Kangaroo 763656Recalls & reported problems

Cardinal Health, Inc. · Joey Pump Set,Anti-free Flow · Class 2

Quick summary

This Cardinal Health Joey Pump Set is a feeding pump set used to deliver enteral nutrition. The 48 reports mostly describe air or gas getting into the tubing, constant pump alarms, and bags that swell, leak, or tear. No deaths or injuries were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
48
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 48 adverse event reports that list this exact device (DI 10884521072305), received between Sep 15, 2017 and Apr 10, 2024 — classified as 48 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak27
  • Air/Gas in Device12
  • Infusion or Flow Problem6
  • Detachment of Device or Device Component2
  • Leak/Splash2
  • Device Displays Incorrect Message1
  • Failure to Prime1
  • Improper Flow or Infusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions25
  • No Consequences Or Impact To Patient16
  • No Known Impact Or Consequence To Patient6
  • Discomfort1

Recent adverse event reports

The 25 most recent of 48 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2024Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THE FEEDING BAG BLEW UP WITH AIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19079872 ↗
  • MalfunctionMar 29, 2024Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THAT 7 FEEDING BAGS HAVE FAILED SO FAR THIS MONTH. IT SEEMS LIKE THEY MOSTLY FAIL IF THEY PAUSE THEIR DAUGHTER'S FEED AND THEN RESTART. THE BAGS BLOW UP WITH AIR INSIDE THE JOEY MACHINE. PER ADDITIONAL INFORMATION PROVIDED ON 27MAR2024 THE BAGS ALSO LEAKED AS WELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19004267 ↗
  • MalfunctionJun 1, 2023Air/Gas in Device; Air/Gas in Device

    PER CUSTOMER, THE PATIENT REPORTED THAT AIR GOT INTO THE PUMP SET TUBING. THERE WAS NO HARM REPORTED AND NO OTHER ISSUES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17043245 ↗
  • MalfunctionOct 11, 2022Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THAT THE PATIENT HAS USED 10 BAGS SO FAR AND ALL ARE CAUSING THE ALARMS ON THE PUMP TO CONSTANTLY GO OFF AND AIR IS STUCK IN THE TUBING. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15579347 ↗
  • MalfunctionMar 24, 2022Air/Gas in Device; Air/Gas in Device

    CUSTOMER REPORTS: TINY AIR BUBBLES APPEARED IN THE FEEDING TUBE SECTION BETWEEN THE PUMP AND THE PATIENT. THE PATIENT WAS EXPERIENCING DISCOMFORT FOR THE LAST WEEK AND MOTHER JUST NOTICED THE AIR BUBBLES IN THE TUBING TODAY. THE PATIENT IS UNABLE TO COMMUNICATE VERBALLY.

    Patient problems: Discomfort

    MAUDE report 13893162 ↗
  • MalfunctionDec 23, 2021Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THAT THERE WAS AIR IN THE TUBING (BETWEEN PUMP AND PATIENT). IT WAS APPROXIMATELY 7-8 INCHES OF AIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13078761 ↗
  • MalfunctionDec 23, 2021Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THAT THERE WAS AIR IN THE TUBING (BETWEEN PUMP AND PATIENT). IT WAS APPROXIMATELY 7-8 INCHES OF AIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13078751 ↗
  • MalfunctionJul 23, 2021Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THAT THE BAGS WERE CONTINUOUSLY CAUSING AN ERROR THAT SHOWED THAT THERE WAS AIR STUCK IN THE LINE. THE BAGS WERE UNABLE TO BE USED CORRECTLY. THERE WAS NO PATIENT INJURY. ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER STATED THAT THERE WAS AIR IN THE LINE OF THE FEEDING SET. THEY TRIED TO EXCHANGE THE PUMP TO SEE IF THAT WAS THE ISSUE BUT THE BAGS CONTINUED TO HAVE AIR IN THE LINE AFTER THAT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12220906 ↗
  • MalfunctionJun 14, 2021Air/Gas in Device; Air/Gas in Device

    CUSTOMER REPORTS: AIR ENTERED INTO THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11991629 ↗
  • MalfunctionMay 13, 2021Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THAT AIR IN THE TUBING IS FOUND TO BE INTERMITTENT. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11817027 ↗
  • MalfunctionApr 30, 2021Infusion or Flow Problem; Infusion or Flow Problem

    THE CUSTOMER REPORTED LARGE AIR GAPS IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11753049 ↗
  • MalfunctionApr 26, 2021Infusion or Flow Problem; Infusion or Flow Problem

    THE CUSTOMER REPORTED AIR WAS GETTING IN THE TUBING. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11726689 ↗
  • MalfunctionMar 24, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT THE SETS ARE LEAKING FROM THE BOTTOM OF THE BAG WHERE THE TUBE IS. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11563421 ↗
  • MalfunctionMar 15, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT FOOD WAS LEAKING WHERE THE BAG CONNECTS TO THE TUBING. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11490250 ↗
  • MalfunctionMar 4, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT THE TUBING IS NOT CONNECTED ALL THE WAY AND THERE IS A SLIGHT HOLE CAUSING THE BAG TO LEAK. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11414213 ↗
  • MalfunctionFeb 26, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT THE TUBING IS NOT CONNECTED ALL THE WAY AND THERE IS A SLIGHT HOLE CAUSING THE BAG TO LEAK. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11388393 ↗
  • MalfunctionFeb 25, 2021Infusion or Flow Problem; Infusion or Flow Problem

    THE CUSTOMER REPORTED AIR WAS IN THE LINE OF THE FEEDING SET. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11379307 ↗
  • MalfunctionFeb 10, 2021Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn

    THE CUSTOMER REPORTED THE BAG SPLIT AND LEAKED THE FORMULA. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11308189 ↗
  • MalfunctionJan 29, 2021Fluid/Blood Leak; Air/Gas in Device; Fluid/Blood Leak; Air/Gas in Device

    THE PATIENT'S MOTHER REPORTED THAT THERE WERE ISSUES WITH THE FEEDING SETS TUBINGS. THE ROD WHICH IS USUALLY INSERTED OR THE ROUND PIECE WHICH IS INSERTED INTO THE FEEDING PUMP CAME OFF THE TUBING ON IT'S OWN, WHEN CHANGING THE BAG IT LEAKED EVERY WHERE, AND THERE WERE AIR BUBBLES IN THE LINE. THERE WERE NO INJURIES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11254862 ↗
  • MalfunctionJan 20, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT THEY FOUND LITTLE CRACKS IN THE TUBING OF THE ENTERAL FEEDING SET. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 15-JAN-2021 STATED THAT THE CRACKS ON THE TUBING WERE ALL THE WAY THROUGH CAUSING THE TUBES TO LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11205672 ↗
  • MalfunctionJan 15, 2021Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT WHEN THEY ARE PLUGGED INTO THE PUMP FOR FEEDING, A LEAK WAS OBSERVED. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11178714 ↗
  • MalfunctionJan 13, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT WHEN THEY ARE PLUGGED INTO THE PUMP FOR FEEDING, A LEAK WAS OBSERVED. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11158486 ↗
  • MalfunctionDec 31, 2020Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THAT THERE WERE AIR GAPS THAT WERE 3-4 INCH INCHES LONG APPEARING IN THE TUBING OF THE FEEDING SET BETWEEN THE PUMP AND THE PATIENT. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11100671 ↗
  • MalfunctionDec 31, 2020Infusion or Flow Problem; Infusion or Flow Problem

    THE CUSTOMER REPORTED THERE WERE AIR GAPS 3-4 INCHES LONG APPEARING IN THE TUBING BETWEEN THE PUMP AND THE PATIENT. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11100467 ↗
  • MalfunctionDec 31, 2020Infusion or Flow Problem; Infusion or Flow Problem

    THE CUSTOMER REPORTED THERE WERE AIR GAPS 3-4 INCHES LONG APPEARING IN THE TUBING BETWEEN THE PUMP AND THE PATIENT. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11100425 ↗

Other PUMP, INFUSION, ENTERAL devices

All PUMP, INFUSION, ENTERAL devices →