Microstream PM35MN01Recalls & reported problems
Oridion Medical 1987 Ltd. · MICROSTREAM CAPNOSTREAM 35 PORTABLE RESPIRATORY MONITOR(EURO · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10884521551589), received between Jul 14, 2023 and Mar 19, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement8
- Low Readings4
- Device Alarm System2
- Erratic or Intermittent Display2
- Failure to Calibrate2
- High Readings2
- Display or Visual Feedback Problem1
- Failure to Run on Battery1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- Insufficient Information3
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2026Failure to Run on Battery; High Readings
IT WAS REPORTED THAT PRIOR TO USE, THE BATTERY AVAILABILITY FAILED AND THE ETCO2 VALUE WAS HIGHER THAN EXPECTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24637204 ↗MalfunctionMay 27, 2025Low Readings; Low Readings
IT WAS REPORTED THAT THE DEVICE HAD LOW CO2. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22088344 ↗MalfunctionJan 16, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT, PRIOR TO USE, THE UNIT HAD A CO2 VALUE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21167401 ↗MalfunctionNov 7, 2024Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE'S CO2 PARAMETER WAS INCORRECT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20630859 ↗MalfunctionNov 7, 2024Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, PRIOR TO USE, THE PRODUCT HAD INCORRECT CO2 READINGS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20627409 ↗MalfunctionNov 6, 2024Incorrect Measurement; Failure to Calibrate; Incorrect Measurement; Failure to Calibrate
ACCORDING TO THE REPORTER, PRIOR TO USE, THE CO2 READING OF THE UNIT WAS NOT ACCURATE, AS THE CO2 CALIBRATION FAILED. NOW SUCCESSFULLY PERFORMED THE CO2 CALIBRATION, HENCE IT SHOWED THE ACCURACY OF CO2. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20624777 ↗MalfunctionNov 6, 2024High Readings; High Readings
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE HAD HIGHER ETCO2 THAN EXPECTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20623521 ↗MalfunctionNov 5, 2024Low Readings; Low Readings
ACCORDING TO THE REPORTER, PRIOR TO USE, THE UNIT HAD LOWER END-TIDAL CARBON DIOXIDE (ETCO2). A SIMULATOR WAS USED TO DETERMINE THE READINGS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20615413 ↗MalfunctionOct 7, 2024Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, THE DEVICE DISPLAYED INCORRECT CO VALUE COMPARED TO EXPECTED. THERE WAS NO REPORTED PATIENT INVOLVEMENT OR OUTCOME.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20383622 ↗MalfunctionAug 27, 2024Low Readings; Low Readings
ACCORDING TO THE REPORTER, PRIOR TO USE. THE MONITOR SHOWED A LOWER CO2 READING. AND A PUMP TEST WAS REQUESTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20080014 ↗MalfunctionJul 18, 2024Erratic or Intermittent Display; Incorrect Measurement; Failure to Calibrate; Erratic or Intermittent Display; Incorrect Measurement; Failure to Calibrate
ACCORDING TO THE REPORTER, THE UNIT END TIDAL CARBON DIOXIDE (ETCO2) VALUES WERE INCORRECT AND IT HAD CALIBRATION ISSUE. ANOTHER MACHINE WAS USED TO CONFIRM THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19778666 ↗MalfunctionMar 25, 2024Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, PRIOR TO USE, THE PRODUCT GAVE INCORRECT CO2 READINGS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18973726 ↗MalfunctionMar 12, 2024Device Alarm System; Device Alarm System
ACCORDING TO THE REPORTER, THE ALARM ON THE MONITOR WAS TRIGGERED, BUT THERE WAS NO AUDIO SOUND PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: Insufficient Information
MAUDE report 18889388 ↗MalfunctionDec 20, 2023Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, THE UNIT DID NOT GIVE CORRECT PARAMETERS. THERE WAS NO REPORTED PATIENT INVOLVEMENT AND OUTCOME.
Patient problems: Insufficient Information
MAUDE report 18368249 ↗MalfunctionJul 14, 2023Device Alarm System; Erratic or Intermittent Display; Display or Visual Feedback Problem; Low Readings; Device Alarm System; Erratic or Intermittent Display; Display or Visual Feedback Problem; Low Readings
ACCORDING TO THE REPORTER, THE UNIT HAD ABNORMAL ETCO2 READINGS AS LOW READINGS WITH A DISPLAYED READINGS OF 15 - 16 MMHG COMPARED TO TO 33-35 MMHG. READINGS WERE COMPARED WITH ARTERIAL BLOOD GAS (ABG) AND OTHER CO2 MODULE DEVICE. ALSO, LOW ETCO2 READINGS REFERS TO NOT SHOWING READINGS PROPERLY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17330399 ↗
Other ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE devices
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 deaths8,444 reports
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall607 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE485 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE471 reports
- NA GF-210RNihon Kohden Corporation · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE386 reports
- Trilogy IT2110X21BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls30 reports
- Trilogy Trilogy Evo Ventilator, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls26 reports
- Trilogy Evo Trilogy Evo Ventilator with oxygen blending model, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page