BD SmartSite 42521ERecalls & reported problems
BD Switzerland Sàrl · Gravity Set, Check Valve 3 Needle-Free Valves, 4-Way Stopcock, 2 Ganged Vented/ Nonvented, 346 cm (136 in) · Class 2
All models of BD SmartSite (29)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10885403274251), received between Nov 21, 2018 and Dec 19, 2025 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak6
- Disconnection5
- Complete Blockage3
- Leak/Splash3
- Delivered as Unsterile Product1
- Incomplete or Inadequate Connection1
- Loose or Intermittent Connection1
- Restricted Flow rate1
Patient problems
- No Consequences Or Impact To Patient8
- No Clinical Signs, Symptoms or Conditions4
- No Information2
- Hemorrhage/Blood Loss/Bleeding1
- Insufficient Information1
- No Patient Involvement1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 19, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE GRAVITY SET HAD PACKAGE OPEN SEAL / STERILITY CONCERNS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEW PACKAGING WITH THE PERFORATED OPENING. SAYING IT CAN CAUSE CROSS CONTAMINATION IN THE STERILE FIELD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23850844 ↗MalfunctionMay 9, 2024Complete Blockage; Complete Blockage
IT WAS REPORTED THAT BD ALARIS SMARTSITE GRAVITY SET HAD FLOW ISSUES THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: NO FLOW THROUGH SET. WILL HAVE TO GET LOT # FROM (B)(6) DURING YOUR FOLLOW UP. PER (B)(6), 5 OCCURRENCES OVER THE PAST 2 WEEKS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19278614 ↗MalfunctionNov 11, 2023Complete Blockage; Complete Blockage
NO FURTHER INFORMATION WAS PROVIDED. MATERIAL#: 42521E, BATCH#: (10)23069291. IT WAS REPORTED BY CUSTOMER THAT THE TIP OF THE DRIP CHAMBER IS OCCLUDED. THIS HAS HAPPENED ON 2 TUBINGS OF THE SAME LOT. VERBATIM: MATERIAL NO.: 42521E, LOT NO.: (10)23069291. CUSTOMER STATED THAT THE TIP OF THE DRIP CHAMBER IS OCCLUDED. THIS HAS HAPPENED ON 2 TUBINGS OF THE SAME LOT. RCC RECEIVED COMPLAINT VIA PHONE. PIR ATTACHED. ADDITIONAL INFO RECEIVED ON 02-NOV-2023. 1. WHAT IS THE DATE OF THE EVENT? A. MULTIPLE OCCURRENCES. 2. IS THE SAMPLE OR PHOTOGRAPHS AVAILABLE FOR EVALUATION? IF YES, PLEASE PROVIDE SHIPPING ADDRESS FOR SENDING A RETURN LABEL FROM US RCC FOR INVESTIGATION PURPOSES. A. I HAVE THE SAMPLES OF THE TUBING AVAILABLE. (B)(6). 3. WAS THE PRODUCT RECEIVED IN THIS CONDITION? A. YES, THE PRODUCT WAS OCCLUDED COMING OUT OF THE PACKAGE. 4. IF NOT, WHAT WAS THE PRODUCT BEING USED FOR WHEN THE REPORTED ISSUE WAS OBSERVED? A. THE TUBING WAS USED TO HYDRATE A PATIENT BEFORE OR AFTER A SURGICAL PROCEDURE. 5. PLEASE CONFIRM WHETHER OR NOT THERE WAS ANY PATIENT IMPACT. A. NO. 6. WHEN DID THE REPORTED ISSUE HAPPEN? BEFORE USE, DURING USE, AFTER USE? A. THE FIRST INCIDENT WAS REPORTED TO ME ON (B)(6
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18117141 ↗MalfunctionMay 18, 2023Leak/Splash; Separation Problem; Leak/Splash; Separation Problem
IT WAS REPORTED THAT THE BD ALARIS SMARTSITE GRAVITY SET EXPERIENCED COMPONENT SEPARATION AND LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONNECTION POINTS AND NOT TIGHT SO THEY COME APART AND IV FLUIDS AND/OR BLOOD DRIPS FROM THEM WHEN THEY COME APART.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16959376 ↗MalfunctionAug 18, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE GRA 15DP CKV 3SS W/2 4W STPCK DEHP FREE EXPERIENCED COMPONENT SEPARATION, AND LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: STANDARD BD SMARTSITE GRAVITY IV SET HAVE DEFECTIVE STOPCOCKS THAT WILL NOT TIGHTEN. THEY ARE USED AS THE STANDARD IV SRT ONCOLD LINES IN THE OR. DESPITE TIGHTENING THEY WILL STILL BECOME LOOSE AND CAUSE POTENTIAL HARM TO PATIENT. ON THIS SPECIFIC EVENT I HAD TIGHTENED THE STOPCOCKS PRIOR TO PUSHING INDUCTION MEDICATIONS. THEY APPEARED TO BE SECURELY TIGHTENED. WE HAD PREPPED THE L ARM HIDING THE IV SITE AND TUBING IN THE SAME ARM WE WERE PLACING A LARGE BORE MICRO-PUNCTURE CATHETER. WHILE PLACING THE LINE (NEEDLE IN ARM), WE NOTICED THAT THE IV WAS DRIPPING WELL DESPITE THERE BEING A TOURNIQUET ON THE ARM. WE THEN MOVED THE DRAPE AND NOTICED THE STOPCOCKS HAD DISCONNECTED AND INTRAVENOUS FLUID AND APPROXIMATELY 100CC OF PATIENT'S BLOOD WERE LOST FROM THE IV SET DISCONNECTION. THERE ARE MANY OTHER INSTANCES, MOST DON'T GET PUT INTO OUR QMS SYSTEM. I DON'T HAVE THE PARTICULAR LOT NUMBER BUT AS I SAID IT HAPPENS TOO FREQUENTLY TO BE A ONE TIME MANUFACTURING ISSUE. THERE IS SOME ELEMENT OF DESIGN WHICH PREVENTS
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 12339021 ↗MalfunctionJan 15, 2021Complete Blockage; Restricted Flow rate
IT WAS REPORTED THAT THE GRA 15DP CKV 3SS W/2 4W STPCK DEHP FREE TUBING WOULD NOT PRIME DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THESE TUBINGS WOULD NOT PRIME. I CLAMPED THE TUBING, ATTACHED A SYRINGE TO THE MOST PROXIMAL HUB TO THE DRIP CHAMBER AND TRIED TO PULL FLUID FORWARD AND ONLY GOT RESISTANCE. IT APPEARS AS IF THE TUBING IS BLOCKED WHERE IT ATTACHES TO THE DRIP CHAMBER." "THE TUBING HAD BEEN SPIKED INTO AN IV BAG HOWEVER NEVER IN A PATIENT ROOM."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11181343 ↗MalfunctionNov 2, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED THAT GRA 15DP CKV 3SS W/2 4W STPCK DEHP FREE DISCONNECTED AND LOST BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 42521E BATCH NO: UNKNOWN. IT WAS REPORTED THAT A GRAVITY SET BECAME DISCONNECTED WHEN A PATIENT WAS BEING TRANSFERRED, RESULTING IN BLOOD PRODUCT LOST. DR. (B)(6) HAS JUST INFORMED ME OF ANOTHER INCIDENT WITH BD GRAVITY SET. WHEN TRANSPORTING THE PATIENT OVER TO THE PACU, GRAVITY SET BECAME DISCONNECTED AND AS A RESULT, BLOOD PRODUCT WAS LOST. SHE HAS INFORMED ME OF 3 DIFFERENT INSTANCES OF PATIENT CARE CONCERNS DUE TO POOR QUALITY OF PRODUCT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10775341 ↗MalfunctionMar 9, 2020Fluid/Blood Leak; Incomplete or Inadequate Connection; Fluid/Blood Leak; Incomplete or Inadequate Connection
IT WAS REPORTED THAT DURING PRIMING AND PRIOR TO PATIENT USE, A LEAK WAS OBSERVED AS THE DISTAL LUER CONNECTOR OF THE STOPCOCK WOULD NOT TIGHTEN. THE EVENT OCCURRED AT THE REHABILITATION AND ORTHOPEDIC INSTITUTE. AS THIS EVENT OCCURRED DURING PRIMING THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT, AND SUBSEQUENTLY; NO PATIENT HARM OR IMPACT AS A RESULT.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 9806344 ↗MalfunctionFeb 14, 2020Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT THE STOPCOCK DISCONNECTS UNINTENTIONALLY FROM THE SET WHILE AN UNSPECIFIED FLUID WAS INFUSING. IT WAS NOTED THAT THE EVENT CAUSED ISSUES WITH CONTAMINATION AND PATIENT CONFIDENCE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9714699 ↗MalfunctionFeb 14, 2020Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT THE STOPCOCK DISCONNECTS UNINTENTIONALLY FROM THE SET WHILE AN UNSPECIFIED FLUID WAS INFUSING. IT WAS NOTED THAT THE EVENT CAUSED ISSUES WITH CONTAMINATION AND PATIENT CONFIDENCE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9714361 ↗MalfunctionFeb 14, 2020Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT THE STOPCOCK DISCONNECTS UNINTENTIONALLY FROM THE SET WHILE AN UNSPECIFIED FLUID WAS INFUSING. IT WAS NOTED THAT THE EVENT CAUSED ISSUES WITH CONTAMINATION AND PATIENT CONFIDENCE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9714274 ↗MalfunctionFeb 14, 2020Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT THE STOPCOCK DISCONNECTS UNINTENTIONALLY FROM THE SET WHILE AN UNSPECIFIED FLUID WAS INFUSING. IT WAS NOTED THAT THE EVENT CAUSED ISSUES WITH CONTAMINATION AND PATIENT CONFIDENCE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9714272 ↗MalfunctionNov 27, 2019Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT THE CONNECTIONS IS COMING APART WHICH IS CAUSING SAFETY CONCERNS FOR THE PROVIDERS. THE ANESTHESIOLOGIST THOUGHT HE WAS ADMINISTERING TO THE PATIENT BUT WAS ACTUALLY GOING ON THE FLOOR. STAFF IS HAVING TO TIGHTEN EACH CONNECTION WHICH IS VERY TIME CONSUMING. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Information
MAUDE report 9383512 ↗MalfunctionNov 21, 2019Disconnection; Disconnection
IT WAS REPORTED THAT THE CONNECTIONS COMING ARE LOOSE WHICH IS CAUSING SAFETY CONCERNS FOR THE PROVIDERS. STAFF IS HAVING TO TIGHTEN EACH CONNECTION WHICH IS VERY TIME CONSUMING. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Information
MAUDE report 9353564 ↗MalfunctionNov 21, 2018Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED A GENERAL COMPLAINT WITH NO SPECIFICS THAT THE 3 WAY STOP COCK CONNECTIONS ON INFUSION SETS ARE RANDOMLY LOOSE FROM THE TUBING ALLOWING FOR LEAKAGE OF IV FLUID AND BLOOD. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8094533 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page