Skip to content

Shiley 70XLTCPRecalls & reported problems

Covidien LP · Tracheostomy Tube XLT Cuffed,Proximal with Disposable Inner Cannula · Class 2

All models of Shiley (387)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10884522027908), received between Jun 27, 2017 and Dec 9, 2025 — classified as 4 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Defective Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Component3
  • Material Split, Cut or Torn2
  • Use of Device Problem2
  • Gas/Air Leak1
  • Human-Device Interface Problem1

Patient problems

  • Unintended Extubation3
  • Hypoxia2
  • No Clinical Signs, Symptoms or Conditions2
  • Abnormal Blood Gases1
  • Bradycardia1
  • Extubate1
  • Insufficient Information1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 9, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT DURING USE OF THE TRACHEOSTOMY TUBE, A PATIENT WAS FOUND TO BE BRADYCARDIC, HYPOXIC, AND HYPERCAPNIC. THE TUBE HAD A CUFF LEAK THAT WOULD NOT STAY INFLATED, AND THE OLD TUBE WAS FOUND TO HAVE A HOLE IN THE CUFF. THE TUBE WAS REPLACED.

    Patient problems: Abnormal Blood Gases; Bradycardia; Hypoxia; Unintended Extubation

    MAUDE report 23762753 ↗
  • MalfunctionMay 16, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT A DEFECTIVE TRACH TUBE WAS IDENTIFIED PRIOR TO USE DURING PRE-OPERATIVE PREPARATION. THE SPECIFIC ISSUE WAS THAT THE CUFF WOULD NOT INFLATE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038685 ↗
  • MalfunctionFeb 4, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE'S CUFF HAD A INFLATION OR DEFLATION ISSUE. IT WAS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: Insufficient Information

    MAUDE report 21303247 ↗
  • MalfunctionJan 31, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF HAD BREAKAGE DURING USE. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 18615970 ↗
  • MalfunctionOct 26, 2022Defective Component; Material Split, Cut or Torn; Defective Component; Material Split, Cut or Torn

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE UNIT WAS PRE-INFLATED AND WAITED FOR ONE MINUTE THEN FOUND THAT THE FILLED CUFF DEFLATED. THE PROCESS WAS REPEATED SEVERAL TIMES AND DETERMINED THAT THE CUFF WAS DAMAGED. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15677888 ↗
  • InjuryMay 11, 2021Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE PRODUCT KINKED RESULTING IN MEDICAL AIRWAY EMERGENCY. THE INNER CANNULA WAS AN INTEGRAL PART OF THE 22MM ATTACHMENT FOR VENTILATOR OR SELF-INFLATING BAG. PATIENT UNDERWENT UNPLANNED EMERGENCY AIRWAY SURGERY TO CORRECT.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11804064 ↗
  • InjuryFeb 9, 2021Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE HERNIATED WHICH CAUSED THE PATIENT TO HAVE A SERIOUS ISSUE. CUFF BULGED ON ONE SIDE AND PATIENT DECOMPENSATED, WENT INTO ASYSTOLE SLEEP, O2 DROPPED ENOUGH THAT HEARTBEAT BECAME IRREGULAR, PROBABLY THE CUFF DID NOT SEAL DURING SURGERY RESULTING IN LOSS OF SUPPLIED OXYGEN. RE CANNULATED AND USED ENDOTRACHEAL (ET) TUBE, A NEW TRACH TUBE WAS PLACED AND SECURED. THE PATIENT WAS INJURED BUT RECOVERED.

    Patient problems: Hypoxia; Unintended Extubation

    MAUDE report 11300832 ↗
  • InjuryJun 27, 2017Human-Device Interface Problem; Human-Device Interface Problem

    IT WAS REPORTED THAT, WHILE THE 980 VENTILATOR WAS BEING USED ON A PATIENT, COMPLICATIONS WITH THE PATIENT OCCURRED. THE CUSTOMER STATED THAT THE VENTILATOR PERFORMED AND ALARMED AS INTENDED WITH NO MALFUNCTION. IT IS THE CUSTOMER'S BELIEF THAT THE LARYNX MIGRATED OVER TIME, CAUSING THE SHILEY 8.0 CUFFED TRACH TO DISLODGE BETWEEN TRACHEAL RINGS AND ADIPOSE TISSUE, RESULTING IN BILATERAL PNEUMOTHORACES AND SUBCUTANEOUS EMPHYSEMA. AS A RESULT, THE PATIENT WAS TAKEN TO THE OPERATING ROOM AND A 2ND TRACH WAS PLACED (SHILEY 7 XLT) WHICH WAS REPORTEDLY STILL NOT LONG ENOUGH. AN EMERGENCY CRICOIDECTOMY WAS PERFORMED AND AN ENDOTRACHEAL TUBE WAS PLACED WHILE THE STOMA WAS OPENED MORE AND A BOVINA TRACH WAS EVENTUALLY PLACED.

    Patient problems: Extubate

    MAUDE report 6671478 ↗

Other Tube tracheostomy and tube cuff devices