MaxZero MZXT5307Recalls & reported problems
BD Switzerland Sàrl · Minibore Pressure reated Extension Set, removeable IV Connector, T-Connector 15 cm (6 in) · Class 2
All models of MaxZero (47)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10885403471155), received between Aug 8, 2023 and Aug 18, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash4
- Connection Problem2
- Material Separation2
- Complete Blockage1
- Defective Component1
- Infusion or Flow Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Insufficient Information1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 18, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD MAXZERO PRESSURE RATED EXTENSION SET WAS DISCONNECTING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER ABOUT THE J LOOP CONNECTION DISCONNECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22836259 ↗MalfunctionAug 7, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD MAXZERO PRESSURE RATED EXTENSION SET WAS DISCONNECTING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT THE J LOOP CONNECTION DISCONNECTS AND ALLOWS THE PATIENT TO BLEED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22743033 ↗MalfunctionJun 23, 2025Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT BD MAXZERO PRESSURE RATED EXTENSION SET WAS KINKED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT THE COUPLE OF OUR NURSES HAVE EXPRESSED CONCERNS ABOUT THE NEW T-CONNECTOR WE HAVE ON A PATIENT WITH A PICC. IT SEEMS TO BE VERY FLEXIBLE AND GETS TWISTED UP/KINKED VERY EASILY AND IN THIS CASE, WAS CAUSING AN OCCLUSION AND INABILITY TO INFUSE A MEDICATION FLUSH AND HAD 2 T-CONNECTORS BREAK WHERE THE T-CONNECTOR ATTACHES TO THE FLUIDS AND NOT THE END WHERE IT MEETS THE PORT TUBING. ADDITIONALLY, THERE ARE ISSUES AROUND THE LIPIDS AND BLOOD GETTING IN THE SMALL RESERVOIR OF THE T- CONNECTOR WHERE LABS ARE SUPPOSED TO BE OBTAINED FROM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22312665 ↗MalfunctionJun 23, 2025Complete Blockage; Leak/Splash; Complete Blockage; Leak/Splash
IT WAS REPORTED THAT BD MAXZERO PRESSURE RATED EXTENSION SET WAS KINKED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT THE COUPLE OF OUR NURSES HAVE EXPRESSED CONCERNS ABOUT THE NEW T-CONNECTOR WE HAVE ON A PATIENT WITH A PICC. IT SEEMS TO BE VERY FLEXIBLE AND GETS TWISTED UP/KINKED VERY EASILY AND IN THIS CASE, WAS CAUSING AN OCCLUSION AND INABILITY TO INFUSE A MEDICATION FLUSH AND HAD 2 T-CONNECTORS BREAK WHERE THE T-CONNECTOR ATTACHES TO THE FLUIDS AND NOT THE END WHERE IT MEETS THE PORT TUBING. ADDITIONALLY, THERE ARE ISSUES AROUND THE LIPIDS AND BLOOD GETTING IN THE SMALL RESERVOIR OF THE T- CONNECTOR WHERE LABS ARE SUPPOSED TO BE OBTAINED FROM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22312664 ↗MalfunctionJun 23, 2025Connection Problem; Connection Problem
IT WAS REPORTED THAT BD MAXZERO EXTENSION SET WAS SEPARATED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT THE SPIN COLLAR IS VERY STIFF AND DOES NOT SPIN FREELY. THIS CONNECTOR DOES NOT SECURELY FIT THE BARD MINILOC REF (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22312548 ↗MalfunctionJun 23, 2025Connection Problem; Connection Problem
IT WAS REPORTED THAT BD MAXZERO EXTENSION SET WAS SEPARATED THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT THE SPIN COLLAR IS VERY STIFF AND DOES NOT SPIN FREELY. THIS CONNECTOR DOES NOT SECURELY FIT THE BARD MINILOC REF (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22312547 ↗MalfunctionJun 17, 2025Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER THE SPIN COLLAR IS VERY STIFF AND DOES NOT SPIN FREELY. THE CONNECTOR DOES NOT SECURELY FIT THE MINILOC. STAFF REPORT IT "POPPING OFF" AND THEN BLEEDING VIA THE CATHETER ON TO BED LINES. IT WAS REPORTED THIS OCCURRED WITH THREE DEVICES. THIS REPORT ADDRESSES ALL THREE DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22243514 ↗MalfunctionJul 8, 2024Material Separation; Material Separation
IT WAS REPORTED THAT BD BD MAXZERO MULTI-FUSE EXTENSION SET WITH NEEDLELESS CONNECTOR WAS SEPARATING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT THE CLEAR END THAT SPINS TO CONNECT THE PORT TO THE CATHETER IS COMING COMPLETELY OFF OF THE DEVICE. SO THE REST OF THE T-PORT CANNOT STAY ATTACHED TO THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19697666 ↗MalfunctionJun 26, 2024Material Separation; Material Separation
NO ADDITIONAL INFO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19615061 ↗MalfunctionFeb 9, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD EXTENSION SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: CUSTOMER SHARED THAT THE EXTENSION SET AT THE T-PORT CONNECTION WAS LEAKING BLOOD DURING ASPIRATING. OR IT WOULD LEAK SALINE. CUSTOMER SAID THIS HAS HAPPENED ON 8 EXTENSION SETS OPENED BY NURSES IN THE ED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18674929 ↗MalfunctionAug 8, 2023Defective Component; Defective Component
IT WAS REPORTED WHILE USING BD MAXZERO MINIBORE PRESSURE RATED EXTENSION SET, REMOVABLE IV CONNECTOR THERE WAS COMPONENT SEPARATION. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INFUSION RN STATE THAT THE PRODUCT IS DEFECTIVE. HE STATES THAT THE SPIN COLLAR IS ABLE TO BE REMOVED WHEREAS THE OLD SMARTSITE PRODUCT (20043E) YOU COULD NOT REMOVE. ¿T-PORT FELL APART WHEN HOOKING TO IV CATHETER. HUB CAME APART FROM LUER LOCK.¿ ON THE SI EVENT.
Patient problems: Insufficient Information
MAUDE report 17483781 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page