Nellcor N600X-1Recalls & reported problems
Covidien LP · OXIMAX N-600x Pulse Oximeter · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521181229), received between Oct 4, 2017 and Jun 4, 2021 — classified as 1 death, 3 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K123581); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Alarm System2
- High Readings1
- Temperature Problem1
Patient problems
- Bronchopneumonia1
- Diminished Pulse Pressure1
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 4, 2021Device Alarm System; Device Alarm System
ACCORDING TO THE REPORTER, POSTOPERATIVELY, THE UNIT DID NOT ALARM. THIS WAS USED ON A PATIENT PARALYZED FROM NECK DOWN, WHO WAS ON A 24/7 VENTILATOR WITH FITTED PACEMAKER AND WAS FED USING A PEG TUBE. THE PATIENT DIED BECAUSE OF PNEUMONIA, WHICH WAS NOT ATTRIBUTED TO THE DEVICE USED.
Patient problems: Bronchopneumonia
MAUDE report 11940049 ↗DeathJan 12, 2021High Readings; High Readings
DURING HOME MONITORING USE THE N600X-1 SPO2 PULSE RATE ALARM DID NOT GO OFF AT THE TIME OF PATIENT¿S DEATH. THE PATIENT WAS PARALYZED FROM THE NECK DOWN, VENTILATOR DEPENDENT WITH A PACEMAKER AND PEG TUBE FOR FEEDINGS. THE PACEMAKER WAS 10 YEARS OLD. THE PATIENT CAREGIVERS INFORMED THE POLICE THAT THE DEVICE DID NOT ALARM AT THE TIME OF DEATH, AND THAT ANOTHER MONITOR (MINDRAY) WAS BEING USED AT THE SAME TIME TO MONITOR THE HEARTBEAT WHICH WAS DECREASING. IT WAS REPORTED THAT THE N600X-1 WAS READING 60BPM, WHICH WAS WHAT THE PACEMAKER WAS SET TO. FOLLOWING THE EVENT, NHS TRUST PERFORMED 24-HOUR TESTING OF THE N600X-1 AND SIMULATED THE REDUCTION OF PULSE AND SATURATION AND THE DEVICE ALARMED AS NORMAL. PER THE REPORTER THE PATIENTS CAUSE OF DEATH WAS PNEUMONIA. THE DEVICE IT NOT RELATED TO THE CAUSE OF DEATH.
Patient problems: Diminished Pulse Pressure
MAUDE report 11157932 ↗MalfunctionMay 21, 2020Temperature Problem; Temperature Problem
ACCORDING TO THE REPORTER, THE DEVICE WAS PLUGGED AND THE DEVICE SHORTED OUT. A BURNING SMELL WAS NOTED. THERE WAS NO ALLEGATION OF PATIENT DEATH OR SERIOUS INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10079542 ↗MalfunctionOct 4, 2017Device Alarm System; Device Alarm System
MEDTRONIC RECEIVED A REPORT STATING A N600X-1 PULSE OXIMETER FAILED TO ALARM. THERE WAS NO PATIENT HARM REPORTED WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6914353 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123581. One submission often covers many models, so check the product and lot details in each record.
- Z-2268-2015 · Jun 29, 2015 · TerminatedOxiMax N-560 Pulse Oximeter. For continuous or spot check monitoring of functional arterial oxygen saturation (SpO2) and pulse rate.
- Z-2267-2015 · Jun 29, 2015 · TerminatedOxiMax N-65 Handheld Pulse Oximeter. N65, N65-1, N65P, and N65P-1. For continuous or spot check monitoring of functional arterial oxygen saturation (SpO2) and pulse rate.
Other Oximeter devices
- Trilogy Trilogy100 Ventilator, InternationalRespironics, Inc. · Oximeter8 deaths2,493 reports
- Trilogy Trilogy 202 Ventilator, United StatesRespironics, Inc. · Oximeter2 deaths1,594 reports
- HAMILTON-G5 159001Hamilton Medical AG · Oximeter18 deaths1,440 reports
- Trilogy Trilogy100 Ventilator, GermanyRespironics, Inc. · Oximeter696 reports
- Model EMV+ Vent, Army, MRI 799-EMVP-04Zoll Medical Corporation · Oximeter1 death605 reports
- Trilogy Evo, O2, Japan JP2100X16BRespironics, Inc. · Oximeter4 recalls5 deaths342 reports
- Model EMV+ Vent, Army, COMMERCIAL 799-EMVP-01-01Zoll Medical Corporation · Oximeter1 death497 reports
- HAMILTON-G5 159002Hamilton Medical AG · Oximeter3 deaths451 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.