Skip to content

Nellcor MAXN-NSRecalls & reported problems

Covidien LP · Neonatal-Adult SpO2 Sensor · Class 2

All models of Nellcor (64)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884522041638), received between Aug 17, 2018 and Nov 5, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K012891); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem1
  • Erratic or Intermittent Display1
  • High Readings1
  • Low Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Injury1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2024Erratic or Intermittent Display; Low Readings; Erratic or Intermittent Display; Low Readings

    ACCORDING TO THE REPORTER, DURING USE, THE PATIENT PULSE RATE WAS NORMALLY 120 AND WOULD JUMP SIGNIFICANTLY HIGHER AND INTERMITTENTLY TO AS HIGH AS 225 THAT RESULTS IN TACKY ALARM. THE SPO2 VALUE WAS ALSO SEEN TO DROP INTERMITTENTLY TO LOWER VALUES, AS LOW AS 50 FROM ITS NORMAL VALUE OF 97. CHANGING THE CABLES, SENSORS AND THE MONITORS DID NOT RESOLVE THE ISSUE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20611939 ↗
  • MalfunctionDec 1, 2020High Readings; High Readings

    ACCORDING TO THE REPORTER, THE DEVICE WAS NOT READING CORRECTLY WHEN HOOKED UP TO THE PATIENT. IT WAS STATED THAT THE READING WAS HIGHER THAN THE ACTUAL READING WHEN A RUBBER PIECE WAS ADDED OR REMOVED. THERE WAS NO ANY ALLEGATION OF PATIENT DEATH OR SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10927245 ↗
  • InjuryAug 17, 2018Contamination /Decontamination Problem; Contamination /Decontamination Problem

    ACCORDING TO THE REPORTER, THE DEVICE A CAUSED A BURN ON THE INFANT. THE CUSTOMER REPORTED THAT THE DEVICE WAS USED FOR ABOUT 36 HOURS ON THE INFANT¿S FOOT.

    Patient problems: Injury

    MAUDE report 7793495 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K012891. One submission often covers many models, so check the product and lot details in each record.

  • Z-0664-05 · Oct 18, 2004 · Terminated
    Nellcor brand OxiMax Pulse Oximeter, model number N-595 Made in Ireland; Distributed by Tyco Healthcare Group, LP, Nellcor Puritan Bennett, Inc., Pleasanton, CA 94588 U.S.A.

Other Oximeter devices

All Oximeter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page