Skip to content

TaperGuard 18765SRecalls & reported problems

Covidien LP · Tracheal Tube with Stylet Murphy Eye · Class 2

All models of TaperGuard (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521164543), received between Feb 4, 2020 and Aug 5, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Disconnection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Disconnection1
  • Gas/Air Leak1

Patient problems

  • Extubate1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT PRIOR TO USAGE, THE ENDOTRACHEAL TUBE'S CONNECTOR PART WAS LOOSE AND CAN BE DETACHED EASILY. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22717122 ↗
  • InjuryFeb 4, 2020Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, DURING USE, THE DEVICE HAD LEAK. EXTUBATION AND RE-INTUBATION WERE PERFORMED AND THE ISSUE WAS RESOLVED. THE PRODUCT WAS CHECKED AFTER EXTUBATION, THE TUBE TIP WAS CLOSED AND WAS PLACED IN WATER. WHEN THE PRESSURE WAS APPLIED FROM THE SLIP JOINT PART, BUBBLES CAME OUT FROM THE ROOT PART OF THE INFLATION LINE. THE HOLE PART THAT WAS OPENED TO ATTACH THE INFLATION LINE TO THE TUBE WAS BEING PENETRATED.

    Patient problems: Extubate

    MAUDE report 9667399 ↗

Other Tube, tracheal (w/wo connector) devices

All Tube, tracheal (w/wo connector) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page