Alaris, SmartSite 10813621Recalls & reported problems
BD Switzerland Sàrl · Pump Infusion Set, Back Check Valve, 3 Ganged 4-Way Stopcocks, 4 Y-Sites 404 cm (159 in) · Class 2
All models of Alaris, SmartSite (160)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10885403234514), received between Jan 13, 2020 and Jun 3, 2026 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (5 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 19, 2026Class IOpen, Classified
Due to discrepancies in specific attributes in the performance data for infusion sets, which may lead to inappropriate pump performance.
Z-3080-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Jul 8, 2025Class IOpen, Classified
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Z-2180-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash5
- Material Separation4
- Material Twisted/Bent2
- Infusion or Flow Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Information3
- No Consequences Or Impact To Patient2
- No Patient Involvement1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 3, 2026Material Twisted/Bent
IT WAS REPORTED THAT THE BD ALARIS PUMP MODULE SMARTSITE INFUSION SET HAD A KINK / BEND. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE ARE HAVING ISSUES AGAIN WITH SET 10813621, LOT # 26015041 & 25095372. TUBING ALL SEEMS TO BE KINKED/BENT AT THE 3-PORT MANIFOLD WHERE THE TUBING GOES INTO THE MANIFOLD. IT APPEARS THAT THE TUBING IS BENDING BECAUSE THE PACKAGE ISN¿T LONG ENOUGH TO ACCOMMODATE THE LENGTH OF THE TUBING. WERE THERE ANY ADVERSE EVENTS OR PATIENT HARM ASSOCIATED WITH THIS OCCURRENCE? NO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25378242 ↗MalfunctionJun 3, 2026Material Twisted/Bent
IT WAS REPORTED THAT THE BD ALARIS PUMP MODULE SMARTSITE INFUSION SET HAD A KINK / BEND. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE ARE HAVING ISSUES AGAIN WITH SET 10813621, LOT # 26015041 & 25095372. TUBING ALL SEEMS TO BE KINKED/BENT AT THE 3-PORT MANIFOLD WHERE THE TUBING GOES INTO THE MANIFOLD. IT APPEARS THAT THE TUBING IS BENDING BECAUSE THE PACKAGE ISN¿T LONG ENOUGH TO ACCOMMODATE THE LENGTH OF THE TUBING. WERE THERE ANY ADVERSE EVENTS OR PATIENT HARM ASSOCIATED WITH THIS OCCURRENCE? NO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25378237 ↗MalfunctionMay 11, 2026Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE PUMP INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT 04/19 WHILE DOING MY SAFETY CHECKS BEFORE STARTING CARES, THEY NOTICED FLUID ON THE FLOOR UNDERNEATH THE BED, AND IT WAS VERY STICKYLIKE TPN. THEY SAW IT DRIPPING FROM THE END OF THE BED. AFTER TRACING MY LINES, THEY FOUND DROPS ON THE IV TUBING AND AROUND THE JUNCTION WHERE THE MED PORT ON THE IV PUMP INFUSION SET IS LOCATED. THEY SHOWED ACTION RN A. XXXXX AND NOTIFIED THE NNP A. XXXXX. GRABBED A BLOOD GLUCOSE WHICH WAS 90 AND THE NNP ORDERED NEW FLUIDS TO HANG. THEY PLACED A STICKER WHERE THEY NOTICED THE LEAKING OCCURRING ON TO THE TUBING EXTENSION SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25144070 ↗MalfunctionMar 16, 2026Infusion or Flow Problem
IT WAS REPORTED BY CUSTOMER THEY EXPERIENCED FLOW ISSUES WITH ALARIS SMARTSITE AND RE-STUCK A PATIENT UNDER THE ASSUMPTION THE IV WAS FAILING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24607727 ↗MalfunctionFeb 24, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS PUMP MODULE SMARTSITE INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT 10813621 SET DOESN'T HAVE THE ANTIREFLUX VALVES AND FLUIDS ARE LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21455118 ↗MalfunctionJan 30, 2024Leak/Splash; Material Separation; Leak/Splash; Material Separation
NO ADDITIONAL INFORMATION WAS PROVIDED. MATERIAL#: 10813621, LOT#: UNKNOWN. IT WAS REPORTED BY THE CUSTOMER THAT THE ONE OF THEM LEAKED FROM THE CLEAR PLASTIC PART AND THE CLEAN SPLIT IN HALF. VERBATIM: 2 EVENT WITH TUBING 10813621 IN THE PAST MONTH. ONE OF THEM LEAKED FROM THE CLEAR PLASTIC PART AND THE CLEAN SPLIT IN HALF. SEE PICTURES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18610061 ↗MalfunctionJan 30, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD ALARIS PUMP MODULE SMARTSITE INFUSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: VERBATIM: 2 EVENT WITH TUBING 10813621 IN THE PAST MONTH. ONE OF THEM LEAKED FROM THE CLEAR PLASTIC PART AND THE CLEAN SPLIT IN HALF. SEE PICTURES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18609833 ↗MalfunctionFeb 17, 2020Material Separation; Material Separation
IT WAS REPORTED THAT THERE WAS A TUBING SEPARATION. THE EVENT OCCURRED IN THE OPERATING ROOM AS IT WAS BEING "REMOVED". IT WAS NOTED THAT PLASMA-LYTE WAS INFUSING KVO AT THE TIME OF THE EVENT. PER THE REPORTER; "TRIED TO REMOVE A PIECE OF TAPE THAT WAS WRAPPED AROUND THE TUBING AND THE TUBING WAS SEVERED BY THE TAPE". IT WAS CONFIRMED THAT THE TAPE WAS NOT PRESENT UPON OPENING OF PACKAGING, AND WAS USED TO "SECURE". THERE WAS NO HARM TO THE USER, AS "THE TUBING WAS CLAMPED PROXIMALLY AND DISTALLY AFTER IT SEPARATED". IT WAS FURTHER CONFIRMED DURING FOLLOW UP THAT PATIENT CARE WAS NOT DELAYED AND THAT THIS EVENT DID NOT CONTRIBUTE TO, OR RESULT IN A SERIOUS ADVERSE IMPACT TO PATIENT. HOWEVER; IT WAS ALSO NOTED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Information; No Patient Involvement
MAUDE report 9717601 ↗MalfunctionJan 26, 2020Material Separation; Material Separation
IT WAS REPORTED THAT THE ALARIS INFUSION TUBING SPONTANEOUSLY SEPARATED.
Patient problems: No Information
MAUDE report 9630508 ↗MalfunctionJan 26, 2020Material Separation; Material Separation
IT WAS REPORTED THAT THE ALARIS INFUSION TUBING SPONTANEOUSLY SEPARATED WHILE MEDICATION WAS BEING INFUSED. THE SEPARATION WAS PICKED UP WHEN THE PUMP KEPT ALARMING AIR IN LINE. NO ENTRAINED AIR REACHED THE PATIENT. THERE HAS BEEN NO FURTHER IMPACT TO PATIENT RESPONSE OR ADDITIONAL EVENT INFORMATION MADE AVAILABLE TO DATE.
Patient problems: No Consequences Or Impact To Patient; No Information
MAUDE report 9630496 ↗MalfunctionJan 13, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE STOPCOCKS ¿ARE BACKWARDS¿. IT WAS OBSERVED THAT "THESE ARE NO TAB FOR OFF AND SHORT TABS FOR ON." THE CLINICIAN STATES THAT "UNLESS YOU TURN OFF THE STOPCOCK AFTER SETTING IT UP, YOU¿LL GET SPLASHED WITH BLOOD/FLUID/ETC., WHEN YOU TAKE OFF THE WHITE PORT CAP." IT WAS REPORTED THAT THERE WAS NO USER OR PATIENT HARM. THE EVENT DID NOT RESULT IN A DELAY THAT SIGNIFICANTLY IMPACTED PATIENT OUTCOME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9580867 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page