Capnostream20p DLCS08651-03Recalls & reported problems

Oridion Medical 1987 Ltd. · Portable Bedside Capnography Demo Loaner · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521771697), received between Jan 4, 2022 and Sep 5, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K123690); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Alarm System2
  • Failure to Power Up2
  • Temperature Problem2
  • Connection Problem1
  • Device Damaged Prior to Use1
  • Display or Visual Feedback Problem1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 5, 2025Temperature Problem; Temperature Problem

    AN INCIDENT WAS REPORTED INVOLVING A UNIT EXPERIENCING INTERNAL COMPONENT ISSUES. THE UNIT EMITTED A SMOKE SMELL, INDICATIVE OF A BURNING INTERNAL COMPONENT, AND WAS FUNCTIONING INTERMITTENTLY. ADDITIONALLY, THE POWER CORD WAS FOUND TO BE DISCONNECTED FROM THE UNIT, FAILING TO ESTABLISH A PROPER CONNECTION. AS PART OF TROUBLESHOOTING, THE UNIT WAS TURNED OFF TO PREVENT FURTHER COMPLICATIONS. NO VISIBLE SMOKE WAS OBSERVED, AND NO POWER SURGE OCCURRED DURING THE EVENT. NO PATIENT WAS INVOLVED IN THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22975340 ↗
  • MalfunctionAug 21, 2025Device Alarm System; Display or Visual Feedback Problem; Failure to Power Up; Manufacturing, Packaging or Shipping Problem; Device Alarm System; Display or Visual Feedback Problem; Failure to Power Up; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT, PRIOR TO USE, THE MONITOR FAILED TO FULLY POWER ON. THE SCREEN DID NOT DISPLAY ANY IMAGES, AND ONLY ONE AMBER LIGHT TURNED ON. THE MONITOR DID NOT PRODUCE ANY ALARMING SOUNDS. IT WAS POWER CYCLED, AND A NEW BATTERY WAS USED TO ISOLATE THE FAILURE. THIS WAS AN OUT-OF-BOX FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22860657 ↗
  • MalfunctionMar 19, 2022Device Alarm System; Device Damaged Prior to Use; Connection Problem; Device Alarm System; Device Damaged Prior to Use; Connection Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE UNIT HAD A LOOSE SPO2 CONNECTOR AND A BROKEN GAS PORT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13825069 ↗
  • MalfunctionJan 4, 2022Failure to Power Up; Temperature Problem; Failure to Power Up; Temperature Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE UNIT WOULD NOT POWER ON. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THERE WAS A BURNT SMELL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13150561 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123690. One submission often covers many models, so check the product and lot details in each record.

  • Z-1577-2016 · Apr 13, 2016 · Terminated