DAR 444SP02014 (UDI 10884521810785)Recalls & reported problems
Covidien LP · Closed Suction System, Replaceable Catheter Endotracheal Length · Class 1
All models of DAR (35)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521810785), received between Jan 17, 2024 and Mar 1, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak2
- Crack1
Patient problems
- Hypoventilation1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 1, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, AN AIR LEAK WAS HEARD IN THE INTUBATION TUBE. OBSERVATION OF THE DEVICE REVEALED THAT THE CLOSED SYSTEM WAS BROKEN ON THE RESPIRATOR SIDE. THE SYSTEM WAS CHANGED IMMEDIATELY AND FIVE MINUTES AFTER FITTING THE NEW CLOSED SYSTEM, THE SAME PROBLEM WAS NOTED WITH A LEAK. WHEN THE CLOSED SYSTEM WAS REPLACED WITH A HYPOXEMIC CONNECTOR, THE CLOSED SYSTEM BROKE, CRUMBLING INTO THE VENTILATOR CIRCUIT AND MAKING IT IMPOSSIBLE TO RECONNECT THE PATIENT AND VENTILATE. THE DOCTOR HAD TO INTERVENE URGENTLY, AND THE PATIENT WAS VENTILATED USING A BAVU (SELF-FILLING BALLOON WITH ONE-WAY VALVE). THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Insufficient Information
MAUDE report 18817963 ↗InjuryMar 1, 2024Crack; Crack
ACCORDING TO THE REPORTER, AN AIR LEAK WAS HEARD IN THE INTUBATION TUBE. OBSERVATION OF THE DEVICE REVEALED THAT THE CLOSED SYSTEM WAS BROKEN ON THE RESPIRATOR SIDE. THE SYSTEM WAS CHANGED IMMEDIATELY AND FIVE MINUTES AFTER FITTING THE NEW CLOSED SYSTEM, THE SAME PROBLEM WAS NOTED WITH A LEAK. WHEN THE CLOSED SYSTEM WAS REPLACED WITH A HYPOXEMIC CONNECTOR, THE CLOSED SYSTEM BROKE, CRUMBLING INTO THE VENTILATOR CIRCUIT AND MAKING IT IMPOSSIBLE TO RECONNECT THE PATIENT AND VENTILATE. THE DOCTOR HAD TO INTERVENE URGENTLY, AND THE PATIENT WAS VENTILATED USING A BAVU (SELF-FILLING BALLOON WITH ONE-WAY VALVE). THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Hypoventilation
MAUDE report 18817900 ↗MalfunctionJan 17, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE BREATHING SYSTEM SUCTION HAD AIR LEAKAGE IN THE CONNECTION BETWEEN THE SYSTEM AND THE SWIVEL CONNECTOR DURING USE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18527382 ↗
Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices
- Ballard* Closed Suction System 93826Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall70 reports
- Ballard* Closed Suction System 92431Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL78 reports
- Medline DYNCPDS14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall16 reports
- Ballard* Turbo-Cleaning Closed Suction System 98965Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall16 reports
- Ballard* Turbo-Cleaning Closed Suction System 98263Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall11 reports
- Ballard* Closed Suction System for Adults 98310Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall1 death9 reports
- Medline DYNCPTP14TMedline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall6 reports
- Medline DYNCPTP14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page