Alaris, SmartSite 11532269Recalls & reported problems
BD Switzerland Sàrl · Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, Low Sorbing Tubing (PE Lined), 2 Y-Sites 318 cm (125 in) · Class 2
All models of Alaris, SmartSite (160)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This BD Alaris SmartSite pump infusion set has 340 reports, 6 of them injuries and no deaths. The reports mostly describe fluid leaking or splashing, parts separating, and blocked or broken tubing. There are 3 open recalls: performance data discrepancies that may affect pump performance, pump flow and bolus accuracy outside established ranges, and sets labeled DEHP-free that contain DEHP, a plastic softener.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 333
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 333 adverse event reports that list this exact device (DI 10885403232343), received between Aug 1, 2018 and Sep 29, 2026 — classified as 6 injuries, 326 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (183 reports).
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 19, 2026Class IOpen, Classified
Due to discrepancies in specific attributes in the performance data for infusion sets, which may lead to inappropriate pump performance.
Z-3080-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Jul 8, 2025Class IOpen, Classified
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Z-2180-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Oct 11, 2023Class IIOpen, Classified
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
Z-0349-2024 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash183
- Fluid/Blood Leak38
- Complete Blockage24
- Material Separation24
- Backflow17
- Break15
- Defective Component9
- Separation Problem9
Patient problems
- No Clinical Signs, Symptoms or Conditions241
- No Consequences Or Impact To Patient53
- Insufficient Information11
- No Patient Involvement10
- No Information7
- Chemical Exposure5
- Blood Loss4
- Hypersensitivity/Allergic reaction2
Recent adverse event reports
The 25 most recent of 333 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Complete Blockage
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936980 ↗MalfunctionSep 22, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50860024 ↗MalfunctionSep 11, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50744351 ↗MalfunctionSep 5, 2026Material Deformation
MalfunctionAug 26, 2026Break; Separation Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50462634 ↗MalfunctionAug 26, 2026Break; Separation Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50462619 ↗MalfunctionJun 16, 2026Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT AADS ALARIS IN-LINE FILTER LEAKING AT SHIFT ONSET. AADS INFUSION STOPPED, REPLACED WITH NEW MAINTENANCE INFUSION AND ALARIS TUBING. AADS BAG AND TUBING KEPT AND PLACED IN NICU DIRTY UTILITY ROOM FOR NICU EDUCATORS TO ASSESS. OF NOTE, LINE WAS JUST CHANGED 1-2 HOURS PRIOR TO LEAKING FILTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25526689 ↗MalfunctionJun 16, 2026Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT AADS FILTER NOTED TO BE STICKER WHEN LINE CHECKED. FILTER CLEANSED AND STICKINESS RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25525215 ↗MalfunctionJun 15, 2026Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT INLINE FILTER WAS LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25511622 ↗MalfunctionJun 11, 2026Material Separation; Air/Gas in Device
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: AN IV TUBING CONNECTOR DISCONNECTED WHILE AN RN WAS ADDRESSING AN AIR-IN-LINE ALARM. THE LINE WAS CLAMPED, REPLACED, AND THE INFUSION CONTINUED WITHOUT AFFECTING PATIENT CARE; THE EQUIPMENT WAS RETAINED FOR INVESTIGATION. CAT# OF PRODUCT BEING COMPLAINED: (B)(6). DESCRIPTION OF PRODUCT: SET CHECK VALVE 0.2MIC 2NF 87/6IN 2PC MALE L/L (B)(4) EA/CA LOT OR S/N: UNKNOWN COMPLAINT NOTICED: DURING / AFTER USE PROBLEM FREQUENCY: 1ST TIME PATIENT INJURY: NO QTY AFFECTED: (B)(4) EA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25474092 ↗MalfunctionMay 28, 2026Leak/Splash; Material Split, Cut or Torn
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS DAMAGED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT COMPLETED STERILE LINE CHANGE WITH 2 RNS PRESENT PER UNIT POLICY FOR CENTRAL LINE. ONCE TPN WAS RUNNING ON PUMP, RN NOTICED FLUID DRIPPING ONTO FLOOR. WHEN LINE WAS TRACED, RN FOUND THAT AT ONE OF THE JUNCTIONS ON THE LINE HAD A PINHOLE LEAK CAUSING TPN TO LEAK OUT AND AIR TO GET INTO LINE BREAKING STERILE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25321885 ↗MalfunctionMay 26, 2026Material Separation
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TUBING DETACHED FROM THE POINT OF ATTACHMENT ON THE FILTER. INJURIES OR ADVERSE EVENT: NO. ITEM: 11532269. QUANTITY AFFECTED:1 EACH. SERIAL/LOT NUMBER: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25291686 ↗MalfunctionMay 26, 2026Material Separation
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TUBING DETACHED FROM THE POINT OF ATTACHMENT ON THE FILTER. INJURIES OR ADVERSE EVENT: NO. ITEM: 11532269. QUANTITY AFFECTED:1 EACH. SERIAL/LOT NUMBER: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25291656 ↗MalfunctionMay 20, 2026Separation Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS SEPARATED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY THE CUSTOMER THAT WHILE TRYING TO GET AIR OUT OF THE LINE WHILE IT WAS IN THE PUMP THE TUBING TOTALLY DISCONNECTED AND SPILLED TPN EVERYWHERE. WE HAD TO BREAK IN TO THE LINE AGAIN ON A PICC TO REPLACE THIS WHOLE TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25231979 ↗MalfunctionMay 8, 2026Leak/Splash
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CHEMOTHERAPY IV TUBING LEAKED FROM THE FILTER. IN THIS CASE IT WAS ON THE BEDDING BUT DID NOT TOUCH OR BURN THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25130984 ↗MalfunctionMay 8, 2026Leak/Splash
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE 24-HOUR CHEMOTHERAPY LEAKED FROM THE FILTER IN THE LVP 20D LOW SORB 2SS 0.2MICRON CV TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25130866 ↗MalfunctionMay 8, 2026Leak/Splash
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CHEMOTHERAPY IV TUBING LEAKED FROM THE FILTER. IN THIS CASE, THE CHEMOTHERAPY DRIPPED ONTO THE PATIENT CAUSING A MINOR BURN.
Patient problems: Burn(s)
MAUDE report 25130863 ↗MalfunctionMay 7, 2026Crack; Leak/Splash
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD CRACKED TUBING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING OR AFTER PRIMING A PATIENT'S MEDICATION, NURSES OBSERVED LEAKAGE CAUSED BY A BROKEN TUBE. IN MOST CASES, THE TUBING EITHER SNAPPED WHILE INSERTED IN THE ALARIS PUMP OR THE ENDS OF THE TUBING WERE FOUND TO BE CRACKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25116209 ↗MalfunctionMay 7, 2026Leak/Splash; Material Separation
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING OR AFTER PRIMING A PATIENT'S MEDICATION, NURSES OBSERVED LEAKAGE CAUSED BY A BROKEN TUBE. IN MOST CASES, THE TUBING EITHER SNAPPED WHILE INSERTED IN THE ALARIS PUMP OR THE ENDS OF THE TUBING WERE FOUND TO BE CRACKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25116172 ↗MalfunctionApr 29, 2026Product Quality Problem; Separation Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET HAD FLOW ISSUES.THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM.IT WAS REPORTED BY CUSTOMER THAT BROKE WHEN SHE PRIMED THE NEW LINE WITH A NEW BAG OF AMIODARONE; THE PART OF THE LINE ATTACHED TO THE DISTAL (PATIENT SIDE) PORT OF THE FILTER DETACHED WHILE PRIMING THE LINE; THE NURSE NOTICED BLOOD BACK FLOWED INTO THE LINE AND THE LINE WAS DISCONNECTED AT THE DISTAL END (THE PATIENT SIDE) OF THE FILTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25028838 ↗MalfunctionApr 28, 2026Separation Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS SEPARATED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT BROKE WHEN SHE PRIMED THE NEW LINE WITH A NEW BAG OF AMIODARONE; THE PART OF THE LINE ATTACHED TO THE DISTAL (PATIENT SIDE) PORT OF THE FILTER DETACHED WHILE PRIMING THE LINE; THE NURSE NOTICED BLOOD BACK FLOWED INTO THE LINE AND THE LINE WAS DISCONNECTED AT THE DISTAL END (THE PATIENT SIDE) OF THE FILTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25021735 ↗MalfunctionApr 28, 2026Separation Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS SEPARATED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT BROKE WHEN SHE PRIMED THE NEW LINE WITH A NEW BAG OF AMIODARONE; THE PART OF THE LINE ATTACHED TO THE DISTAL (PATIENT SIDE) PORT OF THE FILTER DETACHED WHILE PRIMING THE LINE; THE NURSE NOTICED BLOOD BACK FLOWED INTO THE LINE AND THE LINE WAS DISCONNECTED AT THE DISTAL END (THE PATIENT SIDE) OF THE FILTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25021691 ↗MalfunctionApr 28, 2026Separation Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS SEPARATED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING . IT WAS REPORTED BY CUSTOMER THAT BROKE WHEN SHE PRIMED THE NEW LINE WITH A NEW BAG OF AMIODARONE; THE PART OF THE LINE ATTACHED TO THE DISTAL (PATIENT SIDE) PORT OF THE FILTER DETACHED WHILE PRIMING THE LINE; THE NURSE NOTICED BLOOD BACK FLOWED INTO THE LINE AND THE LINE WAS DISCONNECTED AT THE DISTAL END (THE PATIENT SIDE) OF THE FILTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25021151 ↗MalfunctionApr 24, 2026Fluid/Blood Leak
IT WAS REPORTED, IV TUBING LEAKING IV FLUIDS DESPITE ROLLER CLAMP BEING CLAMPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24983550 ↗MalfunctionApr 17, 2026Break; Leak/Splash
IT IS REPORTED, ALARIS SMARTSITE, TIP OF THE PLASTIC SPIKE BROKE OFF IN THE OPENING OF THE IV BAG AFTER SPIKING IT CAUSING THE MEDICATION HAVING TO BE THROWN AWAY.
Patient problems: Exposure to Body Fluids
MAUDE report 24911040 ↗
FDA clearance and classification
The FDA regulates product code FPA (set, administration, intravascular) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the Alaris, SmartSite 11532269
Has the Alaris, SmartSite 11532269 been recalled?
Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-3080-2026 (Class I), began on Aug 19, 2026: Due to discrepancies in specific attributes in the performance data for infusion sets, which may lead to inappropriate pump performance. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Alaris, SmartSite 11532269?
The FDA's MAUDE database lists 333 adverse event reports for this device, received between Aug 1, 2018 and Sep 29, 2026. In those reports, the most common specific device problems are “Leak/Splash” (183), “Fluid/Blood Leak” (38) and “Complete Blockage” (24); the most common patient problems are “Chemical Exposure” (5), “Blood Loss” (4) and “Hypersensitivity/Allergic reaction” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Alaris, SmartSite 11532269?
None of the reports are classified as deaths, 6 as injuries and 326 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Alaris, SmartSite 11532269, and is it still sold?
The FDA's device database lists BD Switzerland Sàrl as the company responsible for it (brand Alaris, SmartSite, model 11532269). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death14,517 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall10,688 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths7,526 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths6,716 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths6,585 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,659 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death3,850 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular18 deaths3,660 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page