Skip to content

Alaris, SmartSite 2403-0007Recalls & reported problems

BD Switzerland Sàrl · Pump Infusion Set, Half Set, Y-Site 220 cm (87 in) · Class 2

All models of Alaris, SmartSite (160)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885403232473), received between Jul 8, 2021 and Jul 8, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Free or Unrestricted Flow” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 8, 2025Class IOpen, Classified

    Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

    Z-2180-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Oct 11, 2023Class IIOpen, Classified

    Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).

    Z-0349-2024 · Root cause: Error in labeling · Recall details · FDA record ↗

Most reported problems

Device problems

  • Free or Unrestricted Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 8, 2021Free or Unrestricted Flow; Free or Unrestricted Flow

    IT WAS REPORTED THAT THE AS LVP HALF SS HAD AIR BUBBLES IN THE LINE WHILE INFUSING BLINATUMOMAB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "RN CALLED UNIT QI NURSE AND EDUCATOR INTO PT ROOM DUE TO PUMP ALARMING " AIR IN LINE". PT WAS INFUSING BLINATUMOMAB THROUGHT CURLIN TUBING CONNECTED TO AN ALARIS 1/2 SET. PER RN, PUMP HAD BEEN BEEPING FREQUENTLY SINCE LAST EVENING. ON INSPECTION OF THE TUBING, A FEW BUBBLES WERE SEEN BELOW THE ALARIS PUMP AND ONE SMALL ONE ABOVE. NO AIR WAS NOTICED IN THE CURLIN TUBING. WHEN THE "RESTART" BUTTON WAS PUSHED THE INFUSION WOULD CONTINUE TO INFUSE (WITHOUT THE BEEP FOR A WHILE). THE PHYSICIAN WAS NOTIFIED AND PT TO BE MOVED TO THE CURLIN HOME PUMP ONCE DISCHARGE WAS CONFIRMED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12139690 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page