Nellcor N65-1Recalls & reported problems
Covidien LP · OXIMAX N-65 Handheld Pulse Oximeter · Class 2
All models of Nellcor (64)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10884521096738), received between Apr 23, 2018 and Jul 3, 2020 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Poor Quality Image” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Poor Quality Image3
- Low Readings2
- Display or Visual Feedback Problem1
- Image Display Error/Artifact1
- Temperature Problem1
Patient problems
- No Patient Involvement6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2020Image Display Error/Artifact; Temperature Problem; Image Display Error/Artifact; Temperature Problem
ACCORDING TO THE REPORTER, THE DEVICE HAD AN ERROR CODE 279 AND A SMOKE WAS OBSERVED. ONE OF THE SPRING WAS LOOSEN AND SHORTED TO THE MAIN BOARD. THEY NEED TO MANUALLY TURN OFF THE DEVICE. NO FLAME BUT A LITTLE BIT OF SMOKE WAS CONFIRMED. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10227695 ↗MalfunctionJun 5, 2020Low Readings; Low Readings
ACCORDING TO THE REPORTER, THE UNIT HAD LOW READING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 10123912 ↗MalfunctionMar 5, 2019Low Readings; Low Readings
ACCORDING TO THE REPORTER, THE UNIT HAD AN INACCURATE READING. THE CUSTOMER REPORTED THAT THE READING IS LOWER THAN NORMAL. IT WAS I NDICATED THAT THERE WAS NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 8390975 ↗MalfunctionNov 30, 2018Display or Visual Feedback Problem; Poor Quality Image; Display or Visual Feedback Problem; Poor Quality Image
ACCORDING TO THE REPORTER, THE DEVICE HAD A DEFECTIVE SCREEN DISPLAY AND INCOMPLETE NUMBERS. IT WAS REPORTED THAT NUMBER NINE WAS DISPLAYED AS FIVE. THE CUSTOMER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8119580 ↗MalfunctionOct 3, 2018Poor Quality Image; Poor Quality Image
ACCORDING TO THE REPORTER, THE UNIT'S LIQUID CRYSTAL DISPLAY(LCD) DISPLAY WAS INCOMPLETE AND THERE WAS A MISSING SEGMENT. THE CUSTOMER REPORTED THAT IT PROVIDED READINGS IN ACTUAL MEASUREMENTS. THE LACKED DISPLAY WAS THE 10 DIGIT PART OF PULSE READING. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 7929499 ↗MalfunctionApr 23, 2018Poor Quality Image; Poor Quality Image
ACCORDING TO THE REPORTER, THE DEVICE HAD MISSING SEGMENTS AT THE CENTER OF THE READINGS. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 7450917 ↗
Other Oximeter devices
- Trilogy Trilogy100 Ventilator, InternationalRespironics, Inc. · Oximeter8 deaths2,493 reports
- Trilogy Trilogy 202 Ventilator, United StatesRespironics, Inc. · Oximeter2 deaths1,594 reports
- HAMILTON-G5 159001Hamilton Medical AG · Oximeter18 deaths1,440 reports
- Trilogy Trilogy100 Ventilator, GermanyRespironics, Inc. · Oximeter696 reports
- Model EMV+ Vent, Army, MRI 799-EMVP-04Zoll Medical Corporation · Oximeter1 death605 reports
- Trilogy Evo, O2, Japan JP2100X16BRespironics, Inc. · Oximeter4 recalls5 deaths342 reports
- Model EMV+ Vent, Army, COMMERCIAL 799-EMVP-01-01Zoll Medical Corporation · Oximeter1 death497 reports
- HAMILTON-G5 159002Hamilton Medical AG · Oximeter3 deaths451 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page