Skip to content

Palindrome 8888145046CPRecalls & reported problems

Mozarc Medical US LLC · Precision H Chronic Catheter Kit,Symmetrical Tip, Heparin Coating and Tal VenaTrac Stylet,14.5 Fr/Ch (4.8 mm) x 33 cm · Class 2 · Implant

All models of Palindrome (133)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521158122), received between Nov 30, 2020 and Feb 17, 2022 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Frayed” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

48 other recalls cover devices cleared under the same FDA submission (K123196); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 8, 2022Class IOpen, Classified

    There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

    Z-1313-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Frayed1
  • Material Puncture/Hole1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Unintended Radiation Exposure1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 17, 2022Material Puncture/Hole; Material Puncture/Hole

    ACCORDING TO THE REPORTER, THE PATIENT WAS IN DIALYSIS CENTER FOR TREATMENT, AFTER TREATMENT, THEY TRIED TO FLUSH HOWEVER, THE CATHETER HAD LEAKED THE FLUSHING FLUID AT THE BLUE EXTENSION NEAR THE ADAPTER, AND NO OTHER DEFECTS/DAMAGES FOUND ON THE PRODUCT WHERE THE LEAK WAS OBSERVED. THE CATHETER WAS EXPLANTED WITHOUT PATIENT COMPLICATION, THE CATHETER WAS NOT REPLACED BECAUSE FURTHER TREATMENT WAS NOT NECESSARY AND THEY SENT THE PATIENT BACK TO HOSPITAL. THE CLAMPS MOVED TO A NEW LOCATION AFTER EACH TREATMENT. OCTENISEPT WAS THE CLEANING AGENT USED ON THE DEVICE, AND TYPICALLY UTILIZED TO CLEAN THE ADAPTERS. THE CLEANING AGENT WAS ALLOWED TO DRY THOROUGHLY PRIOR TO APPLYING OINTMENT TO THE AREA, AND NO OTHER PRODUCTS BEING UTILIZED WITH THE DEVICE. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT, NO BLOOD LOSS, NO INFECTION AND NO AIR EMBOLISM. NO DIAGNOSTIC PROCEDURE PERFORMED, NO MEDICATION GIVEN, AND NO SURGICAL PROCEDURES REQUIRED DUE TO EVENT. THE CATHETER WAS NOT REPAIRED, NO TEGO UTILIZED, NO LUER ADAPTER ISSUE AND NOTHING UNUSUAL OBSERVED ON THE DEVICE PRIOR TO USE. THE PATIENT FINAL OUTCOME WAS NORMAL. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13553094 ↗
  • InjuryNov 30, 2020Material Frayed; Product Quality Problem; Material Frayed; Product Quality Problem

    ACCORDING TO THE REPORTER, DURING INSERTION, THE STYLET OF THE FIRST CATHETER COULD NOT BE PLACED INTO THE CATHETERS' LUMEN AND THE GUIDE WIRE COULD NOT BE INSERTED INTO THE CATHETER. IT WAS THEN REPLACED BY A SECOND KIT (WITH THE SAME LOT NUMBER), THEY TRIED TO INSERT THE GUIDE WIRE AND STYLET OF THE SECOND KIT BUT HAD THE SAME ISSUE. THERE WAS NO LEAK AND TEGO WAS NOT UTILIZED. STERILE SODIUM CHLORIDE (NACL) WAS USED AS CLEANING AGENT ON THE DEVICE AND THERE WAS NO LUER ADAPTER ISSUE. THE INSERTION SITE WAS TREATED WITH BRAUNOL PRIOR TO PRODUCT PLACEMENT. THERE WAS NO DAMAGE TO THE DEVICES' INDIVIDUAL EXTERNAL PACKAGING (STILL SEALED) AND TWO CATHETERS CAME WITH BOXES WHEN IT WAS RECEIVED BY THE CUSTOMER. THERE WAS NOTHING UNUSUAL OBSERVED ON THE DEVICES PRIOR TO USE AND FLUSHING WAS DONE WITHOUT PROBLEM. THERE WERE NO OTHER DEFECTS/DAMAGES FOUND ON THE PRODUCTS. THE TUNNELER AND GUIDE WIRES USED WAS THE ONE INCLUDED IN THE KITS. THERE WAS BLOOD LOSS OF 10 ML AND BLOOD TRANSFUSION WAS NOT REQUIRED. THERE WERE NO INTERVENTION/TREATMENT REQUIRED AS A RESULT OF THE EVENT. WHEN THE GUIDE WIRE WAS REMOVED, THE GUIDE WIRE WAS NOT INTACT (IT WAS DAMAGED). AN X-RAY MACHINE WAS USED TO CH

    Patient problems: Unintended Radiation Exposure

    MAUDE report 10918980 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123196. One submission often covers many models, so check the product and lot details in each record.

  • Z-1887-2025 · Apr 23, 2025 · Open, Classified
    Palindrome Precision SI Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1885-2025 · Apr 23, 2025 · Open, Classified
    Palindrome Precision Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1889-2025 · Apr 23, 2025 · Open, Classified
    MAHURKAR Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1886-2025 · Apr 23, 2025 · Open, Classified
    Palindrome Precision HSI Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1888-2025 · Apr 23, 2025 · Open, Classified
    Palindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1635-2024 · Mar 15, 2024 · Open, Classified
    Palindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin Coating and Tal VenaTrac Stylet- Intended for acute and chronic hemodialysis, aph
  • Z-1634-2024 · Mar 15, 2024 · Open, Classified
    Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip, Heparin Coating, Silver Ion Sleeve and Tal VenaTrac Stylet- Intended for acute a
  • Z-1312-2022 · Jun 8, 2022 · Open, Classified
    Palindrome Precision Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
  • Z-1314-2022 · Jun 8, 2022 · Open, Classified
    Palindrome Precision SI Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
  • Z-1315-2022 · Jun 8, 2022 · Open, Classified
    Palindrome Precision HSI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
  • Z-2374-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 44 cm, Item Code 8888541044P
  • Z-2371-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 23 cm, Item Code 8888541023P
  • Z-2372-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 28 cm, Item Code 8888541028P
  • Z-2375-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 55 cm, Item Code 8888541055P
  • Z-2373-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 33 cm, Item Code 8888541033P

…and 33 more.

Other Catheter, hemodialysis, implanted, coated devices

All Catheter, hemodialysis, implanted, coated devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page