Skip to content

JACKSON-PRATT JP-2212Recalls & reported problems

Cardinal Health 200, LLC · JP CHAN DRN SIL FLT 7MM FULL W/TRO · Class 1

All models of JACKSON-PRATT (20)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885380182495), received between May 10, 2023 and Jan 12, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 14, 2022Class IIOpen, Classified

    Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.

    Z-0975-2023 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break3

Patient problems

  • Insufficient Information2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 12, 2024Break; Break

    MW5148639: THE NEURO APRN(ADVANCED REGISTERED NURSE PRACTITIONER) ATTEMPTED TO REMOVE THE JP (JACKSON-PRATT) DRAIN WHICH BROKE AT THE INSERTION SITE AND A DRAIN FRAGMENT REMAINED IN THE PATIENT. THE NEUROSURGEON ATTEMPTED TO REMOVE THE FRAGMENT AT THE BEDSIDE WITHOUT SUCCESS. THE PT RETURNED TO THE OR(OPERATING ROOM) FOR SURGICAL REMOVAL OF THE JP DRAIN FRAGMENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 18505281 ↗
  • MalfunctionMay 19, 2023Break; Break

    CUSTOMER REPORTED THAT DURING REMOVAL OF JP DRAIN FROM THE PATIENT, IT BROKE. THIS IS 1 OF 3 EVENTS THAT HAS HAPPENED AT OUR INSTITUTION WITH THE SAME DRAIN. EVENT DATE IS UNKNOWN. NO FURTHER INFORMATION WAS PROVIDED WHEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 16969507 ↗
  • MalfunctionMay 10, 2023Break; Break

    DURING REMOVAL OF JP DRAIN FROM THE PATIENT'S SCALP, THE JUNCTION BETWEEN THE CLEAR DRAIN TUBING AND FLUTED WHITE PIECE APPEARED TO BE SEVERED AND EMERGED FROM SCALP, LEAVING THE LENGTH OF THE SUBGALEAL PORTION RETAINED IN THE WOUND. THIS IS THE 3RD EVENT THAT THIS HAS HAPPENED AT OUR INSTITUTION WITH THE SAME DRAIN. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 16907876 ↗

FDA clearance and classification

The FDA regulates product code GBX (catheter, irrigation) under 21 CFR 878.4200 as a Class I device. 21 CFR 878.4200 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the JACKSON-PRATT JP-2212

Has the JACKSON-PRATT JP-2212 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0975-2023 (Class II), began on Dec 14, 2022: Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the JACKSON-PRATT JP-2212?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between May 10, 2023 and Jan 12, 2024. In those reports, the most common specific device problem is “Break” (3); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the JACKSON-PRATT JP-2212?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the JACKSON-PRATT JP-2212 FDA approved?

Devices of this type (product code GBX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cardinal Health 200, LLC.

Who makes the JACKSON-PRATT JP-2212, and is it still sold?

The FDA's device database lists Cardinal Health 200, LLC as the company responsible for it (brand JACKSON-PRATT, model JP-2212). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources and updates