Skip to content

ACHIEVE A1411Recalls & reported problems

Carefusion 2200, Inc · ACHIEVE NEEDLE BIOP SOFT TISSUE 14GX11CM · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10885403088346), received between Aug 13, 2025 and Nov 6, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 5 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 5 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Fire9
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 6, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    DURING A BREAST BIOPSY USING THE ACHIEVE BIOPSY NEEDLE, THE RELEASE FIRING MODE WAS VERY WEAK RESULTING IN NO SAMPLE CAPTURES. THE CUSTOMER INDICTED THAT THEY BELIEVED THE SPRING COULD HAVE BEEN DEFECTIVE. A DIFFERENT NEEDLE FROM ANOTHER LOT WAS THEN USED AND WAS REPORTED TO WORK WELL. NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23487920 ↗
  • MalfunctionNov 3, 2025Failure to Fire; Failure to Fire

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO TAKE BIOPSIES THE BIOPSY DEVICE FAILED TO COLLECT A SAMPLE. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23458586 ↗
  • MalfunctionNov 3, 2025Failure to Fire; Failure to Fire

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO TAKE BIOPSIES THE BIOPSY DEVICE FAILED TO COLLECT A SAMPLE. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23458573 ↗
  • MalfunctionNov 3, 2025Failure to Fire; Failure to Fire

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO TAKE BIOPSIES THE BIOPSY DEVICE FAILED TO COLLECT A SAMPLE. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23458405 ↗
  • MalfunctionNov 3, 2025Failure to Fire; Failure to Fire

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO TAKE BIOPSIES THE BIOPSY DEVICE FAILED TO COLLECT A SAMPLE. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23458383 ↗
  • MalfunctionAug 27, 2025Failure to Fire; Failure to Fire

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO TAKE BIOPSIES THE BIOPSY DEVICE FAILED TO COLLECT A SAMPLE. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22910000 ↗
  • MalfunctionAug 27, 2025Failure to Fire; Failure to Fire

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO TAKE BIOPSIES THE BIOPSY DEVICE FAILED TO COLLECT A SAMPLE. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22909996 ↗
  • MalfunctionAug 27, 2025Failure to Fire; Failure to Fire

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO TAKE BIOPSIES THE BIOPSY DEVICE FAILED TO COLLECT A SAMPLE. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22909974 ↗
  • MalfunctionAug 27, 2025Failure to Fire; Failure to Fire

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO TAKE BIOPSIES THE BIOPSY DEVICE FAILED TO COLLECT A SAMPLE. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22909966 ↗
  • MalfunctionAug 13, 2025Failure to Fire; Failure to Fire

    THE ACCOUNT ALLEGES THAT WHEN ATTEMPTING TO TAKE BIOPSIES THE BIOPSY DEVICE FAILED TO COLLECT A SAMPLE. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22794317 ↗

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →