TaperGuard 18770SRecalls & reported problems

Covidien LP · Tracheal Tube with Stylet Murphy Eye · Class 2

Quick summary

This Covidien TaperGuard 18770S is a tracheal tube with a stylet, used to keep a patient's airway open during ventilation. The 39 reports mostly describe air leaking from the cuff or pilot balloon and parts such as connectors coming loose or cracking, sometimes before use and sometimes during intubation, with a few reports of the tube coming out unexpectedly. Four reports involve injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
39
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 10884521103849), received between Mar 24, 2017 and Dec 13, 2024 — classified as 4 injuries, 35 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem14
  • Gas/Air Leak8
  • Defective Component5
  • Disconnection4
  • Inflation Problem4
  • Air Leak3
  • Crack2
  • Deflation Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • No Patient Involvement11
  • Unintended Extubation11
  • Extubate4
  • Insufficient Information3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2024Crack; Crack

    ACCORDING TO THE REPORTER, PRIOR TO USAGE, THERE WAS A CRACK FOUND FROM THE ENDOTRACHEAL TUBE'S CONNECTOR INSERTION POINT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20924307 ↗
  • MalfunctionSep 20, 2024Disconnection; Disconnection

    ACCORDING TO THE REPORTER, THE CONNECTOR OF THE ENDOTRACHEAL TUBE CAME OFF SEVERAL TIMES. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273807 ↗
  • MalfunctionAug 14, 2024Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, DURING USE, THE CUFF OF THE ENDOTRACHEAL TUBE LEAKED AIR AND WAS UNABLE TO INJECT AND REMOVED AIR IN THE PILOT BALLOON. REPLACEMENT OF THE TUBE WAS DONE.

    Patient problems: Unintended Extubation

    MAUDE report 19981270 ↗
  • MalfunctionAug 13, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE ENDOTRACHEAL TUBE'S CUFF DEFLATED WHEN AIR WAS INJECTED DURING PRE-USAGE INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19972810 ↗
  • MalfunctionJun 25, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE ENDOTRACHEAL TUBE'S PILOT BALLOON HAD INFLATION OR DEFLATION ISSUE DURING THE PROCEDURE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19605276 ↗
  • MalfunctionMay 20, 2024Use of Device Problem; Gas/Air Leak; Use of Device Problem; Gas/Air Leak

    ACCORDING TO THE REPORTER, AFTER TWO DAYS OF PATIENT INTUBATION, IT WAS OBSERVED THAT THERE WAS A VOICE LEAK, AND THE ENDOTRACHEAL TUBE'S CUFF PRESSURE WAS INCREASED, BUT THE SYMPTOMS DID NOT IMPROVE, AND WHEN THE CUFF PRESSURE WAS FINALLY INCREASED TO 75 CMH2O, THE CUFF "RUPTURED." THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 19355596 ↗
  • MalfunctionNov 22, 2023Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE ENDOTRACHEAL TUBE'S CUFF WAS NOT INFLATING DUE TO LEAK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18194771 ↗
  • MalfunctionOct 2, 2023Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, AFTER INTUBATION, THE CUFF PRESSURE WAS MEASURED BUT THE ENDOTRACHEAL TUBE'S INFLATION LINE GOT DAMAGED, UNABLE TO CHECK PRESSURE. THE PATIENT HAD A REPLACEMENT OF THE TUBE.

    Patient problems: Unintended Extubation

    MAUDE report 17857291 ↗
  • MalfunctionMar 27, 2023Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE'S TUBE TIP WAS FOUND DEFORMED OR CRUSHED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16622247 ↗
  • MalfunctionMar 24, 2023Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICE'S CUFF RUPTURED DURING PROCEDURE. THERE WAS NO PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 16610954 ↗
  • MalfunctionFeb 23, 2023Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE WAS BENT AND THE SUCTION CATHETER COULD NOT BE INSERTED. REPLACEMENT OF THE TUBE WAS DONE.

    Patient problems: Unintended Extubation

    MAUDE report 16428991 ↗
  • MalfunctionNov 7, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE CUFF WAS TESTED PRIOR TO USE AND SUCCESSFULLY INFLATED AND DEFLATED USING A SYRINGE WITH 20 CMH2O PRE SSURE. THE DEVICE WAS USED BY THE PATIENT BUT, WHEN EXTUBATED, THE CUFF WAS FOUND NOT CONTRACTING. THE TUBE REPLACEMENT WAS NO LONGER REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation

    MAUDE report 15745773 ↗
  • MalfunctionOct 5, 2022Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE DEVICE WAS CONNECTED TO AN AUTOMATIC CUFF PRESSURE GAUGE AND USED, BUT AN ALARM OCCURRED. THE WATER SEEMED TO BE COMING OUT WHEN THE CUFF WAS DE-AERATED. THE PHYSICIAN TRIED TO INFLATE THE CUFF, BUT IT WOULD NOT INFLATE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation

    MAUDE report 15546572 ↗
  • MalfunctionAug 5, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE CUFF WAS PRE-TESTED WITH AIR USING A HAND CUFF PRESSURE GAUGE BUT THE CUFF DID NOT INFLATE SMOOTHLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15177611 ↗
  • MalfunctionJul 25, 2022Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE PATIENT WAS INTUBATED AFTER CHECKING THE CUFF, BUT IT WAS FOUND THAT A LEAK OCCURRED AFTER THE INTUBATION. DEALT WITH ANOTHER TUBE. THERE WAS NO PATIENT HARM.

    Patient problems: Unintended Extubation

    MAUDE report 15097681 ↗
  • MalfunctionJun 1, 2022Crack; Disconnection; Crack; Disconnection

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE CONNECTOR DISCONNECTED WHEN THE PRODUCT WAS OPENED. THERE WAS A CRACK ON THE TUBE OF THE SLIP JOINT PART. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14568983 ↗
  • MalfunctionMay 25, 2022Defective Device; Defective Device

    ACCORDING TO THE REPORTER, NORMALLY, THE STYLET SHOULD HAVE BEEN INSERTED FROM THE 15 MM CONNECTOR SIDE, BUT IT WAS FOUND TO BE INSERTED FROM THE TUBE TIP SIDE. THERE WAS NO PROBLEM WITH THE TUBE ITSELF, SO IT WAS USED. THEY CHECKED THE IN-HOSPITAL INVENTORY TO SEE WHETHER OR NOT THERE WAS A SIMILAR ONE, BUT THERE WAS NO SIMILAR ONE. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 14498673 ↗
  • MalfunctionMay 4, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICE'S CUFF HAD INFLATION/DEFLATION ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14273244 ↗
  • MalfunctionFeb 28, 2022Disconnection; Disconnection

    ACCORDING TO THE REPORTER, DURING USE, THE 15MM CONNECTOR PIECE DETACHED FROM THE TUBE WHILE THE PATIENT WAS INTUBATED. THE PATIENT WAS REINTUBATED.

    Patient problems: Unintended Extubation

    MAUDE report 13620593 ↗
  • MalfunctionFeb 9, 2022Disconnection; Insufficient Device Problem Information; Disconnection; Insufficient Device Problem Information

    ACCORDING TO THE REPORTER, DURING USE, THE 15MM CONNECTOR PIECE DETACHED FROM THE TUBE WHILE THE PATIENT WAS INTUBATED. THE PATIENT WAS REINTUBATED.

    Patient problems: Unintended Extubation

    MAUDE report 13502405 ↗
  • MalfunctionDec 14, 2021Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, DURING TESTING, THE CUFF OF THE DEVICE DID NOT INFLATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13006703 ↗
  • MalfunctionOct 22, 2021Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, AFTER INTUBATION, THE CUFF OF THE DEVICE HAD AN AIR LEAKAGE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12682887 ↗
  • MalfunctionSep 15, 2021Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A KINKED TUBE. THERE WAS NO PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 12475384 ↗
  • MalfunctionJul 1, 2021Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE HAD A CUFF INFLATION DEFLATION ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12101977 ↗
  • InjuryNov 23, 2020Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICE'S CUFF DID NOT INFLATE. THE PATIENT WAS REINTUBATED.

    Patient problems: Unintended Extubation

    MAUDE report 10886571 ↗

Other Tube, tracheal (w/wo connector) devices