DAR 444SP02014 (UDI 10884521822894)Recalls & reported problems

Covidien LP · Closed Suction System, Replaceable Catheter Endotracheal Length · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521822894), received between Aug 4, 2025 and Jan 21, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak2

Patient problems

  • Low Oxygen Saturation1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 21, 2026Gas/Air Leak

    IT WAS REPORTED THAT LEAKAGE OCCURRED ON THE WHITE ATTACHMENT PIECE TO THE ENDOTRACHEAL TUBE OF THE SUCTION CATHETER, AND NO AIR WAS ENTERING THE LUNGS. THE PROBLEM WAS SOLVED BY CHANGING THE SUCTION SYSTEM. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24140301 ↗
  • InjuryAug 4, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT DURING THE INSTALLATION OF A CLOSED SUCTION SYSTEM SEVERAL LEAKS WERE OBSERVED ON THE FAN SIDE DEVICE. THE VENTILATOR WAS SET TO DELIVER A VOLUME OF 420, BUT ONLY 160 WAS ACTUALLY DELIVERED. THE PATIENT EXPERIENCED INITIAL INSTABILITY AND A SLIGHT WORSENING OF DESATURATION, WHICH WAS QUICKLY CORRECTED. TWO NURSES AND A PHYSIOTHERAPIST ATTEMPTED TO TIGHTEN THE SYSTEM AND REINFORCE THE TUBING, BUT DUE TO THE PATIENT'S INSTABILITY, THE SYSTEM WAS REPLACED WITH A TRACHEOSTOMY. THE SYSTEM WAS THEN REPLACED WITH A STANDARD CATHETER MOUNT PRODUCT, WHICH RESTORED NORMAL VENTILATION VOLUMES.

    Patient problems: Low Oxygen Saturation

    MAUDE report 22705315 ↗