Alaris, VersaSafe 10796814Recalls & reported problems
BD Switzerland Sàrl · Extension Set, 2 Split septum ports 103 cm (41 in) · Class 2
All models of Alaris, VersaSafe (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885403234330), received between Apr 20, 2022 and May 5, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Separation2
- Leak/Splash1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2023Leak/Splash; Material Separation; Leak/Splash; Material Separation
IT WAS REPORTED THAT THE BD ALARIS VERSASAFE EXTENSION SET EXPERIENCED DISCONNECTION AND LEAKED. THIS IS 2 OF 2 RELATED EVENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: POSTED BY USER, INTAKE ON 4/12/2023 12:48:21 AM INTAKE NOTE AFTER INSERTING A PIV AND FLUSHING THROUGH THE SLIP-TIP T-CONNECTOR, IT WAS ATTACHED AND SECURED WITH ADDITIONAL TAPE AT THE CONNECTION. APPROX 5 MINS LATER, BLOOD WAS NOTED ON THE ARMBOARD. UPON FURTHER ASSESSMENT, IT WAS DETERMINED THAT THE T-CONNECTOR HAD MIGRATED OUT SLIGHTLY AND BLOOD WAS BACKFLOWING OUT OF THE CATHETER. THE T-CONNECTOR WAS TIGHTENED AND FLUSHED, HOWEVER WHILE RETAPING, IT COULD BE SEEN FLOATING BACK OUT OF THE CATHETER. AT THIS TIME, THE NNP WAS AT BEDSIDE AND ASKED THIS RN TO UTILIZE OUR REGULAR SCREW ON TYPE OF T-CONNECTOR. THE T-CONNECTOR WAS CHANGED OUT AND THE PIV RETAPED WITH NO FURTHER ISSUES OF BLOOD LEAKING OUT. HAD I NOT BEEN AT BEDSIDE WHEN THIS HAPPENED, THE INFANT COULD HAVE LOST A LOT OF BLOOD DURING THE TIME THAT WOULD PASS BEFORE MY NEXT PIV CHECK 55 MINS LATER. IT IS A SAFETY ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16882186 ↗MalfunctionMay 5, 2023Material Separation; Material Separation
IT WAS REPORTED THAT THE BD ALARIS VERSASAFE EXTENSION SET EXPERIENCED DISCONNECTION. THIS IS 1 OF 2 RELATED EVENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: POSTED BY USER, INTAKE ON (B)(6) 2023 2:20:05 PM. INTAKE NOTE: RN OBTAINED PIV AND HAD STARTED A BLOOD TRANSFUSION. ORDERS RECEIVED TO RETAPE THE ETT, AS RT WAS POSITIONING PATIENT TO RETAPE, RT NOTICED THE "SLIP-TIP" HAD DETACHED FROM THE CATHETER HUB. ALSO, TO NOTE, PATIENT HAD RECEIVED A BOLUS OF AN ORDERED SEDATION MEDICATION AND IT HAD NOT TAKEN EFFECT YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16882182 ↗MalfunctionApr 20, 2022Separation Problem; Separation Problem
IT WAS REPORTED THAT THE BD ALARIS VERSASAFE EXTENSION SET EXPERIENCED COMPONENT SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RN NOTICED BLOOD BACKING UP IN THE TUBING AND THAT THE NEW KIND OF IV TUBING WE ARE USING BROKE. THE DISCONNECTION WAS NOT AT A SITE THAT NORMALLY COMES APART. IT WAS WHERE A NEEDLESS HUB AND THE TUBING MEET. THE TUBING SLIPPED APART.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14164119 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗