Skip to content

MaxZero MZ5306Recalls & reported problems

BD Switzerland Sàrl · Minibore pressure rated extension set, IV connector, no clamp 18 cm (7.0 in) · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885403230592), received between Aug 30, 2018 and Sep 24, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Blocked Connection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Blocked Connection1
  • Fluid/Blood Leak1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Blocked Connection

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50890861 ↗
  • MalfunctionMay 21, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE MINIBORE PRESSURE, IV CONNECTOR NO CLAMP TUBING WAS CRACKED AND LEAKED MEDICINE DURING USE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 4TH OF 4 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WHILE DOING ASSESSMENT I NOTICED SOME DROPLETS ON THE MALE CONNECTOR OF REMODULIN. I LOOKED CLOSER THINKING IT WAS MAYBE SMALL AIR BUBBLES AND MY GLOVE LOOKED WET. I EVALUATED CLOSER AND NOTICED A CRACK ON THE MAX ZERO OF THE MALE CONNECTOR. I SWITCHED THE CONNECTOR AND REQUESTED A NEW SYRINGE. THE MAX ZERO HAD CRACKED AT THE SEAM. NOT MUCH HAD BEEN LOST WHEN I EVALUATED THE FLOOR, BUT UNSURE OF HOW LONG IT WAS LEAKING. I CONTINUOUSLY CHECKED THE LINE AND A FEW HOURS LATER NOTICED MY GLOVE WAS SLIGHTLY WET WHEN RECHECKING, BUT THIS TIME I NOTICED IT ON THE REGULAR STRAIGHT LINE TUBING. AT THIS TIME THIS TUBING HAD NOT BEEN CHANGED ONLY THE CONNECTOR MALE PIECE HAD BEEN CHANGED. I CHANGED THE ENTIRE LINE AT THIS MOMENT AND THIS STRAIGHT TUBING ALSO NOTED TO HAVE A CRACK AT THE SEAM AS WELL."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11867545 ↗
  • MalfunctionAug 30, 2018Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THE NIGHT RN NOTED PGE LEAKING AT THE "PIGTAIL". THE NURSE HAD TO REQUEST A NEW PGE FROM PHARMACY AND REPLACE THE TUBING THE PGE WAS INFUSING THROUGH. THE CUSTOMER LATER REPORTED THE "PIGTAIL" IS A MZ5306. THERE WAS NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7833029 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates