Palindrome recallZ-1313-2022 · Covidien Llc · initiated Jun 8, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
FDA root cause category: Process control
What the company is doing
Medtronic distributed a recall notification to customers by mail on 06/08/2022. Customers were instructed to immediately quarantine and discontinue use for specific lots of chronic hemodialysis catheters, which are to be returned to Medtronic. If customers have further distributed product, they are asked to forward the information from the recall notification to those recipients. All customers are asked to complete and return the Consignee Product Retrieval Confirmation Form to [email protected]. Additionally, consignees are asked to share the notice within their organization and retain it for their records regarding the retrieval of unused product. If a consignee is aware of any adverse events associated with recalled product, they are to submit a report to FDA via Medwatch. Customers with any questions can contact their Medtronic Representative or call Customer Service at 800-962-9888, option 2. ***Updated 8/16/22*** Medtronic has created a website to help customers identify affected product: https://www.medtronic.com/covidien/en-us/products/c/renal-care-recall-lookup-tool.html#lotLookupTool
Recall details
- Recall number
- Z-1313-2022
- Status
- Open, Classified
- Initiated by the company
- Jun 8, 2022
- Posted by the FDA
- Jul 7, 2022
- Recalling firm
- Covidien Llc
- Quantity
- 26,675 units
- Product code
- MSD
- FDA submissions
- K123196
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, DC, GU, & PR. The countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Republic of Korea, Kosovo, Kuwait, Lebanon, Luxembourg, Macao, Malaysia, Maldives, Malta, Mayotte, Mexico, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Viet Nam, Virgin Islands, and Zimbabwe.
Product description
Palindrome Precision H Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
Code information (models, lots, serial numbers)
Show the full code information (7k characters)
Code Information organized as "Model No., Serial No., GTIN": 8888119365P, 2119400168, 10884521157712; 8888119365P, 1923400245, 20884521157719; 8888119365P, 1923400268, 20884521157719; 8888119365P, 1935100114, 20884521157719; 8888119365P, 2002100145, 20884521157719; 8888119365P, 2002900148, 20884521157719; 8888119365P, 2020400090, 20884521157719; 8888119365P, 2023300173, 20884521157719; 8888119365P, 2026000162, 20884521157719; 8888119365P, 2031800126, 20884521157719; 8888119365P, 2034400186, 20884521157719; 8888119365P, 2104600103, 20884521157719; 8888119365P, 2104600233, 20884521157719; 8888119365P, 2108100057, 20884521157719; 8888119365P, 2116500094, 20884521157719; 8888119365P, 2116500095, 20884521157719; 8888119365P, 2119400168, 20884521157719; 8888119371P, 1935100115, 10884521157750; 8888123405P, 2002100146, 10884521157781; 8888123405P, 2014900155, 10884521157781; 8888123405P, 2104600162, 10884521157781; 8888123405P, 1924000136, 20884521157788; 8888123405P, 1924000157, 20884521157788; 8888123405P, 1927300064, 20884521157788; 8888123405P, 1928800238, 20884521157788; 8888123405P, 2002100146, 20884521157788; 8888123405P, 2014900155, 20884521157788; 8888123405P, 2023300174, 20884521157788; 8888123405P, 2026000173, 20884521157788; 8888123405P, 2028200162, 20884521157788; 8888123405P, 2104600162, 20884521157788; 8888123405P, 2106900134, 20884521157788; 8888123405P, 2115800062, 20884521157788; 8888123405P, 2115800093, 20884521157788; 8888123405P, 2119400133, 20884521157788; 8888123411P, 2119400128, 20884521157825; 8888128455P, 2002100148, 10884521157859; 8888128455P, 1924000181, 20884521157856; 8888128455P, 1927300066, 20884521157856; 8888128455P, 2002100148, 20884521157856; 8888128455P, 2104600164, 20884521157856; 8888128455P, 2127200433, 20884521157856; 8888128461P, 2012100186, 20884521157894; 8888133505P, 1924600127, 20884521157924; 8888145043CP, 1927500054, 10884521158061; 8888145043CP, 2021000087, 10884521158061; 8888145043CP, 2026000094, 10884521158061; 8888145043CP, 2026000166, 10884521158061; 8888145043CP, 2120000069, 10884521158061; 8888145043CP, 1927500053, 20884521158068; 8888145043CP, 1927500054, 20884521158068; 8888145043CP, 2021000087, 20884521158068; 8888145043CP, 2026000093, 20884521158068; 8888145043CP, 2026000094, 20884521158068; 8888145043CP, 2026000166, 20884521158068; 8888145043CP, 2028200155, 20884521158068; 8888145043CP, 2115800061, 20884521158068; 8888145043CP, 2120000069, 20884521158068; 8888145043P, 1917200199, 10884521158078; 8888145043P, 1917200200, 10884521158078; 8888145043P, 1917200201, 10884521158078; 8888145043P, 1917200280, 10884521158078; 8888145043P, 1917200281, 10884521158078; 8888145043P, 1917600109, 10884521158078; 8888145043P, 1917600110, 10884521158078; 8888145043P, 2028200184, 10884521158078; 8888145043P, 2028200227, 10884521158078; 8888145043P, 1917200199, 20884521158075; 8888145043P, 1917200200, 20884521158075; 8888145043P, 1917200201, 20884521158075; 8888145043P, 1917200280, 20884521158075; 8888145043P, 1917200281, 20884521158075; 8888145043P, 1917600109, 20884521158075; 8888145043P, 1917600110, 20884521158075; 8888145043P, 1921300033, 20884521158075; 8888145043P, 1921900180, 20884521158075; 8888145043P, 2028200184, 20884521158075; 8888145043P, 2028200227, 20884521158075; 8888145044CP, 1907800083, 10884521158085; 8888145044CP, 1927500061, 10884521158085; 8888145044CP, 1927500062, 10884521158085; 8888145044CP, 2023300154, 10884521158085; 8888145044CP, 2026000123, 10884521158085; 8888145044CP, 2026000124, 10884521158085; 8888145044CP, 2026000125, 10884521158085; 8888145044CP, 2026000169, 10884521158085; 8888145044CP, 2119400129, 10884521158085; 8888145044CP, 2119400144, 10884521158085; 8888145044CP, 2125800093, 10884521158085; 8888145044CP, 1907800083, 20884521158082; 8888145044CP, 1927500061, 20884521158082; 8888145044CP, 1927500062, 20884521158082; 8888145044CP, 1929400083, 20884521158082; 8888145044CP, 2014900162, 20884521158082; 8888145044CP, 2023300154, 20884521158082; 8888145044CP, 2026000123, 20884521158082; 8888145044CP, 2026000124, 20884521158082; 8888145044CP, 2026000125, 20884521158082; 8888145044CP, 2026000132, 20884521158082; 8888145044CP, 2026000169, 20884521158082; 8888145044CP, 2119400129, 20884521158082; 8888145044CP, 2119400144, 20884521158082; 8888145044CP, 2125800093, 20884521158082; 8888145044P, 1917600112, 10884521158092; 8888145044P, 1917600113, 10884521158092; 8888145044P, 1917600114, 10884521158092; 8888145044P, 1917600115, 10884521158092; 8888145044P, 1917600135, 10884521158092; 8888145044P, 2023300137, 10884521158092; 8888145044P, 2026000174, 10884521158092; 8888145044P, 2028200164, 10884521158092; 8888145044P, 2028200187, 10884521158092; 8888145044P, 2106900214, 10884521158092; 8888145044P, 2106900227, 10884521158092; 8888145044P, 2124500093, 10884521158092; 8888145044P, 1917600112, 20884521158099; 8888145044P, 1917600113, 20884521158099; 8888145044P, 1917600114, 20884521158099; 8888145044P, 1917600115, 20884521158099; 8888145044P, 1917600135, 20884521158099; 8888145044P, 2023300137, 20884521158099; 8888145044P, 2026000174, 20884521158099; 8888145044P, 2028200164, 20884521158099; 8888145044P, 2028200187, 20884521158099; 8888145044P, 2106900214, 20884521158099; 8888145044P, 2106900227, 20884521158099; 8888145044P, 2124500093, 20884521158099; 8888145045CP, 1929400084, 10884521158108; 8888145045CP, 1929400085, 10884521158108; 8888145045CP, 2026000133, 10884521158108; 8888145045CP, 1929400084, 20884521158105; 8888145045CP, 1929400085, 20884521158105; 8888145045CP, 2026000133, 20884521158105; 8888145045CP, 2028200166, 20884521158105; 8888145045CP, 2119400136, 20884521158105; 8888145045CP, 2119400195, 20884521158105; 8888145045P, 1917200197, 10884521158115; 8888145045P, 1917200198, 10884521158115; 8888145045P, 2028200169, 10884521158115; 8888145045P, 2115800087, 10884521158115; 8888145045P, 2115800088, 10884521158115; 8888145045P, 1917200197, 20884521158112; 8888145045P, 1917200198, 20884521158112; 8888145045P, 2028200169, 20884521158112; 8888145045P, 2115800087, 20884521158112; 8888145045P, 2115800088, 20884521158112; 8888145046CP, 1924600100, 10884521158122; 8888145046CP, 1924600100, 20884521158129; 8888145046CP, 1926100280, 20884521158129; 8888145046CP, 1926800384, 20884521158129; 8888145046CP, 1929400094, 20884521158129; 8888145046CP, 2026000139, 20884521158129; 8888145046CP, 2028200167, 20884521158129; 8888145046P, 1916500094, 10884521158139; 8888145046P, 1916500095, 10884521158139; 8888145046P, 1920500226, 10884521158139; 8888145046P, 1921300030, 10884521158139; 8888145046P, 1916500094, 20884521158136; 8888145046P, 1916500095, 20884521158136; 8888145046P, 1921300030, 20884521158136; 8888145068P, 1934400183, 20884521158426; 8888145069P, 2104600231, 20884521158433; 8888145070P, 2026000172, 20884521158440; 8888145071P, 1922100139, 20884521158457
Devices named in this recall
19 devices in the FDA device registry whose identifier appears in the recall's code information.
- Palindrome 8888145044CPMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls2 deaths30 reports
- Palindrome 8888119365PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls2 reports
- Palindrome 8888123405PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls2 reports
- Palindrome 8888133505PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls1 report
- Palindrome 8888123411PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls0 reports
- Palindrome 8888128455PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls0 reports
- Palindrome 8888128461PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls0 reports
- Palindrome 8888119371PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls0 reports
- Palindrome 8888145044PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated1 recall8 reports
- Palindrome 8888145043CPMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated1 recall5 reports
- Palindrome 8888145045CPMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated1 recall2 reports
- Palindrome 8888145046CPMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated1 recall2 reports
- Palindrome 8888145043PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated1 recall1 report
- Palindrome 8888145045PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated1 recall1 report
- Palindrome 8888145046PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated1 recall1 report
- Palindrome 8888145068PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall0 reports
- Palindrome 8888145069PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall0 reports
- Palindrome 8888145070PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall0 reports
- Palindrome 8888145071PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall0 reports
Product lines
Other devices under the same FDA submission
51 devices cleared under K123196 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123409PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death12 reports
- Palindrome 8888145014PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall62 reports
- Palindrome 8888123410PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls9 reports
- Palindrome 8888145040PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall1 death58 reports
- Palindrome 8888123400PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls5 reports
- Palindrome 8888128454PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls5 reports
- Palindrome 8888133504PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls3 reports
- Palindrome 8888119370PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 report
- Palindrome 8888128459PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 90311).
- Z-1312-2022 · Open, ClassifiedPalindrome Precision Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
- Z-1314-2022 · Open, ClassifiedPalindrome Precision SI Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
- Z-1315-2022 · Open, ClassifiedPalindrome Precision HSI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
- Z-1316-2022 · Open, ClassifiedPalindrome Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
- Z-1317-2022 · Open, ClassifiedPalindrome H Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
- Z-1318-2022 · Open, ClassifiedPalindrome SI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
- Z-1319-2022 · Open, ClassifiedPalindrome HSI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
- Z-1320-2022 · Open, ClassifiedMahurkar Chronic Carbothane (Maxid) Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page