Skip to content

BD Pyxis 1147-00Recalls & reported problems

Carefusion 303, Inc. · BD Pyxis MedBank Mini · Class 1

All models of BD Pyxis (26)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

10 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This BD Pyxis MedBank Mini is a medication dispensing cabinet used in hospitals. Many of the 180 reports describe electrical and software problems. The reports include 180 malfunctions, no injuries and no deaths. Reports from the last 12 months number 137, compared with 42 in the 12 months before. There are 10 open recalls, covering overheating fingerprint scanners, possible fire from fluid getting in, and software bugs.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
191
Deaths reported
0
Injuries reported
0
Recalls naming it
10

Summary

The FDA's MAUDE database contains 191 adverse event reports that list this exact device (DI 10885403517181), received between Oct 23, 2024 and Sep 28, 2026 — classified as 191 malfunctions.

The most frequently reported specific device problem is “Application Program Problem” (31 reports).

10 FDA recall records name this device's identifier, 10 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 5 reportsJun 2025: 11 reportsJul 2025: 12 reportsAug 2025: 5 reportsSep 2025: 2 reportsOct 2025: 3 reportsNov 2025: 29 reportsDec 2025: 5 reportsJan 2026: 9 reportsFeb 2026: 12 reportsMar 2026: 14 reportsApr 2026: 10 reportsMay 2026: 6 reportsJun 2026: 20 reportsJul 2026: 21 reportsAug 2026: 11 reportsSep 2026: 11 reportsOct 2026: 0 reports20242025202629
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 30, 2026Open, Classified

    Fingerprint scanner may overheat to a temperature to cause 1st degree burn.

    Z-2304-2026 · Root cause: Component design/selection · Recall details · FDA record ↗

  2. Sep 30, 2025Open, Classified

    Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

    Z-0367-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Jul 10, 2025Open, Classified

    Reports of delayed access to medication in automated dispensing cabinets because of the override/Add Item workflow.

    Z-2321-2025 · Root cause: Use error · Recall details · FDA record ↗

  4. Mar 11, 2025Open, Classified

    Users trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a software issue that results in 2 destock transactions and 2 stock transactions, which results in doubling the bin quantity, which may result in a delay in access to desired medications and/or supplies, which may lead to temporary body function impairment.

    Z-1639-2025 · Root cause: Software design · Recall details · FDA record ↗

  5. Mar 11, 2025Open, Classified

    Due to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect transaction quantity to be recorded. This bug can also result in the CUBIE opening more times than needed for the transaction to be completed, which may result in a patient receiving more than the intended dose of medication.

    Z-1545-2025 · Root cause: Software design · Recall details · FDA record ↗

  6. Jan 8, 2025Open, Classified

    Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

    Z-1227-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  7. Jan 8, 2025Open, Classified

    Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

    Z-1230-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Jan 8, 2025Open, Classified

    Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

    Z-1228-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  9. Jan 8, 2025Open, Classified

    Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.

    Z-1229-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  10. Oct 5, 2023Open, Classified

    If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.

    Z-1552-2025 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Application Program Problem31
  • Electrical /Electronic Property Problem31
  • Computer Software Problem27
  • Labelling, Instructions for Use or Training Problem21
  • Installation-Related Problem19
  • Failure to Eject14
  • Mechanical Problem10
  • Use of Device Problem9

Patient problems

  • No Clinical Signs, Symptoms or Conditions191

Recent adverse event reports

The 25 most recent of 191 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026No Apparent Adverse Event

    It was reported that when using the BD Pyxis¿ MedBank Mini, the user was unable to pull medication.The customer stated that this malfunction occurred while dispensing the medication.There were no adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914942 ↗
  • MalfunctionSep 28, 2026Electrical /Electronic Property Problem; Application Program Problem

    It was reported that when using the BD Pyxis¿ MedBank Mini the station was stuck on blue screen. The customer attempted to reboot the machine and unplugged and replugged the power cable; however, the issue persisted. However, there were no delay or adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911626 ↗
  • MalfunctionSep 24, 2026Use of Device Problem

    It was reported that when using the BD Pyxis¿ MedBank Mini system user was unable to see certain medications. The customer reported that a malfunction took place when the user tried to dispense medication to the patient. However, there were no delay or adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50878469 ↗
  • MalfunctionSep 22, 2026Power Problem

    It was reported that when using the BD Pyxis¿ MedBank Mini cabinet 01 was repeatedly found offline or powered off due to a circuit breaker tripping, resulting in loss of cabinet power and communication.However, there were no delays, adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50855360 ↗
  • MalfunctionSep 21, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50838523 ↗
  • MalfunctionSep 18, 2026Application Program Problem

    It was reported that when using the BD Pyxis¿ MedBank Mini, when the user tried to retrieve medication (CIPROFLOXACIN HCL 250MG TABLET), the screen was frozen. The customer stated that there was a delay in patient care. However, there were no adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50815850 ↗
  • MalfunctionSep 18, 2026Application Program Problem

    It was reported that when using the BD Pyxis¿ MedBank Mini medication was not on profile. The customer reported that a malfunction took place when the user tried to dispense medication to the patient. However, there were no delay or adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50814848 ↗
  • MalfunctionSep 15, 2026Loose or Intermittent Connection

    It was reported that when using the BD Pyxis¿ MedBank Mini, monitor was not turning on. However, there were no delays or adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772148 ↗
  • MalfunctionSep 15, 2026Application Program Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769754 ↗
  • MalfunctionSep 3, 2026Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50587772 ↗
  • MalfunctionSep 3, 2026Labelling, Instructions for Use or Training Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50582058 ↗
  • MalfunctionAug 28, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50494179 ↗
  • MalfunctionAug 27, 2026Electrical /Electronic Property Problem; Complete Loss of Power

    It was reported that when using the BD Pyxis¿ MedBank Mini system was powered off. The customer stated that there was a delay in patient care. There were no adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50475923 ↗
  • MalfunctionAug 25, 2026Application Program Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50429192 ↗
  • MalfunctionAug 14, 2026Electrical /Electronic Property Problem; Loose or Intermittent Connection

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50252390 ↗
  • MalfunctionAug 14, 2026Labelling, Instructions for Use or Training Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50249308 ↗
  • MalfunctionAug 14, 2026Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50247748 ↗
  • MalfunctionAug 13, 2026Power Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50231464 ↗
  • MalfunctionAug 11, 2026Complete Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50206344 ↗
  • MalfunctionAug 10, 2026Communication or Transmission Problem

    It was reported that when using the BD Pyxis¿ MedBank Mini reported being prompted for a witness when removing controlled medications. Review showed that Tramadol 50MG was configured to require a witness for issue and cycle count. Med bank my qLink (MQL) indicated the facility was not syncing, and remote smart service (RSS) showed the facility had been offline for approximately 4 days and 7 hours. The specific user involved could not be identified, and witness bypass settings could not be verified. The customer reported that a malfunction took place when the user tried to dispense medication to the patient. However, there were no delay or adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50189675 ↗
  • MalfunctionAug 7, 2026Mechanical Problem; Failure to Align

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50170264 ↗
  • MalfunctionAug 3, 2026Application Program Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50113513 ↗
  • MalfunctionJul 28, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50054559 ↗
  • MalfunctionJul 27, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50035840 ↗
  • MalfunctionJul 27, 2026Installation-Related Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50032624 ↗

Other CABINET, TABLE AND TRAY, ANESTHESIA devices

All CABINET, TABLE AND TRAY, ANESTHESIA devices →

Sources and updates