Kangaroo B5FDRecalls & reported problems

Cardinal Health 200, LLC · Kangaroo Omni Feeding Set, For use with ENfit Connection System, 500 mL. · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885380177125), received between Feb 6, 2025 and Jan 13, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Air/Gas in Device3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 13, 2026Air/Gas in Device

    THE CUSTOMER REPORTED BUBBLES WERE OBSERVED IN THE SET DURING FEEDING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24052909 ↗
  • MalfunctionMar 7, 2025Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THEY WERE USING THE KANGAROO BAG (FOR THE KANGAROO PUMP) TO SET UP A FEED FOR A PATIENT. WHEN THEY PRIMED THE BAG, AIR WAS NOTICED IN THE LINE AND THEY FOUND THE BAG TO BE LEAKING. THEY NOTICED A CUT IN THE TUBING THAT WAS CONTRIBUTING TO THE LEAKING. THEY TRIED TO TAPE THE CUT TO SEE IF THE FEED COULD STILL RUN, BUT THE LEAK PERSISTED. THEY TRANSFERRED THE MILK TO A NEW KANGAROO BAG THAT WAS NOT LEAKING AND PUT THE OTHER ONE TO THE SIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21546571 ↗
  • MalfunctionFeb 6, 2025Air/Gas in Device; Air/Gas in Device

    THE CUSTOMER REPORTED THE PUMP IS PULLING AIR THROUGH INSTEAD OF MILK. BIOMED FOUND WHEN THEY HOOKED UP THE CASSETTE, AS SOON AS THE PUMP STARTED AND IT TRIED TO LOCK THE CASSETTE ON, IT PUSHED THE CASSETTE AWAY MAKING IT SIT AT A WEIRD ANGLE. THE CASSETTE SEEMS TO HAVE A BIT OF DEFORMITY ON THE INNER PLASTIC PART. THEY FOUND THE ISSUE SEEMED TO BE THE COMBINATION OF THE AFFECTED CASSETTE AND THE PUMP. THE CASSETTE ON A DIFFERENT PUMP DOES NOT HAVE THE ISSUE, AND THE PUMP WITH A NEW CASSETTE DOES NOT HAVE THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21322419 ↗

Other Pump, infusion, enteral devices