Vital Sync VITALSYNCSW01 (UDI 10884521780668)Recalls & reported problems

Covidien LP · VITAL SYNC V2 SOFTWARE · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521780668), received between Mar 28, 2024 and Mar 28, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K160718); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Computer Software Problem1
  • Device Alarm System1
  • Labelling, Instructions for Use or Training Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 28, 2024Device Alarm System; Computer Software Problem; Labelling, Instructions for Use or Training Problem; Device Alarm System; Computer Software Problem; Labelling, Instructions for Use or Training Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, MULTIPLE TILES HAD MULTIPLE PATIENTS ADMITTED TO THEM AND ALARMS FOR THE CURRENT PATIENT DID NOT ALERT ON THE TILE DUE TO MANY PATIENTS AND THE ALARMS WERE GETTING PUSHED OUT OF THE TILE. IT WAS FOUND OUT THAT PATIENTS WERE NOT GETTING DISCHARGE FROM THE REMOTE SURVEILLANCE SYSTEM BEFORE NEW PATIENT WAS ADMITTED. AN INTERNAL STAFF WAS ON THE SITE AND PATIENTS THAT WERE ASSIGNED TO THE SAME TILE WERE DISCHARGED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18995495 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160718. One submission often covers many models, so check the product and lot details in each record.

  • Z-1557-2026 · Jan 30, 2026 · Open, Classified
    HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital

Other TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY devices

All TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY devices →