AirLifeRecalls & reported problems across all models
Carefusion Corporation · 23 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 41
- Deaths reported
- 0
- Injuries reported
- 15
- Recalls
- 0
We found 23 models of AirLife by Carefusion Corporation in the FDA's device registry (GUDID). Together they are named in 41 adverse event reports in the FDA's MAUDE database, received between Jan 4, 2019 and Nov 27, 2023 — classified as 15 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported AirLife problems
Device problems
- Contamination /Decontamination Problem6
- Defective Device6
- Infusion or Flow Problem4
- Disconnection3
- Leak/Splash3
- No Pressure3
- Defective Component2
- Fire2
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Consequences Or Impact To Patient5
- Low Oxygen Saturation4
- No Known Impact Or Consequence To Patient4
- No Information2
- Airway Obstruction1
- Bradycardia1
- Fever1
AirLife models
23 models, most reported first. Each model has its own page with its reports, recalls and device details.
- AirLife 2D0737HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)6 reports
- AirLife 2K8004Cardiopulmonary Resuscitation Aid Kit4 reports
- AirLife 002620HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)3 reports
- AirLife CC10NEBULIZER (DIRECT PATIENT INTERFACE)3 reports
- AirLife 002002HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)2 reports
- AirLife 002502NEBULIZER (DIRECT PATIENT INTERFACE)2 reports
- AirLife 4681-504HEATER, BREATHING SYSTEM W/WO CONTROLLER (NOT HUMIDIFIER OR NEBULIZER2 reports
- AirLife 3D0868NEBULIZER (DIRECT PATIENT INTERFACE)2 reports
- AirLife CK0010Humidifier nebulizer kit2 reports
- AirLife 5257NEBULIZER (DIRECT PATIENT INTERFACE)2 reports
- AirLife 002006HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)1 report
- AirLife 002433NEBULIZER (DIRECT PATIENT INTERFACE)1 report
- AirLife 2K8805Cardiopulmonary Resuscitation Aid Kit1 report
- AirLife 2K8038Cardiopulmonary Resuscitation Aid Kit1 report
- AirLife 2K8040Cardiopulmonary Resuscitation Aid Kit1 report
- AirLife 2K8085ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 report
- AirLife FN2438NEBULIZER (DIRECT PATIENT INTERFACE)1 report
- AirLife 2K7000Ventilator, Emergency, Manual (Resuscitator)1 report
- AirLife CK0005Humidifier nebulizer kit1 report
- AirLife 5504BGL1VENTILATOR, NON-CONTINUOUS (RESPIRATOR)1 report
- AirLife NIV041SVENTILATOR, NON-CONTINUOUS (RESPIRATOR)1 report
- AirLife AH102Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer1 report
- AirLife AH180Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer1 report
Recent AirLife adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryNov 27, 2023Defective Component; Defective Component
IT WAS REPORTED TO VYAIRE MEDICAL THAT PRIOR TO EXTUBATION, THE RESPIRATORY THERAPIST (RT) WAS SETTING UP OXYGEN TO MATCH WHAT WAS ON THE VENT. SHORTLY AFTER, A LOUD WHISTLE WAS HEARD FROM THE SAFETY PRESSURE VALVE. RT CHECKED EQUIPMENT AND NOTED THERE WAS NO HOLE ON 002006 - HUMIDIFIER EMPTY 6 PSI BLUE CAP 50/CS LEADING TO THE CANNULA TO DELIVER OXYGEN. WHEN RT GRABBED A NEW ONE FROM THE RESPIRATORY CLOSET, HE NOTED IT HAD THE SAME DEFECT. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM/INJURY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18212236 ↗MalfunctionAug 2, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE 002002 - NEBULIZER WITH AIR ENTRAINMENT WAS LEAKING FROM BOTTLE. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM/INJURY ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17443036 ↗MalfunctionJul 28, 2023Component Missing; Component Missing
IT WAS REPORTED TO VYAIRE MEDICAL THAT ONE OF THE AH102 - CIRCUIT, ADULT, HEATED SINGLE LIMB WAS MISSING THE TIP OR THE ATTACHMENT. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM/INJURY ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17419972 ↗MalfunctionMar 20, 2023Fitting Problem; Fitting Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE HUMIDIFIER KIT 500ML 12/CS ADAPTER DOES NOT FIT PROPERLY INTO THE FLOWMETER. DUE TO THIS, A LEAKAGE OCCURRED WHICH CAUSED A LOW OXYGEN SATURATION. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: Low Oxygen Saturation
MAUDE report 16573827 ↗MalfunctionFeb 6, 2023Unexpected Therapeutic Results; Unexpected Therapeutic Results
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE 2433 - MM10 SVN W/ADULT MASK & 7FT TBG 50/CS DOESN'T WORK; IT DOESN'T MIST THE MEDICATION. THIS HAPPENED TWICE IN THE ICU (INTENSIVE CARE UNIT). THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM/INJURY ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16306002 ↗MalfunctionOct 1, 2022Pressure Problem; Pressure Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE BABY SAFE HYP INF, .5L, 60CMH20, MANO, NRPR DOES NOT PROVIDE ACCURATE PRESSURE. THE PRESSURE NEEDED TO BE INCREASED TO MAINTAIN THE FLOW TO THE BAG. AT THIS TIME, THERE IS NO INFORMATION REGARDING PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15523576 ↗InjuryAug 11, 2022Device Alarm System; Device Alarm System
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE ADULT HEATER DUAL LIMB CIRCUIT CAUSED A VENTILATOR'S HEATER TO ALARM WHILE CONNECTED TO A PATIENT. THE END-USER ATTEMPTED TO CONNECT THE CIRCUIT TO OTHER HEATERS, AND ALL EXPERIENCED THE SAME ALARM ISSUE. THE END-USER THEN REMOVED THE CIRCUIT FROM THE PATIENT AND REPLACED THE CIRCUIT WITH A NEW ONE, AND THE ISSUE WAS RESOLVED. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15214792 ↗MalfunctionAug 8, 2022Moisture or Humidity Problem; Moisture or Humidity Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE 002002 - NEBULIZER WITH AIR ENTRAINMENT DOES NOT HUMIDIFY THE AIR DESPITE BEING REPLACED AND SECURING THE CONNECTION. THE HUMIDIFIED AIR IS NEEDED FOR A PATIENT WITH TRACHEOSTOMY. THERE WERE NO REPORTS OF INJURY/HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15189074 ↗MalfunctionAug 2, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem
THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THERE WAS A RUST-COLORED OR DARK-RED SUBSTANCE ATTACHED TO THE TO THE HUMIDIFICATION CHAMBER'S INNER PLASTIC AND FLOATING INSIDE THE WATER OF THE HUMIDIFICATION CHAMBER WHILE ON A PATIENT. NO HARM IS ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15154086 ↗MalfunctionAug 2, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem
THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THERE WAS A RUST-COLORED OR DARK-RED SUBSTANCE ATTACHED TO THE TO THE HUMIDIFICATION CHAMBER'S INNER PLASTIC AND FLOATING INSIDE THE WATER OF THE HUMIDIFICATION CHAMBER WHILE ON A PATIENT. NO HARM IS ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15154085 ↗MalfunctionMay 20, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem
THE CUSTOMER REPORTED THAT THE WATER STERILE F/INHALATION 2000ML 6/CS EXPERIENCED DEPOSITS IN THE CHAMBER DURING USE. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14459601 ↗MalfunctionMay 20, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem
THE CUSTOMER REPORTED THAT THE WATER STERILE F/INHALATION 2000ML 6/CS EXPERIENCED DEPOSITS IN THE CHAMBER DURING USE. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14459425 ↗MalfunctionApr 14, 2022Fire; Fire
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE 4681-504-INSPIRATORY LINE HTD 8FT ADULT 20/CS STARTED TO BURNED WHILE CONNECTED TO A PATIENT. THE CUSTOMER CONFIRM THAT THERE IS NO HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14116193 ↗MalfunctionDec 17, 2021Fire; Fire
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE 4681-504-INSPIRATORY LINE HTD 8FT ADULT 20/CS STARTED TO FIRE WHILE CONNECTED TO A PATIENT. THE CUSTOMER CONFIRM THAT THERE IS NO HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13025437 ↗MalfunctionOct 19, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem
THE CUSTOMER REPORTED THAT THE WATER STERILE F/INHALATION 2000ML 6/CS EXPERIENCED DEPOSITS IN THE CHAMBER DURING USE. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12661884 ↗
Other Carefusion Corporation products
All Carefusion Corporation devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page