Skip to content

Alaris, SmartSite 2433-0007Recalls & reported problems

BD Switzerland Sàrl · Pump Infusion Set, Back Check Valve, 2 Y-Sites, 297 cm, (117 in) · Class 2

All models of Alaris, SmartSite (160)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885403227974), received between May 18, 2020 and May 18, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 6, 2019Class ITerminated

    An incomplete occlusion can be caused by a variation in the wall thickness of the pumping segment of the affected infusion sets used with the. The issue has the potential to lead to unintended delivery of medication when the pump module is not in running status or can result in faster than expected drug delivery flow when the pump is infusing.

    Z-1770-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 18, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED FROM THE HOI MED SURGICAL FLOOR THAT A PRIMARY TUBING SET LEAKED. THERE WAS NO REPORT OF HARM TO THE PATIENT OR ANY MEDICAL PROFESSIONAL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10064418 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates