002900149, 10884521014244; 8888145252, 2009300077, 10884521014244; 8888145252, 2014900160, 10884521014244; 8888145252, 2015400078, 10884521014244; 8888145252, 2019500178, 10884521014244; 8888145252, 2019500270, 10884521014244; 8888145252, 2020400103, 10884521014244; 8888145252, 2020400104, 10884521014244; 8888145252, 2023300107, 10884521014244; 8888145252, 2027200208, 10884521014244; 8888145252, 2028000088, 10884521014244; 8888145252, 2030300103, 10884521014244; 8888145252, 2034400173, 10884521014244; 8888145252, 2104600208, 10884521014244; 8888145252, 2104600211, 10884521014244; 8888145252, 2104800077, 10884521014244; 8888145252, 2106900104, 10884521014244; 8888145252, 1717900145, 20884521014241; 8888145252, 1719300094, 20884521014241; 8888145252, 1720000121, 20884521014241; 8888145252, 1720700109, 20884521014241; 8888145252, 1721300168, 20884521014241; 8888145252, 1724300117, 20884521014241; 8888145252, 1727700139, 20884521014241; 8888145252, 1728400118, 20884521014241; 8888145252, 1800400103, 20884521014241; 8888145252, 1801700129, 20884521014241; 8888145252, 1805900114, 20884521014241; 8888145252, 1805900133, 20884521014241; 8888145252, 1806600083, 20884521014241; 8888145252, 1807300175, 20884521014241; 8888145252, 1808600069, 20884521014241; 8888145252, 1811500157, 20884521014241; 8888145252, 1813600094, 20884521014241; 8888145252, 1817100080, 20884521014241; 8888145252, 1826200167, 20884521014241; 8888145252, 1826800153, 20884521014241; 8888145252, 1828200107, 20884521014241; 8888145252, 1828200108, 20884521014241; 8888145252, 1828900140, 20884521014241; 8888145252, 1829500126, 20884521014241; 8888145252, 1830300140, 20884521014241; 8888145252, 1832300104, 20884521014241; 8888145252, 1832300105, 20884521014241; 8888145252, 1833100107, 20884521014241; 8888145252, 1833800142, 20884521014241; 8888145252, 1907800085, 20884521014241; 8888145252, 1908400301, 20884521014241; 8888145252, 1913000179, 20884521014241; 8888145252, 1913000180, 20884521014241; 8888145252, 1913000181, 20884521014241; 8888145252, 1914000241, 20884521014241; 8888145252, 1914800206, 20884521014241; 8888145252, 1915800101, 20884521014241; 8888145252, 1917200226, 20884521014241; 8888145252, 1935100108, 20884521014241; 8888145252, 1935100109, 20884521014241; 8888145252, 2002900149, 20884521014241; 8888145252, 2009300077, 20884521014241; 8888145252, 2014900160, 20884521014241; 8888145252, 2015400078, 20884521014241; 8888145252, 2019500174, 20884521014241; 8888145252, 2019500178, 20884521014241; 8888145252, 2019500270, 20884521014241; 8888145252, 2020400103, 20884521014241; 8888145252, 2020400104, 20884521014241; 8888145252, 2023300107, 20884521014241; 8888145252, 2027200208, 20884521014241; 8888145252, 2028000088, 20884521014241; 8888145252, 2030300103, 20884521014241; 8888145252, 2034400154, 20884521014241; 8888145252, 2034400173, 20884521014241; 8888145252, 2104100103, 20884521014241; 8888145252, 2104600208, 20884521014241; 8888145252, 2104600211, 20884521014241; 8888145252, 2104800077, 20884521014241; 8888145252, 2106900104, 20884521014241; 8888145252, 2124600211, 20884521014241; 8888145253, 1715800114, 10884521014275; 8888145253, 1716500174, 10884521014275; 8888145253, 1717200148, 10884521014275; 8888145253, 1717900197, 10884521014275; 8888145253, 1718600099, 10884521014275; 8888145253, 1719300095, 10884521014275; 8888145253, 1720000122, 10884521014275; 8888145253, 1721300154, 10884521014275; 8888145253, 1817700211, 10884521014275; 8888145253, 1828900082, 10884521014275; 8888145253, 1829500105, 10884521014275; 8888145253, 1829500127, 10884521014275; 8888145253, 1832300107, 10884521014275; 8888145253, 1833100105, 10884521014275; 8888145253, 1833800129, 10884521014275; 8888145253, 1917600122, 10884521014275; 8888145253, 1926100269, 10884521014275; 8888145253, 2026000061, 10884521014275; 8888145253, 2027200253, 10884521014275; 8888145253, 2102500101, 10884521014275; 8888145253, 2104100104, 10884521014275; 8888145253, 2124600200, 10884521014275; 8888145253, 1715800114, 20884521014272; 8888145253, 1716500174, 20884521014272; 8888145253, 1717200148, 20884521014272; 8888145253, 1717900197, 20884521014272; 8888145253, 1718600099, 20884521014272; 8888145253, 1719300095, 20884521014272; 8888145253, 1720000122, 20884521014272; 8888145253, 1721300154, 20884521014272; 8888145253, 1817700211, 20884521014272; 8888145253, 1829500127, 20884521014272; 8888145253, 1832300107, 20884521014272; 8888145253, 1833100105, 20884521014272; 8888145253, 1917600122, 20884521014272; 8888145253, 1926100269, 20884521014272; 8888145253, 2027200253, 20884521014272; 8888145253, 2102500101, 20884521014272; 8888145253, 2124600200, 20884521014272; 8888145254, 1715800115, 10884521014329; 8888145254, 1716500175, 10884521014329; 8888145254, 1717800131, 10884521014329; 8888145254, 1717900146, 10884521014329; 8888145254, 1718600100, 10884521014329; 8888145254, 1719300096, 10884521014329; 8888145254, 1720500101, 10884521014329; 8888145254, 1833100104, 10884521014329; 8888145254, 1835300224, 10884521014329; 8888145254, 1903500127, 10884521014329; 8888145254, 1917200227, 10884521014329; 8888145254, 2027200201, 10884521014329; 8888145254, 1715800115, 20884521014326; 8888145254, 1716500175, 20884521014326; 8888145254, 1717800131, 20884521014326; 8888145254, 1717900146, 20884521014326; 8888145254, 1718600100, 20884521014326; 8888145254, 1719300096, 20884521014326; 8888145254, 1720500101, 20884521014326; 8888145254, 1720700110, 20884521014326; 8888145254, 1833100104, 20884521014326; 8888145254, 1835300224, 20884521014326; 8888145254, 1903500127, 20884521014326

Devices named in this recall

20 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

23 devices cleared under K111372 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 90311).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page