DAR 444SP02014 (UDI 10884521822894)Recalls & reported problems
Covidien LP · Closed Suction System, Replaceable Catheter Endotracheal Length · Class 1
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521822894), received between Aug 4, 2025 and Jan 21, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak2
Patient problems
- Low Oxygen Saturation1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 21, 2026Gas/Air Leak
IT WAS REPORTED THAT LEAKAGE OCCURRED ON THE WHITE ATTACHMENT PIECE TO THE ENDOTRACHEAL TUBE OF THE SUCTION CATHETER, AND NO AIR WAS ENTERING THE LUNGS. THE PROBLEM WAS SOLVED BY CHANGING THE SUCTION SYSTEM. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24140301 ↗InjuryAug 4, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT DURING THE INSTALLATION OF A CLOSED SUCTION SYSTEM SEVERAL LEAKS WERE OBSERVED ON THE FAN SIDE DEVICE. THE VENTILATOR WAS SET TO DELIVER A VOLUME OF 420, BUT ONLY 160 WAS ACTUALLY DELIVERED. THE PATIENT EXPERIENCED INITIAL INSTABILITY AND A SLIGHT WORSENING OF DESATURATION, WHICH WAS QUICKLY CORRECTED. TWO NURSES AND A PHYSIOTHERAPIST ATTEMPTED TO TIGHTEN THE SYSTEM AND REINFORCE THE TUBING, BUT DUE TO THE PATIENT'S INSTABILITY, THE SYSTEM WAS REPLACED WITH A TRACHEOSTOMY. THE SYSTEM WAS THEN REPLACED WITH A STANDARD CATHETER MOUNT PRODUCT, WHICH RESTORED NORMAL VENTILATION VOLUMES.
Patient problems: Low Oxygen Saturation
MAUDE report 22705315 ↗
Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices
- Ballard* Closed Suction System 93826Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall70 reports
- Ballard* Closed Suction System 92431Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL78 reports
- Medline DYNCPDS14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall16 reports
- Ballard* Turbo-Cleaning Closed Suction System 98965Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall16 reports
- Ballard* Turbo-Cleaning Closed Suction System 98263Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall11 reports
- Ballard* Closed Suction System for Adults 98310Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall1 death9 reports
- Medline DYNCPTP14TMedline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall6 reports
- Medline DYNCPTP14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall4 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.