Skip to content

Shiley 86234 (UDI 10884521620254)Recalls & reported problems

Covidien LP · Oral/Nasal Tracheal Tube Cuffless Non DEHP, Murphy Eye · Class 2

Quick summary

This Shiley 86234 is a cuffless oral/nasal tracheal tube made by Covidien. The 49 reports mostly describe the tube or its connector breaking, cracking, or coming apart during use — including a piece blocking the airway so a patient could not be ventilated — plus suction catheters getting stuck inside the tube and printed markings wearing off. Two injuries are reported and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
49
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 10884521620254), received between May 24, 2019 and Dec 20, 2023 — classified as 2 injuries, 47 malfunctions.

The most frequently reported specific device problem is “Defective Component” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Component27
  • Contamination /Decontamination Problem10
  • Obstruction of Flow3
  • Material Deformation2
  • Appropriate Device Problem Term/Code Not Available1
  • Crack1
  • Disconnection1
  • Insufficient Device Problem Information1

Patient problems

  • No Patient Involvement40
  • No Clinical Signs, Symptoms or Conditions4
  • Unintended Extubation2
  • Extubate1
  • Insufficient Information1
  • Low Oxygen Saturation1
  • No Consequences Or Impact To Patient1
  • Swelling1

Recent adverse event reports

The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 20, 2023Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem

    ACCORDING TO THE REPORTER, THE ENDOTRACHEAL TUBE'S MARKINGS WORE OFF DURING USE. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 18363237 ↗
  • MalfunctionApr 9, 2021Obstruction of Flow; Obstruction of Flow

    ACCORDING TO THE REPORTER, THE DEVICE WAS INSERTED TO SUCK PATIENT'S SPUTUM, BUT IT WAS HARD TO PULL OUT AND FOUND STUCK AT THE ENDOTRACHEAL TUBE. THE PATIENT HAD A REPLACEMENT OF THE TUBE WITH NO HARM.

    Patient problems: Unintended Extubation

    MAUDE report 11641822 ↗
  • MalfunctionMar 31, 2021Obstruction of Flow; Insufficient Device Problem Information; Obstruction of Flow; Insufficient Device Problem Information

    ACCORDING TO THE REPORTER, THE DEVICE SPUTUM SUCTION TUBE SIZE SIX WAS INSERTED TO SUCK PATIENT'S SPUTUM BUT IT WAS HARD TO PULL OUT AND THE NURSE FELT IT STUCK. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11596612 ↗
  • MalfunctionMar 31, 2021Obstruction of Flow; Obstruction of Flow

    ACCORDING TO THE REPORTER, THE DEVICE WAS INSERTED TO SUCK PATIENT'S SPUTUM, BUT IT WAS HARD TO PULL OUT AND FOUND STUCK AT THE ENDOTRACHEAL TUBE. THE PATIENT HAD A REPLACEMENT OF THE TUBE WITH NO HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation

    MAUDE report 11596575 ↗
  • MalfunctionMar 15, 2021Crack; Crack

    ACCORDING TO THE REPORTER, THE DEVICE HAD A CRACK AT THE END OF THE TUBE TO 15MM CONNECTOR. IT WAS REPORTED THAT WHEN THE 15MM CONNECTOR WAS GOING TO BE CONNECTED TO THE VENTILATOR, THE CRACK WAS FOUND. THE CRACK PART WAS CUT, THEN THE 15 MM CONNECTOR WAS INSERTED INTO THE TUBE AND CONNECTED WITH THE VENTILATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11484656 ↗
  • InjuryNov 4, 2020Material Integrity Problem; Material Integrity Problem

    ACCORDING TO THE REPORTER, AFTER INTUBATION, ADAPTER ON ETT SNAPPED AND A PIECE WAS OBSTRUCTING THE ETT. THE PATIENT COULD NOT BE VENTILATED AND STARTED TO DECOMPENSATE. THE TUBE WAS CUT AND REPLACED CAP RESULTING IN PATIENT STABILIZING. THE TUBE REMAINED IN THE PATIENT.

    Patient problems: Low Oxygen Saturation

    MAUDE report 10783804 ↗
  • MalfunctionOct 23, 2020Disconnection; Disconnection

    ACCORDING TO THE REPORTER, THE DEVICE HUB BROKE OFF THE STEM. RATHER THAN REINTUBATE THE PATIENT, WAS ABLE TO USE THE BALLARD ADAPTER. THE STEM IS STUCK INSIDE OF THE ET TUBE. THE DEVICE WAS FRESH FROM THE PACKAGED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10727749 ↗
  • MalfunctionFeb 14, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE REPORTER, DURING USED, THE DEVICE CUFF DEHP MATERIAL CAUSED GLOTTIS SWELLING.

    Patient problems: Swelling

    MAUDE report 9712001 ↗
  • InjuryDec 20, 2019Loss of or Failure to Bond; Loss of or Failure to Bond

    ACCORDING TO THE REPORTER, DURING USE, THE DEVICE MARKING WAS RUBBING OFF. THE CUSTOMER INDICATED THAT THE PATIENT HAD A HIGH RISK FOR TRAUMA OR EXTUBATION AND DEVICE MISMANAGEMENT. THE CUSTOMER INDICATED THAT THE INCIDENT DEVICE HAS BEEN DISCARDED.

    Patient problems: Extubate

    MAUDE report 9503513 ↗
  • MalfunctionDec 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE HAD TUBE THAT WAS STICKY AND ROUGH. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 9401439 ↗
  • MalfunctionSep 13, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    ACCORDING TO THE REPORTER, THE UNIT HAD STICKY RESIDUE END OF TUBE. CLOUDY, ROUGH SURFACE THROUGHOUT. THERE IS NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9040399 ↗
  • MalfunctionSep 13, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    ACCORDING TO THE REPORTER, THE UNIT HAD STICKY RESIDUE END OF TUBE. CLOUDY, ROUGH SURFACE THROUGHOUT. THERE IS NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9038745 ↗
  • MalfunctionSep 13, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem

    ACCORDING TO THE REPORTER, THE UNIT HAD STICKY RESIDUE END OF TUBE. CLOUDY, ROUGH SURFACE THROUGHOUT. THERE IS NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9038588 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8957502 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8957501 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8957500 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8956874 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8956863 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8956822 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8956782 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8956709 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8956632 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8956588 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8956554 ↗
  • MalfunctionSep 3, 2019Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HAD A STICKY RESIDUE AT THE END OF THE TUBE. IT WAS STATED THAT IT WAS CLOUDY AND HAD A ROUGH SURFACE ALL THROUGHOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8956515 ↗

Other Tube, Tracheal (W/Wo Connector) devices

All Tube, Tracheal (W/Wo Connector) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.