Skip to content

AirLife AH180Recalls & reported problems

Carefusion Corporation · AirLife Adult Heated Wire Circuit Dual-limb, Dual-heated Circuit · Class 2

All models of AirLife (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885403416682), received between Aug 11, 2022 and Aug 11, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Device Alarm System” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Alarm System1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 11, 2022Device Alarm System; Device Alarm System

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE ADULT HEATER DUAL LIMB CIRCUIT CAUSED A VENTILATOR'S HEATER TO ALARM WHILE CONNECTED TO A PATIENT. THE END-USER ATTEMPTED TO CONNECT THE CIRCUIT TO OTHER HEATERS, AND ALL EXPERIENCED THE SAME ALARM ISSUE. THE END-USER THEN REMOVED THE CIRCUIT FROM THE PATIENT AND REPLACED THE CIRCUIT WITH A NEW ONE, AND THE ISSUE WAS RESOLVED. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15214792 ↗

Other Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer devices

All Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page