Shiley 8CN85ERRecalls & reported problems
Covidien LP · Adult Flexible Evac Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula · Class 2
All models of Shiley (387)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10884521726307), received between Jan 31, 2022 and Jun 5, 2026 — classified as 1 injury, 15 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak11
- Use of Device Problem4
- Appropriate Device Problem Term/Code Not Available3
- Defective Component2
- Human-Device Interface Problem1
Patient problems
- Unintended Extubation11
- No Clinical Signs, Symptoms or Conditions4
- Discomfort1
- Low Oxygen Saturation1
- Pain1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2026Use of Device Problem
IT WAS REPORTED THAT PRIOR TO USE ON A PATIENT, THE TRACHEOSTOMY TUBE'S CUFF INTEGRITY TEST WAS PERFORMED AND FAILURE WAS OBSERVED WITH A HOLE AND LEAK IN THE CUFF. ANOTHER UNIT WAS USED AND THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25417448 ↗MalfunctionMar 2, 2026Use of Device Problem; Gas/Air Leak
IT WAS REPORTED THAT INTRA-OPERATIVELY, EVEN THE TRACH TUBE'S CUFF WAS CHECKED BEFORE PLACEMENT, AFTER INTUBATION THE CUFF STARTED TO LOOSE PRESSURE AND FOUND RUPTURED. THE TUBE WAS REPLACED.
Patient problems: Unintended Extubation
MAUDE report 24484889 ↗MalfunctionFeb 17, 2026Gas/Air Leak
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE EXPERIENCED A A LEAKING CUFF DURING USE. THE ISSUE TOOK PLACE IN A HOSPITAL SETTING. THE TUBE WAS REPLACED.
Patient problems: Unintended Extubation
MAUDE report 24372889 ↗MalfunctionFeb 17, 2026Gas/Air Leak
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE EXPERIENCED A A LEAKING CUFF DURING USE. THE ISSUE TOOK PLACE IN A HOSPITAL SETTING. THE TUBE WAS REPLACED.
Patient problems: Unintended Extubation
MAUDE report 24372859 ↗MalfunctionApr 1, 2025Defective Component; Gas/Air Leak; Defective Component; Gas/Air Leak
IT WAS REPORTED THAT A PATIENT WITH A TRACHEOSTOMY IN PLACE EXPERIENCED ISSUES WITH A TRACHEOSTOMY TUBE. THE CUFF OF THE TRACHEOSTOMY TUBE HAD A HOLE AND DEFLATED, LEADING TO A SECOND TRACHEAL LEAK WITHIN LESS THAN 24 HOURS FOR THE SAME PATIENT. THE TUBE WAS REPLACED.
Patient problems: Unintended Extubation
MAUDE report 21738585 ↗MalfunctionJan 3, 2025Gas/Air Leak; Human-Device Interface Problem; Gas/Air Leak; Human-Device Interface Problem
ACCORDING TO THE REPORTER, THE CUFF OF THE TWO TRACHEOSTOMY TUBES DID NOT HOLD PRESSURE AND HAD BEEN CONSISTENTLY LEAKING. THE PRESSURE WAS SET AND MAINTAINED AT A PRESSURE OF 20 TO 30 CMH2O, BUT THE PRESSURE WAS DOWN TO BETWEEN 10 AND 20 WITHIN 24 HOURS, SOMETIMES LESS THAN 8 HOURS. THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Discomfort; Pain
MAUDE report 21064616 ↗MalfunctionNov 13, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, DURING USE, THE TRACHEOSTOMY TUBE'S CUFF WAS FOUND LEAKING. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20674803 ↗MalfunctionSep 10, 2024Defective Component; Gas/Air Leak; Defective Component; Gas/Air Leak
ACCORDING TO THE REPORTER, THE CUFF OF THE TRACHEOSTOMY TUBE COULD NOT SEAL DUE TO A LEAK AFTER ONE TO TWO DAYS OF INSERTION. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20179899 ↗MalfunctionAug 9, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, DURING USE, THERE WERE TRACHEOSTOMY TUBES THAT HAD SECRETIONS LEFT IN THE SUBGLOTTIC LINES. THE REPLACEMENT OF THE TUBE WAS DONE.
Patient problems: Unintended Extubation
MAUDE report 19953570 ↗MalfunctionAug 7, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, DURING USE, THE TRACHEOSTOMY TUBE HAD SECRETION LEFT IN THE SUBGLOTTIC LINES. REPLACEMENT OF THE TUBE WAS DONE.
Patient problems: Unintended Extubation
MAUDE report 19929239 ↗MalfunctionAug 7, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, DURING USE, THERE WERE TRACHEOSTOMY TUBES THAT HAD SECRETION LEFT IN THE SUBGLOTTIC LINES. REPLACEMENT OF THE TUBE WAS DONE.
Patient problems: Unintended Extubation
MAUDE report 19929237 ↗MalfunctionAug 6, 2024Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING INSPECTION, THE TRACHEOSTOMY TUBE'S CUFF WAS DEFLATING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19917124 ↗MalfunctionJun 29, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, DURING USE, THE PATIENT HAD BEEN INTUBATED FOR SEVERAL DAYS BEFORE GOING TO THE SUBGLOTTIC SECRETION UNIT. DURING THE CANNULA CHANGE, THERE WAS A SECRETION LEFT IN THE EVAC CANNULA AND SUCTION LINE AS SHOWN IN THE ATTACHED PHOTO. REPLACEMENT OF THE TUBE WAS DONE.
Patient problems: Unintended Extubation
MAUDE report 19640438 ↗MalfunctionJun 8, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, BOTH TRACHEOSTOMY TUBES' CUFFS HAD A LEAK DURING PATIENT USAGE. THE TUBES WERE REPLACED.
Patient problems: Unintended Extubation
MAUDE report 19493308 ↗InjuryMar 15, 2023Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE PATIENT DESATURATED DUE TO CUFF LEAK DURING THE PROCEDURE. THE TUBE WAS REPLACED.
Patient problems: Low Oxygen Saturation; Unintended Extubation
MAUDE report 16546050 ↗MalfunctionJan 31, 2022Use of Device Problem; Gas/Air Leak; Use of Device Problem; Gas/Air Leak
ACCORDING TO THE REPORTER, DURING ROUTINE CHECK ON THE PATIENT, THE CUFF OF THE DEVICE WAS FOUND DEFLATED. ATTEMPTED TO RE-INFLATE THE CUFF BUT HAD FAILED. THE PATIENT HAD A REPLACEMENT OF THE TUBE. THE EXTUBATED DEVICE HAD CONFIRMED A LEAK ON THE CUFF.
Patient problems: Unintended Extubation
MAUDE report 13392598 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026