Argyle 8888505172Recalls & reported problems
Cardinal Health 200, LLC · Yankauer Open Tip with On/Off Switch · Class 2
All models of Argyle (33)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884527020010), received between Sep 16, 2020 and Jun 5, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Detachment of Device or Device Component1
- Suction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THAT DURING DELIVERY FROM THE DISTRIBUTOR TO THE HOSPITAL, WELDED PARTS BECAME DIS-ATTACHED. THE PRODUCT WAS UNUSED. PER ADDITIONAL INFORMATION RECEIVED, THERE WAS NO MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17058277 ↗MalfunctionMay 19, 2022Break; Break
THE CUSTOMER REPORTED DURING ORTHOPEDIC SURGERY, THE YANKAUER WAS USED FOR BLOOD SALVAGE. AFTER THE OPERATION THEY NOTICED THE TIP WAS CHIPPED. THERE WAS NO EFFECT ON THE PATIENT AND NO MEDICAL INTERVENTION REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14434277 ↗MalfunctionSep 16, 2020Suction Problem; Suction Problem
THE CUSTOMER REPORTED THAT THE YANKAUER WAS USED DURING CARDIAC SURGERY AND IT WAS UNABLE TO SUCTION. A SECOND YANKAUER WAS OPENED AND USED HOWEVER IT WAS ALSO UNABLE TO SUCTION. THE FLUID REACHED A LITTLE OVER THE BENDING POINT OF THE CATHETER BUT WOULD GET STUCK THERE. THE CATHETER WAS USED AGAIN AND THE SUCTION PRESSURE WAS RAISED BUT NOTHING HAD CHANGED. THE YANKAUER HAD TO BE REPLACED WITH ANOTHER MAKERS CATHETER WHICH SUCTIONED WITHOUT A PROBLEM. IT WAS CONFIRMED THAT THERE WERE NO CLOGS OR ISSUES WITH THE TUBE BETWEEN THE CATHETER AND THE ASPIRATOR. THE TUBE'S SHAPE WAS A LITTLE FLATTENED DURING SUCTIONING. TO VERIFY THERE WERE NO ISSUES WITH THE TUBING THE CATHETER WAS DETACHED FROM THE TUBE AND THE SUCTION WITH THE TUBE WAS CHECKED, THERE WERE NO ISSUES WITH THE TUBE. THE USER UNDERSTANDS BOTH OPERATION PROCEDURE AND STRUCTURE OF THE BLACK CONTROL LEVER, AND THOSE HAD BEEN CHECKED AT THE OPERATING FIELD. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10543703 ↗
Other CATHETER AND TIP, SUCTION devices
- J-VAC 2160Ethicon Inc. · CATHETER AND TIP, SUCTION152 reports
- Wayne G56537Cook Incorporated · CATHETER AND TIP, SUCTION2 recalls33 reports
- Bard® Suction Connector EN0066000C. R. Bard, Inc. · CATHETER AND TIP, SUCTION1 recall6 reports
- CLEARVIEW® 31200Medtronic, Inc. · CATHETER AND TIP, SUCTION1 recall0 reports
- ARROW IPN039282Teleflex Incorporated · CATHETER AND TIP, SUCTION1 recall0 reports
- ARROW IPN037088Teleflex Incorporated · CATHETER AND TIP, SUCTION1 recall0 reports
- ARROW IPN922898Teleflex Incorporated · CATHETER AND TIP, SUCTION1 recall0 reports
- Medline Industries, Inc. OR929KMedline Industries, Inc. · CATHETER AND TIP, SUCTION1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page