Alaris 30853Recalls & reported problems

BD Switzerland Sàrl · PCA Administration Set, Microbore Tubing, Segment Check Valve, Anti-siphon Valve 178 cm (70 in) · Class 2

All models of Alaris (85)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10885403273995), received between Oct 4, 2019 and Mar 18, 2020 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak8
  • Crack4
  • Material Split, Cut or Torn1

Patient problems

  • No Consequences Or Impact To Patient6
  • No Information2
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2020Crack; Crack

    IT WAS REPORTED THAT TUBING WAS CRACKED. ALTHOUGH REQUESTED, NO ADDITIONAL EVENT AND/OR PATIENT INFORMATION WAS PROVIDED.

    Patient problems: No Information

    MAUDE report 9851335 ↗
  • MalfunctionMar 16, 2020Crack; Crack

    THE SET WAS RECEIVED AS AN UNEXPECTED RETURN WITH NO OTHER INFORMATION PROVIDED. ALTHOUGH REQUESTED, THERE HAS BEEN NO IMPACT TO PATIENT RESPONSE OR ADDITIONAL EVENT INFORMATION MADE AVAILABLE TO DATE.

    Patient problems: No Information

    MAUDE report 9835735 ↗
  • MalfunctionJan 13, 2020Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn

    IT WAS REPORTED THAT TUBING WAS SLICED. THE RN WAS IN THE ICU, SETTING UP THE PCA TUBING. SHE STARTED TO PRIME THE PCA PUMP AND NOTICED THE TUBING WAS LEAKING. THE RN STOPPED THE PUMP AND CHANGED THE TUBING. IT WAS CONFIRMED THAT THERE WAS NO PATIENT INVOLVEMENT OR HARM AS A RESULT OF THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9582242 ↗
  • MalfunctionNov 25, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A PATIENT WAS ON A PCA PUMP. THE PHYSICIAN'S ASSISTANT NOTIFIED THE NURSE THAT THE PCA PUMP TUBING WAS LEAKING ONTO THE FLOOR. THE NURSE ASSESSED THE TUBING LINE AND FOUND THAT THERE WAS A SMALL LEAK IN THE TUBING. THE TUBING WAS REMOVED. IT WAS FURTHER CONFIRMED DURING FOLLOW UP, THAT THERE WAS NO DELAY IN TREATMENT NOR ANY PATIENT HARM OR IMPACT AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9371757 ↗
  • MalfunctionNov 21, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PCA EXT TUBING LEAKED MULTIPLE TIMES. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM. THE EVENT OCCURRED ON BMT . A NOTE ATTACHED TO BAG THAT STATED; "LEAKING, CHANGED MULTIPLE TIME, NEW TUBING LOT OBTAINED FROM ANOTHER UNIT , INFUSING FINE".

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9359311 ↗
  • MalfunctionNov 21, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PCA EXT TUBING LEAKED MULTIPLE TIMES. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM. THE EVENT OCCURRED ON BMT. A NOTE ATTACHED TO BAG THAT STATED; "LEAKING, CHANGED MULTIPLE TIME, NEW TUBING LOT OBTAINED FROM ANOTHER UNIT, INFUSING FINE".

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9359256 ↗
  • MalfunctionNov 21, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PCA EXT TUBING LEAKED MULTIPLE TIMES. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM. THE EVENT OCCURRED IN THE BMT. ( A NOTE ATTACHED TO BAG THAT STATED; "LEAKING, CHANGED MULTIPLE TIME, NEW TUBING LOT OBTAINED FROM ANOTHER UNIT , INFUSING FINE".)

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9358440 ↗
  • MalfunctionNov 21, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PCA EXT TUBING LEAKED MULTIPLE TIMES. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM. THE EVENT OCCURRED IN THE BMT. ( A NOTE ATTACHED TO BAG THAT STATED; "LEAKING, CHANGED MULTIPLE TIME, NEW TUBING LOT OBTAINED FROM ANOTHER UNIT , INFUSING FINE".)

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9358179 ↗
  • MalfunctionOct 22, 2019Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PRIMARY FLUSH TUBING SET LEAKED FROM THE SIDE OF THE BACK CHECK VALVE WHEN THE PCA INFUSION SET WAS ATTACHED TO THE PRIMARY FLUSH. THE CUSTOMER FURTHER STATED THAT THERE WERE NO ADVERSE EFFECTS CAUSED TO THE PATIENT FROM THE EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9220209 ↗
  • MalfunctionOct 4, 2019Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT THE PCA TUBING SET WAS FOUND TO HAVE SMALL CRACKS WHICH CAUSED THE MEDICATION TO LEAK DURING USE ON A PATIENT. THE CUSTOMER FURTHER CONFIRMED DURING FOLLOW UP, THAT THERE WERE NO ADVERSE EFFECTS CAUSED TO THE ADULT PATIENT FROM THE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9156106 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →