Oridion N85Recalls & reported problems
Covidien LP · Portable Bedside Capnograph/Pulse Oximeter · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884521172043), received between Nov 21, 2018 and Nov 21, 2024 — classified as 1 death, 4 malfunctions.
The most frequently reported specific device problem is “Device Damaged Prior to Use” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Device Damaged Prior to Use1
- Failure to Calibrate1
- High Readings1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Material Split, Cut or Torn1
- Temperature Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Death1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2024Failure to Calibrate; High Readings; Failure to Calibrate; High Readings
ACCORDING TO THE REPORTER, THE DEVICE WAS UNABLE TO CALIBRATE AND GAVE A HIGH READINGS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20754167 ↗MalfunctionAug 10, 2022Temperature Problem; Temperature Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE UNIT BECAME HOT WHEN POWERED ON. THE DEVICE WAS POWER CYCLED FOR TROUBLESHOOTING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15201567 ↗MalfunctionJan 14, 2022Device Damaged Prior to Use; Material Split, Cut or Torn; Device Damaged Prior to Use; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, PRIOR TO USE, THE UNIT HAD A CRACK CASE NEAR TO POWER SWITCH AND PSU CABLE LOW VOLTAGE SIDE LOCKING TAB WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13260611 ↗DeathMar 13, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, THE DEVICE WAS CAUGHT ON FIRE WHICH ALLEGEDLY CAUSED BY THE DEFIBRILLATOR. IT WAS REPORTED THAT THE DEVICE WAS NOT POWERED ON AND WAS NOT CONNECTED TO THE PATIENT. THE DEVICE WAS BURNED/MELTED DUE TO FIRE FROM ELECTRICAL ENERGY ARC FROM DEFIBRILLATOR PAD AND HIGH FLOW OXYGEN FROM NASAL CANNULA IN PATIENT'S BED. THE CUSTOMER REPORTED THAT THE PATIENT HAD BURNS TO HEAD, NECK, FACE, EARS AND SHOULDERS DUE TO THE FIRE AND DIED DUE TO CARDIAC ARREST. THE CAUSE OF DEATH WAS NOT RELATED TO DEVICE.
Patient problems: Death
MAUDE report 8416857 ↗MalfunctionNov 21, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ACCORDING TO THE REPORTER, THE UNIT WAS NOT REALLY USED ON A PATIENT. THERE WAS A CODE BLUE ON A PATIENT, THEY WERE GOING TO USE THIS UNIT ON THE PATIENT TO CHECK SPO2 AND CO2. BUT WHEN THE UNIT WAS TURNED ON, IT WAS RED AND DID NOT GIVE ANY KIND OF READINGS AT ALL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8095641 ↗
Other Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase devices
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 deaths8,444 reports
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall607 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE485 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE471 reports
- NA GF-210RNihon Kohden Corporation · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE386 reports
- Trilogy IT2110X21BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls30 reports
- Trilogy Trilogy Evo Ventilator, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls26 reports
- Trilogy Evo Trilogy Evo Ventilator with oxygen blending model, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page