Alaris, SmartSite 10015861ARecalls & reported problems
BD Switzerland Sàrl · Pump Infusion Set, Bag Access, Non-Vented Low Sorbing Tubing (PE Lined) Back Check Valve, 3 Y-Sites 323 cm (127 in) · Class 2
All models of Alaris, SmartSite (160)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This BD Alaris SmartSite infusion set is used with an infusion pump. Many of the 81 reports describe fluid leaking, connections coming apart and tubing cracking. The reports record no deaths and no injuries, and reports fell from 35 in the prior year to 14 in the latest year. There are 2 open recalls: one for mismatched performance data and one for sets labeled DEHP-free that contain DEHP.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 80
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 80 adverse event reports that list this exact device (DI 10885403232626), received between May 21, 2019 and Sep 24, 2026 — classified as 80 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (35 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 19, 2026Class IOpen, Classified
Due to discrepancies in specific attributes in the performance data for infusion sets, which may lead to inappropriate pump performance.
Z-3080-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Oct 11, 2023Class IIOpen, Classified
Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
Z-0349-2024 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash35
- Connection Problem12
- Defective Component6
- Fluid/Blood Leak6
- Material Separation5
- Complete Blockage4
- Backflow2
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions74
- No Information3
- No Consequences Or Impact To Patient2
- Chemical Exposure1
Recent adverse event reports
The 25 most recent of 80 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Excess Flow or Over-Infusion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889796 ↗MalfunctionJul 6, 2026Crack; Leak/Splash; Material Puncture/Hole
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD DAMAGED / DEFECTIVE TUBING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAD A THIRD INCIDENT WITH BD TUBING FOR THE ALARIS PRODUCTS. THE FIRST TWO WERE CRACKING BACK IN (B)(6) 2026 AND TODAY (B)(6) 2026. OUR HEMATOLOGY/ONCOLOGY OUTPATIENT TEAM IDENTIFIED DURING A CHEMOTHERAPY INFUSION TO A PATIENT THAT A SMALL PIN HOLE IN THE TUBING WAS SEEN WITH CHEMOTHERAPY DRIPPING OUT OF THE TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25717235 ↗MalfunctionJun 3, 2026Material Deformation
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD A BULGE / BALLOON IN SILICONE SEGMENT / PUMP SEGMENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: I AM REPORTING AN INCIDENT WE EXPERIENCED WITH A BD ALARIS PUMP INFUSION SET REF (B)(4). WE ONLY USED ONE TUBING SET. UNFORTUNATELY, WE DO NOT HAVE THE SAVED PACKAGING TO REPORT THE SPECIFIC LOT NUMBER. WE HAVE NARROWED THE LOT DOWN TO EITHER (10)25125451 OR (10)25055230. ADMINISTRATION SET ATTACHED TO BASE SOLUTION BAG AND PRIMED WITH BASE SOLUTION. CHEMOTHERAPY COMPOUNDED APPROPRIATELY. DOSE DELIVERED TO NURSING UNIT VIA TEAM MEMBER HAND-OFF. NURSING ADMINISTERED DOSE VIA APPROPRIATE ALARIS PUMP SET UP PROCEDURES. SEVERAL HOURS AFTER ADMINISTRATION INITIATED PUMP ALARM SOUNDED ALERTING NURSE OF RATE FLOW ERROR. UPON EVALUATION, THE NURSE DISCOVERED BULGING BELOW UPPER PUMP FITMENT IN PUMP SEGMENT LINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25381883 ↗MalfunctionApr 27, 2026Leak/Splash; Material Integrity Problem
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25002569 ↗MalfunctionMar 26, 2026Complete Blockage
IT IS REPORTED, BD ALARIS PUMP INFUSION SET, WHEN LOADED INTO ALARIS PUMP MODULE, MODULES BEEPING OCCLUDED PATIENT SIDE. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24700335 ↗MalfunctionMar 23, 2026Material Integrity Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS DAMAGED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT TUBING DEFORMED, UNABLE TO PRIME DUE TO THE HOLE THAT IS THERE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24662641 ↗MalfunctionMar 10, 2026Material Separation; Air/Gas in Device
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BD ALARIS PUMP INFUSION SET (REF (B)(4)) CONTAINING CHEMOTHERAPY CAME APART WHEN USER ATTEMPTED TO DISLODGE AIR BUBBLES FROM LINE (SOFT SECTION THAT IS PLACED INTO THE ALARIS PUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24559753 ↗MalfunctionMar 10, 2026Leak/Splash; Separation Problem
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24559408 ↗MalfunctionMar 4, 2026Leak/Splash; Material Separation
MalfunctionMar 4, 2026Leak/Splash; Material Separation
MDR MADE IN ERROR WITH INCORRECT GRID UPDATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24508532 ↗MalfunctionFeb 24, 2026Fluid/Blood Leak
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS DAMAGED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY THE CUSTOMER THAT LEAKAGE WAS OBSERVED NEAR THE PROXIMAL PORT IMMEDIATELY AFTER PRIMING THE TUBING WITH MEDICATION, RESULTING IN MEDICATION LOSS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24434801 ↗MalfunctionJan 15, 2026Material Puncture/Hole
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD DAMAGED / DEFECTIVE TUBING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE IS A SMALL PINHOLE LEAK IN THE LOW SORB TUBING RIGHT AROUND WHERE THE HARD PLASTIC COVERING IS LOCATED BELOW ONE OF THE CLAMPS. UNFORTUNATELY, THE SPILL WAS DISCOVERED DURING ADMINISTRATION TO THE PATIENT. THE PATIENT IS FINE. WE HAVE THE BAG IN THE PHARMACY NOW; IT'S FILLED WITH A MEDICATION CALLED KADCYLA. ABOUT 5-10MLS WAS LOST DURING THE SPILL, SO THE BAG IS RELATIVELY FULL WITH DRUG. I CANNOT TELL YOU WHAT THE LOT NUMBER IS UNFORTUNATELY. I CAN ONLY SAY THAT IT IS EITHER LOT (10) 25065355 OR (10)2505502.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24083375 ↗MalfunctionDec 16, 2025Backflow; Obstruction of Flow; Backflow; Obstruction of Flow
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET HAD FLOW ISSUES. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING: IT WAS REPORTED BY THE CUSTOMER THAT THE SECONDARY MEDICATION OF IRON SUCROSE WITH THE PROPER HEAD HEIGHT OF THE SECONDARY AND PRIMARY BAG LOWER RUNNING 0.9% NORMAL SALINE. CHECK VALVE FAILED ON THE PRIMARY INFUSION LINE ABOUT 1/2 WAY THROUGH INFUSION WITH IRON SUCROSE MIGRATING UP TO THE FLUSH BAG. PRIMARY PUMP INFUSION SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23824378 ↗MalfunctionOct 27, 2025Product Quality Problem; Product Quality Problem
MalfunctionOct 27, 2025Leak/Splash; Leak/Splash
MalfunctionOct 19, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT THE DRUG LEAKED FROM THE CONNECTION BETWEEN THE 50ML TEXIUM SYRINGE AND THE PORT ON THE PRIMARY TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23333978 ↗MalfunctionOct 15, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TUBING WAS LEAKING BELOW THE Y-SITE. NO ADVERSE EVENTS OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23297228 ↗MalfunctionSep 9, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET HAD CONNECTION PROBLEM THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT UPON ATTEMPTING TO HANG MEDICATION, TRAZIMERA, SPIKE OF TUBING BECAME DISCONNECTED FROM MEDICATION BAG, ESTIMATED 100ML OF MEDICATION SPILLED UNTO CHAIR, PUMP, AND FLOOR, AREA CLEANED PER PROTOCOL
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23001304 ↗MalfunctionAug 26, 2025Connection Problem; Connection Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS SEPARATED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT THE SPIKE FELL OUT OF THE BAG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22895664 ↗MalfunctionAug 8, 2025Nonstandard Device; Defective Component; Nonstandard Device; Defective Component
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE PUMP INFUSION SET WAS DAMAGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER WAS TRYING TO INFUSE ZENECAR. A PRE-MED TO CHEMO THAT ONLY HAS A 30 MINUTE EXPIRATION TIME. THE DRIP CHAMBER KEPT COLLAPSING ON ITSELF AND NOT ALLOWING THE INFUSION TO FLOW. THE RATE WAS 999.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22759011 ↗MalfunctionJul 31, 2025Obstruction of Flow; Obstruction of Flow
MalfunctionJul 21, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS DAMAGED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING: IT WAS REPORTED BY CUSTOMER THAT THE TUBING BROKE AND SPRAYED CHEMO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22559678 ↗MalfunctionJul 14, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT OBSERVED TUBING HAVE A FINE PINHOLE LEAK FROM WHICH THE TREATMENT WAS LEAKING. UPON FURTHER INSPECTION, THE TREATMENT HAD LEAKED/LEFT A TRAIL FROM THE CHEMO MED ROOM DOWN HALL TO PATIENT'S ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22495667 ↗MalfunctionJul 14, 2025Connection Problem; Connection Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET SEPARATED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER. IT WAS REPORTED BY THE CUSTOMER THAT FAULTY CONNECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22494915 ↗MalfunctionJul 14, 2025Connection Problem; Connection Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET SEPARATED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER. IT WAS REPORTED BY THE CUSTOMER THAT FAULTY CONNECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22494913 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page