Shiley 86550Recalls & reported problems
Covidien LP · Oral/Nasal Tracheal Tube Cuffed Reinforced, Murphy Eye · Class 2
All models of Shiley (387)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10884521211667), received between Apr 18, 2019 and Feb 4, 2025 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem4
- Gas/Air Leak3
- Inflation Problem2
- Material Split, Cut or Torn2
- Defective Component1
- Insufficient Device Problem Information1
- Material Separation1
Patient problems
- No Patient Involvement4
- Unintended Extubation4
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 4, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT DURING PATIENT'S SEDATION, THE ANESTHESIOLOGIST WAS UNABLE TO VISUALIZE THE ETCO2 MEASUREMENT IN THE ANESTHESIA MACHINE, AND THE EQUIPMENT'S BELLOWS WERE NOT HOLDING, INDICATING A LEAK. THE LEAK WAS DETECTED TO BE COMING FROM THE ENDOTRACHEAL TUBE BALLOON, SPECIFICALLY AT THE PILOT BALLOON SEAM. THE ENDOTRACHEAL TUBE WAS SUBSEQUENTLY CHANGED. THE EVENT TOOK PLACE IN A HOSPITAL SETTING.
Patient problems: Unintended Extubation
MAUDE report 21309622 ↗MalfunctionAug 23, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, DURING THE PATIENT'S SEDATION, THE ANESTHESIOLOGIST WAS UNABLE TO VISUALIZE THE END-TIDAL CARBON DIOXIDE (ETCO2) MEASUREMENT ON THE ANESTHESIA MACHINE AND ALSO DETECTED THAT THE EQUIPMENT'S BELLOWS WERE NOT HOLDING, INDICATING A LEAK. AFTER CHECKING THE EQUIPMENT, IT WAS DETERMINED THAT THE LEAK ORIGINATED FROM THE ENDOTRACHEAL TUBE BALLOON, SO THE DEVICE WAS CHANGED.
Patient problems: Unintended Extubation
MAUDE report 20051734 ↗MalfunctionMar 13, 2024Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE THREE ENDOTRACHEAL TUBE'S PILOT BALLOON BLEW UP UPON CONNECTING A SYRINGE AND DEFLATES THE BALLOON AFTER REMOVING THE SYRINGE. IT WAS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: Insufficient Information
MAUDE report 18899546 ↗MalfunctionMar 13, 2024Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE THREE ENDOTRACHEAL TUBE'S PILOT BALLOON BLEW UP UPON CONNECTING A SYRINGE AND DEFLATES THE BALLOON AFTER REMOVING THE SYRINGE. IT WAS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: Insufficient Information
MAUDE report 18899510 ↗MalfunctionMar 13, 2024Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE THREE ENDOTRACHEAL TUBE'S PILOT BALLOON BLEW UP UPON CONNECTING A SYRINGE AND DEFLATES THE BALLOON AFTER REMOVING THE SYRINGE. IT WAS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: Insufficient Information
MAUDE report 18899499 ↗MalfunctionNov 22, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, THE REINFORCED TUBE'S CUFF BROKE DURING RHINOPLASTY SURGERY AND IT WAS NECESSARY TO CHANGE THE TUBE.
Patient problems: Unintended Extubation
MAUDE report 18194925 ↗MalfunctionNov 22, 2023Insufficient Device Problem Information; Material Split, Cut or Torn; Insufficient Device Problem Information; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, THE REINFORCED TUBE'S CUFF BROKE DURING RHINOPLASTY SURGERY AND IT WAS NECESSARY TO CHANGE THE TUBE.
Patient problems: Unintended Extubation
MAUDE report 18194913 ↗MalfunctionApr 26, 2023Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE DEVICE'S CUFF HAD A LEAK DURING PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16820431 ↗MalfunctionSep 30, 2022Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING PROCEDURE, THE CUFF HAD AN INFLATION/DEFLATION ISSUE. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15520019 ↗MalfunctionNov 4, 2020Defective Component; Defective Component
ACCORDING TO THE REPORTER, THE DEVICE HAD A PHYSICAL DAMAGE WHEREIN THE CANNULA WAS DEFORMED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 10783814 ↗MalfunctionDec 30, 2019Material Separation; Material Separation
ACCORDING TO THE REPORTER, WHEN INFLATING, THE DEVICE BALL WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9530011 ↗MalfunctionJun 17, 2019Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, THE UNIT WAS DIFFICULT TO INFLATE AND TO DEFLATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8705376 ↗MalfunctionApr 18, 2019Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, PRE OPERATIVELY, THE PILOT BALLOON WAS INFLATED WITH SYRINGE BUT COULD NOT INFLATE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8530213 ↗
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page