JACKSON-PRATT JP-2190Recalls & reported problems
Cardinal Health 200, LLC · JP CHANNEL DRAIN 19FR, FULL FLUTES · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885380182457), received between Dec 3, 2025 and Dec 3, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Material Separation” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 14, 2022Open, Classified
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
Z-0971-2023 · Root cause: Unknown/Undetermined by firm · FDA record ↗
Most reported problems
Device problems
- Material Separation1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 3, 2025Material Separation; Material Separation
WE HAD AN UNFORTUNATE SITUATION WITH THE CARDINAL HEALTH JACKSON-PRATT CHANNEL DRAINS JP-2190 THAT WE RECENTLY ORDERED. THE CHANNEL PART THAT GOES UNDER THE SKIN, SEPARATED FROM THE PART THAT IS OUTSIDE OF THE SKIN, THEREFORE CAUSING THE CHANNEL PART TO GET STUCK UNDER THE SKIN AND THE DOCTOR HAD TO OPEN THE AREA AND RETRIEVE THE CHANNEL PART OF THE DRAIN.
Patient problems: Foreign Body In Patient
MAUDE report 23710419 ↗
Other CATHETER, IRRIGATION devices
- Navarre Opti-Drain NOD8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION901 reports
- Navarre NNU8LPTBard Access Systems, Inc. · CATHETER, IRRIGATION736 reports
- BLAKE 2227Ethicon Inc. · CATHETER, IRRIGATION331 reports
- Navarre NNU6LPTBard Access Systems, Inc. · CATHETER, IRRIGATION211 reports
- Navarre Opti-Drain NOD10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION178 reports
- Navarre NNU10LPTBard Access Systems, Inc. · CATHETER, IRRIGATION127 reports
- BLAKE 2233Ethicon Inc. · CATHETER, IRRIGATION114 reports
- Vasoshield VH-5001Maquet Cardiovascular LLC · Catheter, Irrigation1 recall27 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.