Alaris, SmartSite 20027ERecalls & reported problems
BD Switzerland Sàrl · Extension Set 0.2 Micron Low Protein Binding Filter Needle-Free Valve · Class 2
All models of Alaris, SmartSite (160)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This BD Alaris SmartSite extension set is a tubing set with a needle-free valve, used to connect IV lines. Many of its 189 reports describe leaks or splashes, blocked flow, and parts coming apart. Most reports describe no effect on the patient. There are 3 injury reports and 1 death report. No recalls name this device. Reports rose from 21 in the year before to 36 in the latest year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 183
- Deaths reported
- 1
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 183 adverse event reports that list this exact device (DI 10885403234859), received between Oct 24, 2018 and Sep 29, 2026 — classified as 1 death, 3 injuries, 179 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (71 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash71
- Fluid/Blood Leak56
- Obstruction of Flow36
- Disconnection35
- Material Separation14
- Complete Blockage12
- Device Contamination with Chemical or Other Material6
- Defective Component4
Patient problems
- No Clinical Signs, Symptoms or Conditions103
- No Consequences Or Impact To Patient63
- No Information10
- No Patient Involvement4
- Insufficient Information3
- Hypoglycemia2
- Anxiety1
- Apnea1
Recent adverse event reports
The 25 most recent of 183 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936979 ↗MalfunctionSep 24, 2026Leak/Splash
It was reported that BD smartsite extension set was leaking.The following information was received by the initial reporter with the following Verbatim.It was reported by customer that while preparing a CSP and priming the line with drug the tubing was found to be leaking out of the 0.2-micron filter. The leaking tubing was identified prior to interaction with a patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889056 ↗MalfunctionSep 17, 2026Leak/Splash
It was reported that BD smartsite extension set was crackedThe following information was received by the initial reporter with the following VerbatimIt was reported by customer that leaking and cracked line.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50797236 ↗MalfunctionSep 3, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50593006 ↗MalfunctionAug 4, 2026Complete Blockage; Connection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50132886 ↗InjuryAug 3, 2026Leak/Splash
It was reported that the BD Alaris SmartSite Extension Set had leakage.The following information was provided by the initial reporter: leaking from extension set.Additional Information Received from the Customer:Please provide the exact event date in MM-DD-YYYY format. If multiple incidents occurred, kindly provide the corresponding event date for each incident. (b)(6)2026.Please provide the Batch# or Lot# number? Unknown.What was the medication/fluid in use at the time of the event? Yes; TPN.Describe any patient¿s harm, injury, complication or negative outcome that occurred as a result of the event. Patient had a low blood glucose and required a D10 bolus and hourly BG checks for several hours. Patient is now back to their baseline and back on their TPN.
Patient problems: Hypoglycemia
MAUDE report 50124253 ↗InjuryAug 3, 2026Leak/Splash
It was reported that the BD Alaris SmartSite Extension Set had leakage.The following information was provided by the initial reporter: leaking from extension setAdditional Information Received from the Customer:Please provide the exact event date in MM-DD-YYYY format. If multiple incidents occurred, kindly provide the corresponding event date for each incident. 6/10/2026.Please provide the Batch# or Lot# number? Unknown.What was the medication/fluid in use at the time of the event? Yes; TPN.Describe any patient¿s harm, injury, complication or negative outcome that occurred as a result of the event. Patient had a low blood glucose and required a D10 bolus and hourly BG checks for several hours. Patient is now back to their baseline and back on their TPN.
Patient problems: Hypoglycemia
MAUDE report 50124251 ↗MalfunctionAug 1, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50108050 ↗MalfunctionJul 10, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25788410 ↗MalfunctionMay 18, 2026Complete Blockage
IT WAS REPORTED THAT BD SMARTSITE SECONDARY SET WAS OCCLUDED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY THE CUSTOMER THAT FLUID WILL NOT FLOW PAST THE DRIP CHAMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25208063 ↗MalfunctionApr 22, 2026Leak/Splash; Separation Problem
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE EXTENSION SET HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FAILURE OF BD EXTENSION SET IDENTIFIED, 2 OCCURRENCES. RN REPORTED LEAKING WHEN EXTENSION SET WAS ATTACHED TO THE FLUSH BAG. THE SECOND THEN CAME APART IN HER HANDS WHEN TRYING TO REPLACE THE 1ST SET, WHICH HAD LEAKED. IT APPEARS THAT THERE IS NO SEALANT PRESENT. PT CODE: 4582. DEVICE CODES: 1250, 2907. REF REPORT: MW5184749.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24957002 ↗MalfunctionApr 22, 2026Leak/Splash
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE EXTENSION SET HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FAILURE OF BD EXTENSION SET IDENTIFIED, 2 OCCURRENCES. RN REPORTED LEAKING WHEN EXTENSION SET WAS ATTACHED TO THE FLUSH BAG. THE SECOND THEN CAME APART IN HER HANDS WHEN TRYING TO REPLACE THE 1ST SET, WHICH HAD LEAKED. IT APPEARS THAT THERE IS NO SEALANT PRESENT. PT CODE: 4582. DEVICE CODES: 1250, 2907. REF REPORT: MW5184749.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24957001 ↗MalfunctionApr 21, 2026Separation Problem
NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24945140 ↗MalfunctionApr 21, 2026Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE EXTENSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT RN REPORTED LEAKING WHEN EXTENSION SET WAS ATTACHED TO THE FLUSH BAG. THE SECOND THEN CAME APART IN HER HANDS WHEN TRYING TO REPLACE THE 1ST SET, WHICH HAD LEAKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24945137 ↗MalfunctionApr 13, 2026Leak/Splash; Material Separation
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE EXTENSION SET HAD SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NURSE WALKED INTO ROOM; THERE WAS A PUDDLE ON THE FLOOR NEXT TO PATIENT. UPON INSPECTION RN REALIZED THAT THE AMIODARONE TUBING WAS WET, UPON INSPECTION THE FILTER HAD BROKEN FROM THE TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24860394 ↗MalfunctionApr 1, 2026Leak/Splash
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24757178 ↗MalfunctionMar 31, 2026Separation Failure; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE EXTENSION SET WAS SEPARATED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY THE CUSTOMER THAT THE TUBING ON ONE SIDE OF THE FILTER DISLODGED FROM THE FILTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24747206 ↗MalfunctionMar 31, 2026Complete Blockage
IT WAS REPORTED THAT BD SMARTSITE SECONDARY SET WAS OCCLUDED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY THE CUSTOMER THAT FLUID WILL NOT FLOW PAST THE DRIP CHAMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24746543 ↗MalfunctionMar 24, 2026Separation Problem
IT WAS REPORTED THAT BD ALARIS SMARTSITE EXTENSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT MARCH 2ND WE HAVE HAD 2 BD 0.2 MICRON LOW PROTEIN BINDING FILTERS (BD SMARTSITE EXTENSION SET) THAT WOULD NOT PRIME/FLUSH AND MARCH 3RD NURSE WAS INFUSING DEXAMETHASONE(NO CYTOTOXICS) THROUGH A BD ALARIS TUBING SET WITH A 0.2 MICRON FILTER (NO PACKAGE BUT SAME PRODUCT AS ABOVE) AND THE IV BEGAN LEAKING AT THE FILTER WHERE THE TUBING CONNECTS TO THE ACTUAL FILTER AND FEBRUARY 20TH SAME FILTER PRODUCT AS ABOVE. WAS LEAKING AND THEN SEPARATED AT THE SPOT WHERE THE TUBING CONNECTS TO THE FILTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24681152 ↗MalfunctionMar 24, 2026Leak/Splash
IT WAS REPORTED THAT BD ALARIS SMARTSITE EXTENSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT (B)(6) WE HAVE HAD 2 BD 0.2 MICRON LOW PROTEIN BINDING FILTERS (BD SMARTSITE EXTENSION SET) THAT WOULD NOT PRIME/FLUSH AND (B)(6) NURSE WAS INFUSING DEXAMETHASONE(NO CYTOTOXICS) THROUGH A BD ALARIS TUBING SET WITH A 0.2 MICRON FILTER (NO PACKAGE BUT SAME PRODUCT AS ABOVE) AND THE IV BEGAN LEAKING AT THE FILTER WHERE THE TUBING CONNECTS TO THE ACTUAL FILTER AND (B)(6) SAME FILTER PRODUCT AS ABOVE. WAS LEAKING AND THEN SEPARATED AT THE SPOT WHERE THE TUBING CONNECTS TO THE FILTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24675148 ↗MalfunctionMar 24, 2026Complete Blockage
IT WAS REPORTED THAT BD ALARIS SMARTSITE EXTENSION SET WAS OCCLUDED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT MARCH 2ND WE HAVE HAD 2 BD 0.2 MICRON LOW PROTEIN BINDING FILTERS (BD SMARTSITE EXTENSION SET) THAT WOULD NOT PRIME/FLUSH AND MARCH 3RD NURSE WAS INFUSING DEXAMETHASONE(NO CYTOTOXICS) THROUGH A BD ALARIS TUBING SET WITH A 0.2 MICRON FILTER (NO PACKAGE BUT SAME PRODUCT AS ABOVE) AND THE IV BEGAN LEAKING AT THE FILTER WHERE THE TUBING CONNECTS TO THE ACTUAL FILTER AND FEBRUARY 20TH SAME FILTER PRODUCT AS ABOVE. WAS LEAKING AND THEN SEPARATED AT THE SPOT WHERE THE TUBING CONNECTS TO THE FILTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24675145 ↗MalfunctionMar 18, 2026Leak/Splash
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24635023 ↗MalfunctionMar 18, 2026Fluid/Blood Leak; Leak/Splash
ADDITIONAL INFORMATION:ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NOCAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN FORMAT DD-MMM-YYYY? (B)(6) 2026.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24632159 ↗MalfunctionMar 12, 2026Obstruction of Flow
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24587272 ↗MalfunctionJan 29, 2026Fluid/Blood Leak
IT IS REPORTED LEAKAGE. DESCRIPTION: BD SMARTSITE EXTENSION SET SHOWED SIGNS OF LEAKAGE AROUND THE EDGES OF THE FILTER DURING AN INFUSION OF PEMBROLIZUMAB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24209461 ↗
Other Set, i.V. Fluid transfer devices
- EXACTAMIX H938740Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall1,350 reports
- EXACTAMIX H938738Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall1,180 reports
- EXACTAMIX H938739Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall972 reports
- EXACTAMIX H938741Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall710 reports
- EXACTAMIX H938737Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall607 reports
- EXACTAMIX H938724Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall513 reports
- EXACTAMIX H938173Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls260 reports
- EXACTAMIX H938176Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls232 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page