DAR 444SP02012 (UDI 10884521822887)Recalls & reported problems
Covidien LP · Closed Suction System, Replaceable Catheter,Endotracheal Length,DETACH. DSE · Class 1
All models of DAR (35)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884521822887), received between Sep 10, 2024 and Mar 12, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak3
- Use of Device Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Low Oxygen Saturation1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMar 12, 2026Use of Device Problem
IT WAS REPORTED THAT DURING A ROUTINE SYSTEM CHANGE INVOLVING THE CLOSE SUCTION CIRCUIT, A PATIENT EXPERIENCED A DROP IN PERIPHERAL OXYGEN SATURATION DUE TO EXISTING SECRETIONS IN THE LUNGS. AN ATTEMPT WAS MADE TO SUCTION THE SECRETIONS ENDOTRACHEALLY, BUT SUCTION COULD NOT BE GENERATED BECAUSE THE BLUE SUCTION VALVE BUTTON WAS NOT FUNCTIONING. THE PERIPHERAL OXYGEN SATURATION CONTINUED TO FALL. PRE-OXYGENATION WAS PERFORMED, AND SECRETIONS WERE SUBSEQUENTLY REMOVED USING CONVENTIONAL STERILE SUCTIONING, AFTER WHICH OXYGEN SATURATION RETURNED TO NORMAL LEVELS.
Patient problems: Low Oxygen Saturation
MAUDE report 24583294 ↗MalfunctionJan 8, 2026Use of Device Problem
IT WAS REPORTED THAT THE BREATHING SUCTION SYSTEM HAD A LEAK OCCURRED DURING PRE-USAGE INSPECTION. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24007472 ↗MalfunctionNov 7, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, DURING USE, THE BREATHING SYSTEM SUCTION HAD A LEAK. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20629462 ↗MalfunctionSep 23, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, DURING USE, AN AIR LEAK WAS CONFIRMED FROM THE VENTILATOR WAVEFORM AND WAS DETECTED IN THE SERPENTINE TUBE PART OF DEVICE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20286443 ↗MalfunctionSep 10, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, DURING USE, THE GAS FROM THE VENTILATOR WAS LEAKING FROM THE DEVICE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20180484 ↗
FDA clearance and classification
The FDA regulates product code BSY (catheters, suction, tracheobronchial) under 21 CFR 868.6810 as a Class I device. 21 CFR 868.6810 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the DAR 444SP02012
Has the DAR 444SP02012 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10884521822887), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the DAR 444SP02012?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Sep 10, 2024 and Mar 12, 2026. In those reports, the most common specific device problems are “Gas/Air Leak” (3) and “Use of Device Problem” (2); the most common patient problem is “Low Oxygen Saturation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the DAR 444SP02012?
None of the reports are classified as deaths, 1 as an injury and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the DAR 444SP02012 FDA approved?
Devices of this type (product code BSY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Covidien LP.
Is the DAR 444SP02012 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the DAR 444SP02012, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand DAR, model 444SP02012). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices
- Ballard* Closed Suction System 93826Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall70 reports
- Ballard* Closed Suction System 92431Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL78 reports
- Ballard* Turbo-Cleaning Closed Suction System 98965Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall15 reports
- Medline DYNCPDS14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall15 reports
- Ballard* Turbo-Cleaning Closed Suction System 98263Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall11 reports
- Ballard* Closed Suction System for Adults 98310Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall1 death9 reports
- Medline DYNCPTP14TMedline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall5 reports
- Medline DYNCPTP14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page