Sofsilk Braided Silk sutures recallZ-0845-2024 · Covidien LP · initiated Dec 21, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
FDA root cause category: Process control
What the company is doing
On December 21, the firm notified affected consignees via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to identify, quarantine, and return all unused, non-expired affected sutures to Medtronic. Please contact [email protected] for the Return Good Authorization (RGA). Customers will receive credit for returned affected product. If purchased from a distributor, please contact your distributor directly to arrange for the return of product back to your distributor. Pass on the recall notice to all those who need to be aware within your organization, or to any organization where the affected sutures have been transferred or distributed. If you have questions regarding this recall, please call Medtronic Quality Assurance at 800-255-6774 option 1, then option 1 ***Updated 4/1/2024*** Firm identified additional models and lots affected by the gamma radiation issue and sent additional notification to customers on January 25, 2024.
Recall details
- Recall number
- Z-0845-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 21, 2023
- Posted by the FDA
- Jan 29, 2024
- Recalling firm
- Covidien LP
- Quantity
- 46,812 US; 68,976 OUS
- Product code
- GAP
- FDA submissions
- K981128
- Distribution
- Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
Product description
Sofsilk Braided Silk sutures: CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 GS-831 SOFSILK* 4-0 BLK 75CM V20 X36 S-176 SOFSILK* 0 BLK 12X45CM PCT X24 S-182 SOFSILK* 5-0 BLK 12X45CM PCT X36 S-184 SOFSILK* 3-0 BLK 12X45CM PCT X36 S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36 S-245 SOFSILK* 2-0 BLK 12X60CM PCT X36 S-246 SOFSILK* 0 BLK 6X60CM PCT X36 S-2782K SOFSILK* 4-0 WHI 45CM SS24 DA S-303 SOFSILK* 4-0 BLK 12X75CM PCT X36 S-304 SOFSILK* 3-0 BLK 12X75CM PCT X36 S-305 SOFSILK* 2-0 BLK 12X75CM PCT X36 S-316 SOFSILK* 0 BLK 12X75CM PCT X24 S-317 SOFSILK* 1 BLK 12X75CM PCT X24 S-605 SOFSILK* 2-0 BLK 2X150CM PCT X36 S-606 SOFSILK* 0 BLK 2X150CM PCT X36 S606-12 SOFSILK* 0 BLK 2X150CM PCT X12 S-610 SOFSILK* 5 BLK 2X150CM PCT X12 SOFSILK 2-0 30 BK CV-24 GS453-2 SOFSLK 4-0 24 BK CV-15 DA VS581-2 ***Added 4/1/24*** CS-792 SOFSILK* 0 BLK 100CM GCC90 X36 S-1172 SOFSILK* 6-0 BLK 45CM C1 X36 S-1735 SOFSILK* 4-0 BLK 30CM C1 DA X12 S-1750K SOFSILK* 6-0 BLK 45CM HE1 DAX12 S-1768K SOFSILK* 7-0 BLK 45CM HE6 DAX12 S-1789K SOFSILK* 4-0 BLK 45CM HE3 X12 S-187 SOFSILK* 1 BLK 6X45CM PCT X36 S-243 SOFSILK* 4-0 BLK 12X60CM PCT X36 S-2780K SOFSILK* 6-0 BLK 45CM SS-24DA S-282 SOFSILK* 5 BLK 3X60CM PCT X36 S-318 SOFSILK* 2 BLK 12X75CM PCT X24 S-608 SOFSILK* 2 BLK 2X150CM PCT X36 VS-872 SOFSILK* 3-0 BLK 75CM CV23 X36 VS872-2 SOFSILK 3-0 30 BK CV-23 VS872-2 SS-1694G SOFSILK 4/0 BLK 45CM P24 X12 SS-5641 SOFSILK* 4-0 BLK 45CM P13 X36 SS-5649G SOFSILK* 6-0 BLK 45CM P13 X12 SS-645 SOFSILK* 4-0 BLK 75CM C12 X36 SS-675 SOFSILK* 3-0 BLK 75CM C16 X36 SS-681 SOFSILK* 6-0 BLK 45CM C12 X36 VS-872 SOFSILK* 3-0 BLK 75CM CV23 X36 XX-5280 SOFSILK* 0 BLK 150CM PCT X36 -- Ti-Cron Coated Braided Polyester Sutures: 3087-31 TICRON* 4-0 BLU 75CM CV-331DA 3226-31 TICRON* 4-0 BLU 60CM CV330 DA 88863026-71 TICRON* 1 BLU 75CM HOS12X36 88863070-51 TICRON* 2-0 BLU 75CM C15X36 88863087-31 TICRON* 4-0 BLU 75CM CV33DA 88863092-71 TICRON* 1 BLU 75CM KV37X36 88863147-83 TICRON 2 BLU 5X75CM KV37X12 88863185-41 TICRON* 3-0 BLU 75CM CV305 88863186-31 TICRON* 4-0 BLU 75CM CV331 88863226-31 TICRON* 4-0 BLU 60CM CV330DA 88863226-41 TICRON* 3-0 BLU 60CM CV330DA 88863229-31 TICRON* 4-0 BLU 75CM CV307DA 88863271-31 TICRON* 4-0 BLU 90CM CV331DA 88863309-71 TICRON* 1 BLU 75CM GS11X36 -- Monosof Monofilament Nylon Sutures: N-63 MONOSOF* 3-0 BLK 12X45CM PCT X36 SN-3695 MONOSOF* 5-0 BLK 45CM P10 X12 -- Steel Monofilament Stainless Steel Sutures 88862224-49 STEEL 7 4X45CM SCC X12
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
Model CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 UDI-DI 20884521083629 Lot D1L1576RY Model GS-831 SOFSILK* 4-0 BLK 75CM V20 X36 UDI-DI 20884521085203 Lot D2B1216RY Model S-176 SOFSILK* 0 BLK 12X45CM PCT X24 UDI-DI 20884521085746 Lot D2D0612RY Model S-182 SOFSILK* 5-0 BLK 12X45CM PCT X36 UDI-DI 20884521085944 Lot D2B0728RY Model S-184 SOFSILK* 3-0 BLK 12X45CM PCT X36 UDI-DI 20884521085968 Lots D2D0972RY D2E2396RY Model S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36 UDI-DI 20884521085975 Lots D2D0957RY D2D0968RY D2D0971RY D2D0984RY D2D0985RY D2E0900RY D2E0902RY Model S-245 SOFSILK* 2-0 BLK 12X60CM PCT X36 UDI-DI 10884521086197 Lot D2E0674RY UDI-DI 20884521086194 Lots D2D1871RY D2E0674RY D2E0901RY Model S-246 SOFSILK* 0 BLK 6X60CM PCT X36 UDI-DI 20884521086200 Lot D2D0980RY Model S-2782K SOFSILK* 4-0 WHI 45CM SS24 DA UDI-DI 20884521086422 Lot D2F2101RY Model S-303 SOFSILK* 4-0 BLK 12X75CM PCT X36 UDI-DI 20884521086460 Lots D2D1872RY D2D2120RY Model S-304 SOFSILK* 3-0 BLK 12X75CM PCT X36 UDI-DI 20884521086477 Lot D2D1991RY Model S-305 SOFSILK* 2-0 BLK 12X75CM PCT X36 UDI-DI 20884521086484 Lots D2E0907RY D2E2685RY Model S-316 SOFSILK* 0 BLK 12X75CM PCT X24 UDI-DI 20884521086491 Lot D2D2436RY Model S-317 SOFSILK* 1 BLK 12X75CM PCT X24 UDI-DI 20884521086514 Lot D2E0587RY Model S-605 SOFSILK* 2-0 BLK 2X150CM PCT X36 UDI-DI 20884521086620 Lot D2E0676RY Model S-606 SOFSILK* 0 BLK 2X150CM PCT X36 UDI-DI 10884521086630 Lot D2B2638RY UDI-DI 20884521086637 Lots D2B2638RY D2D0961RY Model S606-12 SOFSILK* 0 BLK 2X150CM PCT X12 UDI-DI 10884521637887 Lot D2D0961RY Model S-610 SOFSILK* 5 BLK 2X150CM PCT X12 UDI-DI 20884521086675 Lot D2B1535RY Model SOFSILK 2-0 30 BK CV-24 GS453-2 UDI-DI 20884521758121 Lots D2D0254RY D2D1576RY Model SOFSLK 4-0 24 BK CV-15 DA VS581-2 UDI-DI 20884521757971 Lot D2E1431RY ***Added 4/1/24*** Model CS-792 SOFSILK* 0 BLK 100CM GCC90 X36 UDI-DI 20884521084756 Lot D2C0576RY Model S-1172 SOFSILK* 6-0 BLK 45CM C1 X36 UDI-DI 20884521085470 Lot D2J0691RY Model S-1735 SOFSILK* 4-0 BLK 30CM C1 DA X12 UDI-DI 20884521085630 Lot D1L3318FRY Model S-1750K SOFSILK* 6-0 BLK 45CM HE1 DAX12 UDI-DI 20884521085739 Lot D2E2348RY Model S-1768K SOFSILK* 7-0 BLK 45CM HE6 DAX12 UDI-DI 20884521085784 Lot D2B2217RY Model S-1789K SOFSILK* 4-0 BLK 45CM HE3 X12 UDI-DI 20884521085876 Lot D2D0262RY Model S-185 SOFSILK* 2-0 BLK 12X45CM PCT X36 UDI-DI 20884521085975 Lots D2B2878RY D2E2680RY D2E2769RY Model S-187 SOFSILK* 1 BLK 6X45CM PCT X36 UDI-DI 20884521085982 Lot D2A1585RY Model S-243 SOFSILK* 4-0 BLK 12X60CM PCT X36 UDI-DI 20884521086170 Lots D1M1020RY D2D2348RY D2E1492RY Model S-246 SOFSILK* 0 BLK 6X60CM PCT X36 UDI-DI 10884521086203 Lot D2E1375RY UDI-DI 20884521086200 Lots D2E1375RY D2E1488RY Model S-2780K SOFSILK* 6-0 BLK 45CM SS-24DA UDI-DI 20884521086408 Lot D2C1916RY Model S-282 SOFSILK* 5 BLK 3X60CM PCT X36 UDI-DI 20884521086453 Lot D2E1470RY Model S-303 SOFSILK* 4-0 BLK 12X75CM PCT X36 UDI-DI 20884521086460 Lots D2E1495RY D2E1496RY Model S-304 SOFSILK* 3-0 BLK 12X75CM PCT X36 UDI-DI 20884521086477 Lot D2E1472RY Model S-305 SOFSILK* 2-0 BLK 12X75CM PCT X36 UDI-DI 20884521086484 Lots D2C1479RY D2E2772RY Model S-318 SOFSILK* 2 BLK 12X75CM PCT X24 UDI-DI 20884521086538 Lots D1M0153RY D2A1677RY Model SS-1694G SOFSILK 4/0 BLK 45CM P24 X12 UDI-DI 20884521176406 Lot D1M0141FRY Model S-605 SOFSILK* 2-0 BLK 2X150CM PCT X36 UDI-DI 20884521086620 Lot D2E1491RY Model S-608 SOFSILK* 2 BLK 2X150CM PCT X36 UDI-DI 20884521086668 Lots D1M0173RY D1M0635RY D2E1210RY Model VS872 SOFSILK* 3-0 BLK 75CM CV23 X36 UDI-DI 20884521090184 Lot D1M1837RY Model VS872-2 SOFSILK 3-0 30 BK CV-23 VS872-2 UDI-DI 20884521758060 Lot D2B1783RY Model SS-5641 SOFSILK* 4-0 BLK 45CM P13 X36 UDI-DI 20884521086774 Lot D2B2181FRY Model SS-5649G SOFSILK* 6-0 BLK 45CM P13 X12 UDI-DI 20884521086804 Lot D2C1987FRY Model SS-645 SOFSILK* 4-0 BLK 75CM C12 X36 UDI-DI 20884521087580 Lot D1M2531FRY Model SS-675 SOFSILK* 3-0 BLK 75CM C16 X36 UDI-DI 20884521087726 Lot D1K1916FRY Model SS-681 SOFSILK* 6-0 BLK 45CM C12 X36 UDI-DI 20884521087924 Lot D2G0795FRY Model VS-872 SOFSILK* 3-0 BLK 75CM CV23 X36 UDI-DI 20884521090184 Lot D1M1837RY Model XX-5280 SOFSILK* 0 BLK 150CM PCT X36 UDI-DI 20884521106045 Lot D2E1482RY -- Ti-Cron Coated Braided Polyester Sutures: 3087-31 TICRON* 4-0 BLU 75CM CV-331DA UDI-DI 10884521632271 Lot D1M2840RY 3226-31 TICRON* 4-0 BLU 60CM CV330 DA UDI-DI 10884521632547 Lot D2G1007RY 88863026-71 TICRON* 1 BLU 75CM HOS12X36 UDI-DI 20884521059129 Lot D2C1802RY 88863070-51 TICRON* 2-0 BLU 75CM C15X36 UDI-DI 20884521059570 Lot D2C3250FRY 88863087-31 TICRON* 4-0 BLU 75CM CV33DA UDI-DI 10884521059832 Lot D1M2840RY 88863092-71 TICRON* 1 BLU 75CM KV37X36 UDI-DI 20884521059990 Lot D2C2450RY 88863147-83 TICRON 2 BLU 5X75CM KV37X12 UDI-DI 10884521060340 Lot D1K0110RY 88863185-41 TICRON* 3-0 BLU 75CM CV305 UDI-DI 20884521061191 Lot D2C0774RY 88863186-31 TICRON* 4-0 BLU 75CM CV331 UDI-DI 10884521061279 Lot D2G1004RY UDI-DI 20884521061276 Lot D2G1004RY 88863226-31 TICRON* 4-0 BLU 60CM CV330DA UDI-DI 20884521061634 Lot D2G1007RY 88863226-41 TICRON* 3-0 BLU 60CM CV330DA UDI-DI 20884521061641 Lot D2A0451RY 88863229-31 TICRON* 4-0 BLU 75CM CV307DA UDI-DI 20884521061689 Lot D2G1307RY 88863271-31 TICRON* 4-0 BLU 90CM CV331DA UDI-DI 10884521062177 Lot D2F2152RY UDI-DI 20884521062174 Lot D2F2152RY 88863309-71 TICRON* 1 BLU 75CM GS11X36 UDI-DI 20884521062488 Lot D2D2188RY -- Monosof Monofilament Nylon Sutures: N-63 MONOSOF* 3-0 BLK 12X45CM PCT X36 UDI-DI 20884521076607 Lot D2A0862RY SN-3695 MONOSOF* 5-0 BLK 45CM P10 X12 UDI-DI 20884521077994 Lot D2C1333FRY D2H1541FRY -- Steel Monofilament Stainless Steel Sutures 88862224-49 STEEL 7 4X45CM SCC X12 UDI-DI 20884521102351 Lot D3B1366RY
Devices named in this recall
55 devices in the FDA device registry whose identifier appears in the recall's code information.
- Sofsilk S-2782KCovidien LP · SUTURE, NONABSORBABLE, SILK3 recalls0 reports
- Sofsilk S-1768KCovidien LP · SUTURE, NONABSORBABLE, SILK2 recalls5 reports
- Ti-Cron 88863070-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls4 reports
- Ti-Cron 88863309-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls4 reports
- Monosof SN-3695Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE2 recalls3 reports
- Sofsilk S-304Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls2 reports
- Sofsilk S-316Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls2 reports
- Sofsilk S-187Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls1 report
- Sofsilk S-605Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls1 report
- Ti-Cron 88863092-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls0 reports
- Ti-Cron 88863185-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls0 reports
- Ti-Cron 88863226-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls0 reports
- Sofsilk CS-211Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-1172Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-1750KCovidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-176Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-182Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-184Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-185Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-243Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-245Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-246Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-303Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-305Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-317Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk S-608Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk SS-5641Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk SS-645Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk VS-872Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk XX-5280Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Sofsilk VS872-2Covidien LP · SUTURE, NONABSORBABLE, SILK2 recalls0 reports
- Ti-Cron 88863147-83Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall4 reports
- Steel 88862224-49Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall3 reports
- Ti-Cron 88863271-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall2 reports
- Sofsilk GS-831Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall1 report
- Sofsilk SS-5649GCovidien LP · SUTURE, NONABSORBABLE, SILK1 recall1 report
- Sofsilk SS-681Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall1 report
- Ti-Cron 88863026-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863087-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863186-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863226-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863229-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Monosof N-63Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall0 reports
- Sofsilk CS-792Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk S-1735Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk S-1789KCovidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk S-2780KCovidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk S-282Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk S-318Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk S-606Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk S-610Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk SS-675Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk SS-1694GCovidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk VS581-2Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk GS453-2Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
Product lines
Other devices under the same FDA submission
77 devices cleared under K981128 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- McKesson S683GXMckesson Medical-Surgical Inc. · Suture, Nonabsorbable, Silk2 recalls0 reports
- Sofsilk GSJ-33-MCovidien LP · SUTURE, NONABSORBABLE, SILK54 reports
- Sofsilk SS-523Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall2 reports
- Sofsilk VS-871Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall2 reports
- Sofsilk GS-823Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall1 report
- Sofsilk GS-835Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall1 report
- Sofsilk S-244Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall1 report
- Sofsilk SS-623Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall1 report
- Sofsilk SS-689Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall1 report
- Sofsilk VS-806Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall1 report
- Sofsilk CS-490Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
- Sofsilk CS-562Covidien LP · SUTURE, NONABSORBABLE, SILK1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93768).
- Z-0844-2024 · Open, ClassifiedSurgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page