Vital Sync VITALSYNCSW01 (UDI 10884521780668)Recalls & reported problems
Covidien LP · VITAL SYNC V2 SOFTWARE · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521780668), received between Mar 28, 2024 and Mar 28, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Computer Software Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K160718); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Computer Software Problem1
- Device Alarm System1
- Labelling, Instructions for Use or Training Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 28, 2024Device Alarm System; Computer Software Problem; Labelling, Instructions for Use or Training Problem; Device Alarm System; Computer Software Problem; Labelling, Instructions for Use or Training Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, MULTIPLE TILES HAD MULTIPLE PATIENTS ADMITTED TO THEM AND ALARMS FOR THE CURRENT PATIENT DID NOT ALERT ON THE TILE DUE TO MANY PATIENTS AND THE ALARMS WERE GETTING PUSHED OUT OF THE TILE. IT WAS FOUND OUT THAT PATIENTS WERE NOT GETTING DISCHARGE FROM THE REMOTE SURVEILLANCE SYSTEM BEFORE NEW PATIENT WAS ADMITTED. AN INTERNAL STAFF WAS ON THE SITE AND PATIENTS THAT WERE ASSIGNED TO THE SAME TILE WERE DISCHARGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18995495 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160718. One submission often covers many models, so check the product and lot details in each record.
- Z-1557-2026 · Jan 30, 2026 · Open, ClassifiedHealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital
Other TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY devices
- NA ORG-9110ANihon Kohden Corporation · TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY272 reports
- NA GZ-130PNihon Kohden Corporation · TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY1 death203 reports
- NA ORG-9100ANihon Kohden Corporation · TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY106 reports
- Midmark Midmark Digital ECGCardioline SPA · Transmitters And Receivers, Physiological Signal, Radiofrequency1 recall1 report
- Vital Sync VITALSYNCSW03Covidien LP · TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY1 recall1 report
- Vital Sync VITALSYNCSW03Covidien LP · TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY1 recall1 report
- VistaCenter 6200-01Vital Connect, Inc. · Transmitters And Receivers, Physiological Signal, Radiofrequency1 recall1 report
- BioButton v3 (BBN-r, different batch) BIOST06040Biointellisense, Inc. · Transmitters And Receivers, Physiological Signal, Radiofrequency1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.