Monoject 8506YRecalls & reported problems
Cardinal Health, Inc. · MONOJECT SHARPSTAR SAFETY IN-ROOM SHARPS CONTAINER, Counterbalance Lid, Yellow, 5 Quart · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521023260), received between Oct 6, 2021 and Feb 7, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Mechanical Jam2
- Mechanical Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Needle Stick/Puncture1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2023Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED THAT THE STAFF HAD A NEAR MISS (NEEDLE STICK) WHEN TRYING TO PUSH THE VACUTAINER IN. THE CONTAINER THAT WAS NOT OVERFILLED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16315285 ↗MalfunctionFeb 7, 2023Mechanical Problem; Mechanical Problem
THE CUSTOMER STATES AN EMPLOYEE REPORTED THAT THERE WAS A BLOODWORK NEEDLE STICKING (NOT RETRACTED) OUT OF THE SHARPS CONTAINER, WHILE WRITER WAS DISPOSING IV SHARPS AND WAS SUBSEQUENTLY POKED. THERE WAS A VACUTAINER THAT HAD NOT FALLEN COMPLETELY IN THE CONTAINER AND THE TUMBLER HAD NOT ENGAGED. THE STAFF DIDN¿T SEE THE NEEDLE STICKING OUT AND REACHED FOR THE CONTAINER. THEY SAID THAT THE CONTAINER WAS NOT FULL. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2023 STATED THAT AS A PRECAUTION THE EMPLOYEE WENT TO EMERGENCY AND SAW A PHYSICIAN. THEY ALSO HAD BLOOD WORK DONE AS A PRECAUTION, THE RESULTS ARE NOT AVAILABLE.
Patient problems: Needle Stick/Puncture; No Clinical Signs, Symptoms or Conditions
MAUDE report 16315077 ↗MalfunctionOct 14, 2021Mechanical Jam; Mechanical Jam
THE CUSTOMER REPORTED THAT THIS PERMANENT SUBSTITUTE IS VERY DIFFICULT TO USE DUE TO THE SMALL OPENING AND THE COUNTER-BALANCED LID. THE STAFF SPENDS EXCESS TIME ENSURING THE SHARPS DESCEND INTO THE CONTAINER. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER STATED THAT THEY DO NOT LIKE THE DESIGN OF THE COUNTER-BALANCED LID. THE LID IS SLOW TO ACCEPT ITEMS. THEY STOCK THESE CONTAINERS SO THEY HAVE MANY. WHEN SYRINGES WITH LABELS ATTACHED FROM THE OMNICELL ARE PLACED IN THE SHARPS CONTAINERS, THE LABELS END UP STICKING TO THE SIDES OF THE OPENING PREVENTING OTHER ITEMS FROM DROPPING IN. WHEN DEPOSITING LARGE ITEMS, LIKE CVC SHARPS AND NEEDLE/SYRINGES USED TO PERFORM BLOCKS, THE ITEMS GET STUCK IN THE NARROW OPENING. STAFF HAS TO SHAKE AND KNOCK THE CONTAINERS TO GET ITEMS TO FALL IN. THEY DID NOT HAVE ANY OF THESE ISSUES WITH THE OPEN LID DESIGN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12635854 ↗MalfunctionOct 6, 2021Mechanical Jam; Mechanical Jam
THE CUSTOMER REPORTED THAT THE COUNTER-BALANCED LID ON THE SHARPS CONTAINER GETS STUCK AND JAMS. SHARPS ARE NOT COMPLETELY DEPOSITED INTO CONTAINER AND SHARPS ARE LEFT EXPOSED WHICH INCREASES RISK OF NEEDLESTICK INJURY. THESE CONTAINERS ARE LOCATED THROUGHOUT THE HOSPITAL. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12586173 ↗
Other Container, sharps devices
- SharpSafety 8556HCardinal Health, Inc. · Container, Sharps65 reports
- Monoject 8507SACardinal Health, Inc. · Container, sharps13 reports
- BD Recykleen™ Sharps Collector 305160Becton, Dickinson And Company · Container, sharps6 reports
- Monoject 31142222Cardinal Health, Inc. · Container, sharps4 reports
- Monoject 8980SCardinal Health, Inc. · Container, sharps4 reports
- Monoject 8970YCardinal Health 200, LLC · Container, Sharps3 reports
- Monoject 85221RCardinal Health, Inc. · Container, sharps3 reports
- Monoject 8932PG2Cardinal Health, Inc. · Container, sharps3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page