Alaris, SmartSite 11419365Recalls & reported problems

BD Switzerland Sàrl · Pump Infusion Set, Back Check Valve, Manifold, 3-Way Stopcock, 5 Y-Sites 403 cm (159 in) · Class 2

All models of Alaris, SmartSite (160)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885403234736), received between Dec 13, 2023 and Dec 13, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 19, 2026Class IOpen, Classified

    Due to discrepancies in specific attributes in the performance data for infusion sets, which may lead to inappropriate pump performance.

    Z-3080-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jul 8, 2025Class IOpen, Classified

    Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

    Z-2180-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Excess Flow or Over-Infusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2023Excess Flow or Over-Infusion; Excess Flow or Over-Infusion

    NO ADDITIONAL INFORMATION WAS PROVIDED. MATERIAL#: 11419365, BATCH NUMBER#: 23075082. IT WAS REPORTED BY CUSTOMER THAT IN THE BLDG D R&D LAB, AN INFUSION WITH AN ALARIS SYSTEM SET INCORPORATED WITH A DISTAL MANIFOLD (MODEL: 11419365), RESULTED IN AN OVER INFUSION AS THE INFUSION WAS INITIATED ON 1 OUT OF 3 SETS TESTED. THE INFUSION WAS PROGRAMMED TO RUN AT 0.1 ML/HR FOR 96 HOURS AS STATED IN TEST PROTOCOL (B)(4). VERBATIM#: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. IN THE BLDG D R&D LAB, AN INFUSION WITH AN ALARIS SYSTEM SET INCORPORATED WITH A DISTAL MANIFOLD (MODEL: 11419365), RESULTED IN AN OVER INFUSION AS THE INFUSION WAS INITIATED ON 1 OUT OF 3 SETS TESTED. THE INFUSION WAS PROGRAMMED TO RUN AT 0.1 ML/HR FOR 96 HOURS AS STATED IN TEST PROTOCOL (B)(4). DATE OF EVENT: 25OCT2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18322600 ↗