MaxPlus MP5314-CRecalls & reported problems
BD Switzerland Sàrl · Pressure rated extension set, with removable clear needleless connector 18 cm (7 in) · Class 2
Quick summary
This BD MaxPlus pressure-rated extension set is a short tubing set with a removable needleless connector used for IV drug administration. The 27 reports mostly describe the tubing coming apart or disconnecting from the IV catheter port, along with leaking at connections; 11 reports came in the past 12 months compared with 7 in the prior 12 months. No deaths or injuries are described, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 10885403236723), received between Aug 3, 2021 and Aug 17, 2026 — classified as 26 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Connection Problem10
- Leak/Splash6
- Separation Problem5
- Activation, Positioning or Separation Problem3
- Complete Blockage1
- Component Missing1
- Fluid/Blood Leak1
- Incomplete or Inadequate Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- Pain4
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 17, 2026Connection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50280932 ↗MalfunctionAug 17, 2026Connection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50280896 ↗MalfunctionAug 17, 2026Connection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50280895 ↗MalfunctionMar 3, 2026Leak/Splash
IT WAS REPORTED THAT BD MAXPLUS PRESSURE RATED EXTENSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT LEAKED AT IV CONNECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24503164 ↗MalfunctionJan 8, 2026Activation, Positioning or Separation Problem
IT WAS REPORTED THAT BD MAXPLUS PRESSURE RATED EXTENSION SET WAS SEPARATED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT HAD TWO PATIENTS OVER THE WEEKEND OF 12/13-12/14 WHOSE J-LOOPS CAME DETACHED FROM THE IV CATHETER PORT WHILE THE PATIENTS WERE IN POSTPARTUM. THIS MORNING 12/15, ONE OF OUR INDUCTIONS ALSO HAD THEIRS BECOME DISCONNECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24016280 ↗MalfunctionJan 8, 2026Activation, Positioning or Separation Problem
IT WAS REPORTED THAT BD MAXPLUS PRESSURE RATED EXTENSION SET WAS SEPARATEDTHE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIMIT WAS REPORTED BY THE CUSTOMER THAT HAD TWO PATIENTS OVER THE WEEKEND OF 12/13-12/14 WHOSE J-LOOPS CAME DETACHED FROM THE IV CATHETER PORT WHILE THE PATIENTS WERE IN POSTPARTUM. THIS MORNING 12/15, ONE OF OUR INDUCTIONS ALSO HAD THEIRS BECOME DISCONNECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24016279 ↗MalfunctionJan 8, 2026Activation, Positioning or Separation Problem
IT WAS REPORTED THAT BD MAXPLUS PRESSURE RATED EXTENSION SET WAS SEPARATED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING: IT WAS REPORTED BY THE CUSTOMER THAT HAD TWO PATIENTS OVER THE WEEKEND ON (B)(6) WHOSE J-LOOPS CAME DETACHED FROM THE IV CATHETER PORT WHILE THE PATIENTS WERE IN POSTPARTUM. THIS MORNING ON (B)(6), ONE OF OUR INDUCTIONS ALSO HAD THEIRS BECOME DISCONNECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24016277 ↗MalfunctionDec 11, 2025Fluid/Blood Leak; Connection Problem; Fluid/Blood Leak; Connection Problem
IT WAS REPORTED THAT BD MAXZERO PRESSURE RATED EXTENSION SET HAD CONNECTION ISSUES. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT THE NURSE ATTEMPTED TO APPLY THE EXTENSION SET PER PROTOCOL. WHEN THEY WAS SECURING THE EXTENSION DEVICE TO THE INT CATHETER, IT WOULD NOT COMPLETELY SEAL. THIS ALLOWED BLOOD TO CONTINUE TO FLOW AND POOL UNDER THE PT'S ELBOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23783985 ↗MalfunctionDec 3, 2025Connection Problem; Connection Problem
IT WAS REPORTED THAT BD MAXPLUS PRESSURE RATED EXTENSION SET HAD CONNECTION ISSUES THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT THEY CANNOT SCREW THE PRODUCT ON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23710733 ↗MalfunctionDec 3, 2025Connection Problem; Connection Problem
IT WAS REPORTED THAT BD MAXPLUS PRESSURE RATED EXTENSION SET HAD CONNECTION ISSUES. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY THE CUSTOMER THAT THEY EXPERIENCED CONNECTION ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23710732 ↗MalfunctionOct 20, 2025Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT THE BD MAXPLUS PRESSURE RATED EXTENSION SET HAD A LOOSE COMPONENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A ROUTINE IV INSERTION WAS COMPLETED ON THIS PATIENT AT 2125 USING THE 'BD MAXPLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE CLEAR NEEDLELESS CONNECTOR' TO CONNECT TO THE INT. I PAID EXTRA ATTENTION TO TIGHTENING THE CONNECTOR TO THE INT BECAUSE THE RAD TECH WAS AT BEDSIDE WAITING TO TAKE THE PATIENT FOR A CT SCAN WITH CONTRAST. IT IS COMMON FOR THESE EXTENSION SETS TO COME LOOSE DURING CONTRAST ADMINISTRATION IN CT SCAN. TECH CALLED ME AFTER PERFORMING HER TEST INJECTION FOR THE CT SCAN WITH THE AUTO INJECTOR BECAUSE THE PATIENT WAS BLEEDING SIGNIFICANTLY FROM HIS IV. I FOUND THE SCREWING PORTION OF THIS EXTENSION SET COMPLETELY LOOSE AND THE EXTENSION SET WAS OUT OF THE INT HUB ALLOWING THE PATIENT TO FREE BLEED. I REAPPLIED THE EXTENSION SET AND SCREWED IT AS TIGHTLY AS POSSIBLE, CLEANED UP THE PATIENT, AND REAPPLIED A SECURE DRESSING. WHEN RETURNING TO MY DEPARTMENT I RECEIVED ANOTHER CALL FROM RADIOLOGY THAT THE INT IS LEAKING AGAIN WITH THE AUTOINJECTOR TEST FLUSH. I CHANGED THIS EXTENSION SET OUT WITH A NEW ONE AND WE WERE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23337168 ↗MalfunctionOct 16, 2025Connection Problem; Separation Problem; Connection Problem; Separation Problem
IT WAS REPORTED THAT THE BD MAXPLUS PRESSURE RATED EXTENSION SET HAD SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A ROUTINE IV INSERTION WAS COMPLETED ON THIS PATIENT AT 2125 USING THE 'BD MAXPLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE CLEAR NEEDLELESS CONNECTOR' TO CONNECT TO THE INT. I PAID EXTRA ATTENTION TO TIGHTENING THE CONNECTOR TO THE INT BECAUSE THE RAD TECH WAS AT BEDSIDE WAITING TO TAKE THE PATIENT FOR A CT SCAN WITH CONTRAST. IT IS COMMON FOR THESE EXTENSION SETS TO COME LOOSE DURING CONTRAST ADMINISTRATION IN CT SCAN. (B)(6) CALLED ME AFTER PERFORMING HER TEST INJECTION FOR THE CT SCAN WITH THE AUTO INJECTOR BECAUSE THE PATIENT WAS BLEEDING SIGNIFICANTLY FROM HIS IV. I FOUND THE SCREWING PORTION OF THIS EXTENSION SET COMPLETELY LOOSE AND THE EXTENSION SET WAS OUT OF THE INT HUB ALLOWING THE PATIENT TO FREE BLEED. I REAPPLIED THE EXTENSION SET AND SCREWED IT AS TIGHTLY AS POSSIBLE, CLEANED UP THE PATIENT, AND REAPPLIED A SECURE DRESSING. WHEN RETURNING TO MY DEPARTMENT I RECEIVED ANOTHER CALL FROM RADIOLOGY THAT THE INT IS LEAKING AGAIN WITH THE AUTOINJECTOR TEST FLUSH. I CHANGED THIS EXTENSION SET OUT WITH A NEW ONE AND WE WERE ABLE
Patient problems: Pain
MAUDE report 23312269 ↗MalfunctionOct 16, 2025Connection Problem; Separation Problem; Connection Problem; Separation Problem
IT WAS REPORTED THAT THE BD MAXPLUS PRESSURE RATED EXTENSION SET HAD SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A ROUTINE IV INSERTION WAS COMPLETED ON THIS PATIENT AT 2125 USING THE 'BD MAXPLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE CLEAR NEEDLELESS CONNECTOR' TO CONNECT TO THE INT. I PAID EXTRA ATTENTION TO TIGHTENING THE CONNECTOR TO THE INT BECAUSE THE RAD TECH WAS AT BEDSIDE WAITING TO TAKE THE PATIENT FOR A CT SCAN WITH CONTRAST. IT IS COMMON FOR THESE EXTENSION SETS TO COME LOOSE DURING CONTRAST ADMINISTRATION IN CT SCAN. (B)(6) CALLED ME AFTER PERFORMING HER TEST INJECTION FOR THE CT SCAN WITH THE AUTO INJECTOR BECAUSE THE PATIENT WAS BLEEDING SIGNIFICANTLY FROM HIS IV. I FOUND THE SCREWING PORTION OF THIS EXTENSION SET COMPLETELY LOOSE AND THE EXTENSION SET WAS OUT OF THE INT HUB ALLOWING THE PATIENT TO FREE BLEED. I REAPPLIED THE EXTENSION SET AND SCREWED IT AS TIGHTLY AS POSSIBLE, CLEANED UP THE PATIENT, AND REAPPLIED A SECURE DRESSING. WHEN RETURNING TO MY DEPARTMENT I RECEIVED ANOTHER CALL FROM RADIOLOGY THAT THE INT IS LEAKING AGAIN WITH THE AUTOINJECTOR TEST FLUSH. I CHANGED THIS EXTENSION SET OUT WITH A NEW ONE AND WE WERE ABLE
Patient problems: Pain
MAUDE report 23312043 ↗MalfunctionOct 16, 2025Connection Problem; Separation Problem; Connection Problem; Separation Problem
IT WAS REPORTED THAT THE BD MAXPLUS PRESSURE RATED EXTENSION SET HAD SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A ROUTINE IV INSERTION WAS COMPLETED ON THIS PATIENT AT 2125 USING THE 'BD MAXPLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE CLEAR NEEDLELESS CONNECTOR' TO CONNECT TO THE INT. I PAID EXTRA ATTENTION TO TIGHTENING THE CONNECTOR TO THE INT BECAUSE THE RAD TECH WAS AT BEDSIDE WAITING TO TAKE THE PATIENT FOR A CT SCAN WITH CONTRAST. IT IS COMMON FOR THESE EXTENSION SETS TO COME LOOSE DURING CONTRAST ADMINISTRATION IN CT SCAN. (B)(6) CALLED ME AFTER PERFORMING HER TEST INJECTION FOR THE CT SCAN WITH THE AUTO INJECTOR BECAUSE THE PATIENT WAS BLEEDING SIGNIFICANTLY FROM HIS IV. I FOUND THE SCREWING PORTION OF THIS EXTENSION SET COMPLETELY LOOSE AND THE EXTENSION SET WAS OUT OF THE INT HUB ALLOWING THE PATIENT TO FREE BLEED. I REAPPLIED THE EXTENSION SET AND SCREWED IT AS TIGHTLY AS POSSIBLE, CLEANED UP THE PATIENT, AND REAPPLIED A SECURE DRESSING. WHEN RETURNING TO MY DEPARTMENT I RECEIVED ANOTHER CALL FROM RADIOLOGY THAT THE INT IS LEAKING AGAIN WITH THE AUTOINJECTOR TEST FLUSH. I CHANGED THIS EXTENSION SET OUT WITH A NEW ONE AND WE WERE ABLE
Patient problems: Pain
MAUDE report 23312042 ↗MalfunctionOct 16, 2025Connection Problem; Separation Problem; Connection Problem; Separation Problem
IT WAS REPORTED THAT THE BD MAXPLUS PRESSURE RATED EXTENSION SET HAD SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A ROUTINE IV INSERTION WAS COMPLETED ON THIS PATIENT AT 2125 USING THE 'BD MAXPLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE CLEAR NEEDLELESS CONNECTOR' TO CONNECT TO THE INT. I PAID EXTRA ATTENTION TO TIGHTENING THE CONNECTOR TO THE INT BECAUSE THE RAD TECH WAS AT BEDSIDE WAITING TO TAKE THE PATIENT FOR A CT SCAN WITH CONTRAST. IT IS COMMON FOR THESE EXTENSION SETS TO COME LOOSE DURING CONTRAST ADMINISTRATION IN CT SCAN. (B)(6) CALLED ME AFTER PERFORMING HER TEST INJECTION FOR THE CT SCAN WITH THE AUTO INJECTOR BECAUSE THE PATIENT WAS BLEEDING SIGNIFICANTLY FROM HIS IV. I FOUND THE SCREWING PORTION OF THIS EXTENSION SET COMPLETELY LOOSE AND THE EXTENSION SET WAS OUT OF THE INT HUB ALLOWING THE PATIENT TO FREE BLEED. I REAPPLIED THE EXTENSION SET AND SCREWED IT AS TIGHTLY AS POSSIBLE, CLEANED UP THE PATIENT, AND REAPPLIED A SECURE DRESSING. WHEN RETURNING TO MY DEPARTMENT I RECEIVED ANOTHER CALL FROM RADIOLOGY THAT THE INT IS LEAKING AGAIN WITH THE AUTOINJECTOR TEST FLUSH. I CHANGED THIS EXTENSION SET OUT WITH A NEW ONE AND WE WERE ABLE
Patient problems: Pain
MAUDE report 23312033 ↗MalfunctionAug 25, 2025Material Separation; Material Separation
IT WAS REPORTED THAT MAXPLUS PRESSURE RATED EXTENSION SET WAS DISCONNECTING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT WHEN INJECTING THE CONTRAST WITH THE AUTO INJECTOR THAT MANY TIMES THE EXTENSION SET WAS DISCONNECTING FROM THE IV CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22886523 ↗MalfunctionJan 23, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD MAXPLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE NEEDLELESS CONNECTOR WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT LEAKING AT THE HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21223849 ↗MalfunctionSep 12, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD EXTENSION SET SEPARATED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: VERBATIM: BD MAX PLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE CLEAR NEEDLESS CONNECTOR WOULD NOT CONNECT TO IV HUB CAUSING BLOOD TO LEAK OUT AROUND THE CONNECTOR. ISSUE RESOLVED WHEN CHANGED TO SUB ITEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20201016 ↗MalfunctionJun 10, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD MAXPLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE NEEDLELESS CONNECTOR WAS DAMAGED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. NURSES ARE SAYING THAT THE LL CONNECTOR PIECE THAT TURNS TO ATTACH TO THE IV IS VERY HARD TO TURN THIS RESULTS IN SOMETIMES NOT GETTING IT ATTACHED PROPERLY. THEY THINK IT IS BUT THEN IT LEAKS AND THEY HAVE A BLOODY MESS TO DEAL WITH, EVEN WORSE THE PATIENT SEES IT AND GETS WORRIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19507019 ↗MalfunctionJul 15, 2023Leak/Splash; Incomplete or Inadequate Connection; Leak/Splash; Incomplete or Inadequate Connection
IT WAS REPORTED, THAT DURING USE WITH AN UNSPECIFIED AMOUNT OF BD MAXPLUS¿ PRESSURE RATED EXTENSION SET. WITH REMOVABLE NEEDLELESS CONNECTOR, THERE WAS A POOR CONNECTION AND LEAKAGE OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: OUR SURGERY AREA IS SEEING AN ISSUE WITH THE BD MAXPLUS PRODUCT REF#: MP314-C. SEEMS TO HAVE A LEAK /SEAL ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17330473 ↗MalfunctionJun 7, 2023Component Missing; Component Missing
IT WAS REPORTED THAT THE BD MAXPLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE NEEDLELESS CONNECTOR WAS MISSING COMPONENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MTM2 FOUND EXTENSION SET MISSING ONE OF THE ENDS DURING PRODUCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17081232 ↗MalfunctionMar 9, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE BD MAXPLUS¿ PRESSURE RATED EXTENSION SET WITH REMOVABLE NEEDLELESS CONNECTOR CAUSED BLOOD EXPOSURE TO THE STAFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ANY MEDICAL INTERVENTION/ OR TREATMENT DUE TO REPORTED ISSUE? NO TREATMENT- MORE BLOOD EXPOSURE TO STAFF."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16510610 ↗MalfunctionDec 20, 2022Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE BD MAXPLUS¿ PRESSURE RATED EXTENSION SET WITH REMOVABLE NEEDLELESS CONNECTOR TUBING WAS BLOCKED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE TUBING IS BLOCKED SOMEHOW-BRAND NEW BUT CANNOT GET ANYTHING TO GO THROUGH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16018925 ↗MalfunctionDec 1, 2022Obstruction of Flow
THE TUBING IS BLOCKED SOMEHOW-BRAND NEW BUT CANNOT GET ANYTHING TO GO THROUGH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15892947 ↗MalfunctionNov 7, 2022No Flow
TUBING IS BLOCKED, NO LIQUID WILL FLOW THROUGH.
Patient problems: Insufficient Information
MAUDE report 15742800 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports