Alaris, SmartSite 0.00425Recalls & reported problems
BD Switzerland Sàrl · Gravity Set, Check Valve, 3 Needle-Free Valves, 3 Port Manifold, Vented/Nonvented, 335 cm (132 in) · Class 2
All models of Alaris, SmartSite (160)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10885403240294), received between Oct 12, 2021 and May 1, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Leak/Splash2
- Backflow1
- Complete Blockage1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2023Separation Problem; Separation Problem
IT WAS REPORTED THAT THE BD ALARIS¿ SMARTSITE¿ GRAVITY SET CHAMBER DISLODGED FROM THE TUBING WHEN REMOVING IT FROM THE PACKAGING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "GRAVITY TUBE ISSUE ON (B)(6) 2023 OPENED PACKAGING FOR A GRAVITY SET AND THE CHAMBER DISLODGED FROM THE TUBING... NOT USED ON PATIENT NO DELAY IN CARE"
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16850306 ↗MalfunctionOct 12, 2021Backflow; Complete Blockage; Leak/Splash; Backflow; Complete Blockage; Leak/Splash
IT WAS REPORTED THAT THE GRA 15DP 3PSSMNFLD 3SS CV DEHP FREE TUBING SEPARATED AND LEAKED IV FLUID ONTO THE FLOOR, AND THE PATIENT'S BLOOD BACKED UP INTO THE OPEN-ENDED IV TUBING. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "I THINK WE NEED TO BE CLEAR THAT THIS PROBLEM DOES NOT STEM FROM A USER ERROR. THIS SAME ISSUE HAS OCCURRED TO ME IN MORE THAN ONE CASE, THE MOST RECENT OVER LAST WEEKEND. I INJECTED A MEDICATION INTO ONE OF THE PORTS ON THE MULTI-PORT HUB... I TURNED AROUND TO CHART THE MEDICATION. I TURNED BACK AROUND TOWARDS THE PATIENT TO FIND IV FLUID DUMPING ONTO THE FLOOR AND THE PATIENT¿S BLOOD BACKING UP INTO THE FREE HANGING OPEN ENDED IV TUBING. AGAIN, THIS HAS HAPPENED TO ME MORE THAN ONCE. IT HAS HAPPENED IN THE DEPARTMENT MULTIPLE TIMES. THIS HAS NOT BEEN AN ONGOING ISSUE, SO SOMETHING MUST HAVE CHANGED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12615928 ↗MalfunctionOct 12, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE GRA 15DP 3PSSMNFLD 3SS CV DEHP FREE IV TUBING SEPARATED DUE TO ONE END NOT BEING THREADED OR TIGHTENED. THIS OCCURRED WITH 2 SETS DURING USE, AND THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DR. XXXX WAS ABLE TO SHOW ME ANOTHER IV TUBING SET THAT CAME APART. LOOKS LIKE THE ONE END OF TUBING IS NOT THREADED AND IF NOT TIGHTENED, THE TUBING WOULD EASILY COME APART... ON THE LEFT SIDE, BOTH ENDS OF THE CONNECTION ARE THREADED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12615926 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page