Skip to content

Sofsilk S-317Recalls & reported problems

Covidien LP · Wax Coated Braided Silk · Class 2 · Implant

All models of Sofsilk (120)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

11 other recalls cover devices cleared under the same FDA submission (K981128); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Dec 21, 2023Open, Classified

    Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.

    Z-0845-2024 · Root cause: Process control · Recall details · FDA record ↗

  2. Dec 2, 2022Open, Classified

    Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

    Z-1176-2023 · Root cause: Packaging process control · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K981128. One submission often covers many models, so check the product and lot details in each record.

  • Z-1697-2024 · Jan 25, 2024 · Open, Classified
    Sofsilk Braided Silk sutures, Product Number S-2782K. Nonabsorbable, sterile, non-mutagenic surgical suture composed of natural silk fibers, for use in soft tissue approximation an
  • Z-2814-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 4/0 18" BLACK CV-22 DT non-absorbable silk sutures GS46M Sofsilk is intended for use in general soft tissue approximation and/or ligation, including use in the c
  • Z-2812-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 5/0 BLACK 5 X 18" CV-22 D*TACH non-absorbable silk sutures GS43M Sofsilk is intended for use in general soft tissue approximation and/or ligation, including use
  • Z-2813-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 4/0 18" BLACK CV-22 DT non-absorbable silk sutures GS44M Sofsilk is intended for use in general soft tissue approximation and/or ligation, including use in the
  • Z-2817-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 2/0 30" BLACK V-20 DT non-absorbable braided nylon sutures GS66M Sofsilk is intended for use in general soft tissue approximation and/or ligation, including use
  • Z-2816-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 3/0 18" BLACK V-20 DT non-absorbable braided nylon sutures GS63M Sofsilk is intended for use in general soft tissue approximation and/or ligation, including use
  • Z-2819-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 0 BLACK 6X 18" PRE-CUT WAX/RACE non-absorbable braided nylon sutures S196 Sofsilk is intended for use in general soft tissue approximation and/or ligation, inclu
  • Z-2811-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 3/0 18" BLACK CV-25 DT non-absorbable silk sutures GS33M Sofsilk is intended for use in general soft tissue approximation and/or ligation, including use in the c
  • Z-2818-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 3/0 BK 12X18" PRE-CUT WAX-RACE non-absorbable braided nylon sutures S194 Sofsilk is intended for use in general soft tissue approximation and/or ligation, includ
  • Z-2815-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 2/0 BLACK 5x18" V-20 D-TACH non-absorbable silk sutures GS62M Sofsilk is intended for use in general soft tissue approximation and/or ligation, including use in
  • Z-2810-2011 · Jun 3, 2011 · Terminated
    Syneture Sofsilk 0 18" BLACK GS-21 DT non-absorbable braided nylon sutures CS10M Sofsilk is intended for use in general soft tissue approximation and/or ligation, including use i

Other SUTURE, NONABSORBABLE, SILK devices

All SUTURE, NONABSORBABLE, SILK devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page