Alaris 30883Recalls & reported problems
BD Switzerland Sàrl · PCA Administration Set, Microbore Tubing Anti-siphon Valve 234 cm (92 in) · Class 2
All models of Alaris (85)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10885403273957), received between Jun 14, 2019 and Aug 6, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Defective Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Component2
- Break1
- Complete Blockage1
- Crack1
- Device Markings/Labelling Problem1
- Disconnection1
- Failure to Prime1
- Human-Device Interface Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2026Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50150729 ↗MalfunctionAug 6, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50150727 ↗MalfunctionJan 9, 2024Crack; Defective Component; Crack; Defective Component
NO NEW INFORMATION: MATERIAL#: 30883 BATCH NUMBER#: 22095083. IT WAS REPORTED BY CUSTOMER THAT THEY HAD FOUR SEPARATE ISSUES: FIRST A DISCONNECTION FROM A PATIENT, FROM THE PCA TUBING, SECOND AND THIRD, TUBING HAD HOLES AND CRACKS IN IT. FOURTH A BREAK IN THE TUBING WHERE IT CONNECTS TO THE SYRINGE. VERBATIM# "WE HAVE HAD FOUR SEPARATE ISSUES: FIRST A DISCONNECTION FROM A PATIENT, FROM THE PCA TUBING, SECOND AND THIRD, TUBING HAD HOLES AND CRACKS IN IT. FOURTH A BREAK IN THE TUBING WHERE IT CONNECTS TO THE SYRINGE" CUSTOMER REPLY: CONFIRM WHETHER THE FOUR INCIDENTS HAPPENED IN THE SAME DATE. IF NOT, PLEASE PROVIDE THE DATES SEPARATELY. HAPPENED ON DIFFERENT DATES, WE WORK 12 HOUR SHIFTS SO TIMES AS ESTIMATES ONLY: 12/5/23 @ 2300- PCA MICRO BORE TUBING FOUND DISCONNECTED FROM SYRINGE, MEDICATION LEAKED ONTO FLOOR. 12/11/23 @ 0900- PCA TUBING MICRO BORE TUBING LEAKING MEDICATION ON FLOOR, DEFECTS FOUND IN PCA SMALL BORE TUBING. 12/12/23 @ 2130 -PCA TUBING MICRO BORE TUBING LEAKING MEDIATION ON FLOOR, DEFECTS FOUND IN PCA SMALL BORE TUBING. 12/14/23 @ 0800-LUER LOCK ON PCA SYRINGE BROKE OFF FROM PCA TUBING. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCUR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18480473 ↗MalfunctionJan 9, 2024Disconnection; Material Separation; Disconnection; Material Separation
NO NEW ADDITIONAL INFORMATION: MATERIAL#: 30883, BATCH NUMBER#: 22095083. IT WAS REPORTED BY CUSTOMER THAT THEY HAD FOUR SEPARATE ISSUES: FIRST A DISCONNECTION FROM A PATIENT, FROM THE PCA TUBING, SECOND AND THIRD, TUBING HAD HOLES AND CRACKS IN IT. FOURTH A BREAK IN THE TUBING WHERE IT CONNECTS TO THE SYRINGE. VERBATIM# "WE HAVE HAD FOUR SEPARATE ISSUES: FIRST A DISCONNECTION FROM A PATIENT, FROM THE PCA TUBING, SECOND AND THIRD, TUBING HAD HOLES AND CRACKS IN IT. FOURTH A BREAK IN THE TUBING WHERE IT CONNECTS TO THE SYRINGE". CUSTOMER REPLY: CONFIRM WHETHER THE FOUR INCIDENTS HAPPENED IN THE SAME DATE. IF NOT, PLEASE PROVIDE THE DATES SEPARATELY. HAPPENED ON DIFFERENT DATES, WE WORK 12 HOUR SHIFTS SO TIMES AS ESTIMATES ONLY: 12/5/23 @ 2300- PCA MICRO BORE TUBING FOUND DISCONNECTED FROM SYRINGE, MEDICATION LEAKED ONTO FLOOR. 12/11/23 @ 0900- PCA TUBING MICRO BORE TUBING LEAKING MEDICATION ON FLOOR, DEFECTS FOUND IN PCA SMALL BORE TUBING. 12/12/23 @ 2130 -PCA TUBING MICRO BORE TUBING LEAKING MEDIATION ON FLOOR, DEFECTS FOUND IN PCA SMALL BORE TUBING. 12/14/23 @ 0800-LUER LOCK ON PCA SYRINGE BROKE OFF FROM PCA TUBING. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18480472 ↗MalfunctionJan 9, 2024Break; Defective Component; Break; Defective Component
IT WAS REPORTED THAT BD ALARIS PCA ADMINISTRATION SET WAS TUBING WAS DAMAGED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE VERBATIM: "WE HAVE HAD FOUR SEPARATE ISSUES: FIRST A DISCONNECTION FROM A PATIENT, FROM THE PCA TUBING, SECOND AND THIRD, TUBING HAD HOLES AND CRACKS IN IT. FOURTH A BREAK IN THE TUBING WHERE IT CONNECTS TO THE SYRINGE"
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18480470 ↗MalfunctionMar 25, 2021Stretched; Obstruction of Flow; Stretched; Obstruction of Flow
IT WAS REPORTED THAT THE PCA ASV MICROBORE TUBING HAD BALLOONED NEAR WHERE IT ENTERED INTO THE CHANNEL, AND THE PUMP WAS STOPPED WITH THE REMAINING "28.6ML" OF "DOXORUBICIN". THIS COMPLAINT WAS CREATED TO CAPTURE THE 4TH OF 6 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "DATE REPORTED: 2/25; EXTENT OF HARM: REACHED THE INDIVIDUAL: BALLOONING OF IV TUBING PRODUCT. ITEM: ALARIS PUMP IV MODULE & BD ALARIS 3-PORT TUBING: ITEM AND LOT NUMBER: BEING INVESTIGATED CURRENTLY. ITEM CURRENTLY BEING REVIEWED BY CLINICAL ENGINEERING ALONGSIDE WITH THE ALARIS CHANNEL. O DOXORUBICIN MEDICATION PUMP CONTINUED TO STATE THAT "PATIENT SIDE WAS OCCLUDED",NO VISIBLE OCCLUSION OR BENT TUBING, ONCE I OPENED THE CHANNEL ON PUMP, VISIBLE BALLON ON TUBING THAT GOES INSIDE CHANNEL. PUMP WAS STOPPED WITH 28.6ML REMAINING OF MEDICATION."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11571255 ↗MalfunctionJan 15, 2021Complete Blockage; Restricted Flow rate; Complete Blockage; Restricted Flow rate
IT WAS REPORTED THAT THE PCA ASV MICROBORE TUBING WOULD NOT FLUSH. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "TUBING WILL NOT FLUSH."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11182501 ↗MalfunctionMay 7, 2020Failure to Prime
NURSE WAS UNABLE TO GET PCA TUBING TO PRIME. MANUFACTURER RESPONSE FOR PCA TUBING, BD PCA ADMINISTRATION SET (PER SITE REPORTER). QUALITY EVENT REPORT EMAILED TO CUSTOMER FEEDBACK <[email protected]>. RETURN EMAIL RECEIVED FROM CUSTOMER ADVOCACY COMPLAINT MANAGEMENT. COMPLAINT FILE (B)(4). PRODUCT RETURNED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10032218 ↗MalfunctionJun 14, 2019Device Markings/Labelling Problem; Human-Device Interface Problem; Device Markings/Labelling Problem; Human-Device Interface Problem
THE PATIENT WAS POST-OPERATIVE FROM BACK SURGERIES. NURSING HAD THOUGHT THE EPIDURAL INFUSION WAS IV TUBING, AND ATTACHED THE EPIDURAL TUBING TO THE PATIENT'S MIDLINE. THERE WAS NO PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8701902 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page