Shiley 6.5PLCF (UDI 10884521103603)Recalls & reported problems
Covidien LP · Pediatric Tracheostomy Tube Long with TaperGuard Cuff · Class 2
All models of Shiley (387)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10884521103603), received between Jul 16, 2021 and Sep 11, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak5
- Use of Device Problem5
- Defective Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Unintended Extubation4
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2025Defective Component; Defective Component
IT WAS REPORTED THAT THE CUFF HAD A HOLE WHEN THE AIR WAS INSERTED DURING TESTING BEFORE INTUBATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23023171 ↗MalfunctionMay 15, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT DURING USE, THE CUFF OF THE TRACHEOSTOMY TUBE HAD AN AIR LEAK. REPLACEMENT OF THE TUBE WAS DONE.
Patient problems: Unintended Extubation
MAUDE report 22030813 ↗MalfunctionNov 19, 2024Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF "BURST" AFTER 48 HOURS OF BEING CHANGED WITH SIMILAR FAILURE. THE TRACH TUBE WAS REPLACED AGAIN.
Patient problems: Unintended Extubation
MAUDE report 20729392 ↗MalfunctionApr 22, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF PRESSURE WAS UNCONTROLLABLE DURING USE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19154742 ↗MalfunctionApr 22, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF PRESSURE WAS UNCONTROLLABLE DURING USE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19154682 ↗MalfunctionDec 19, 2023Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, AFTER A FEW DAYS OF PATIENT USE, THERE WAS A CRACK FOUND IN TRACHEOSTOMY TUBE'S INFLATION LINE NEAR THE FLANGE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18359994 ↗MalfunctionSep 20, 2023Use of Device Problem; Gas/Air Leak; Use of Device Problem; Gas/Air Leak
ACCORDING TO THE REPORTER, AIR ESCAPES FROM THE TRACHEOSTOMY TUBE'S CUFF AFTER A WEEK AND THERE WAS A CRACK NEAR THE FLANGE OF THE INFLATION LINE. THE TUBE WAS REPLACED.
Patient problems: Unintended Extubation
MAUDE report 17786215 ↗MalfunctionAug 16, 2023Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THERE WAS MINIMAL LEAK OBSERVED FROM THE TRACH'S INFLATION LINE, CAUSING ISSUE WITH CUFF INFLATION/DEFLATION OF THE CUFF DURING USE TO A PATIENT. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation
MAUDE report 17561593 ↗MalfunctionFeb 25, 2022Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THERE WAS A CUFF LEAK. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13600211 ↗MalfunctionJul 16, 2021Use of Device Problem; Use of Device Problem
ACCORDING TO REPORTER, DURING PROCEDURE, THE CUFF WOULD NOT INFLATE PRIOR TO USE ON PATIENT. THE DEFECTIVE PRODUCT WAS NOT USE AND WAS REPLACED BY A NEW PRODUCT. THERE WAS NO PATIENT COMPLICATIONS ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12185542 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page